Intelligent telemetric stent for wireless monitoring of intravascular pressure and its in vivo testing

被引:0
|
作者
Xing Chen
Daniel Brox
Babak Assadsangabi
York Hsiang
Kenichi Takahata
机构
[1] University of British Columbia,Department of Electrical and Computer Engineering
[2] University of British Columbia,Department of Surgery, Vancouver General Hospital
来源
Biomedical Microdevices | 2014年 / 16卷
关键词
Stents; Restenosis; Wireless sensing; Radio-frequency antennas; Pressure sensors; Micro-electro-mechanical systems; Stainless steels;
D O I
暂无
中图分类号
学科分类号
摘要
This paper reports a sensor-integrated telemetric stent targeted at wireless detection and monitoring of restenosis, a common vascular complication induced by stent implantation. The developed “smart” stent incorporates the design and fabrication approaches that raise the practicality of the device, being tested in an in vivo study that validates its operating principle. The stent is produced to have a gold-coated helical-like structure that serves as a high-performance inductor/antenna and integrated with a novel capacitive pressure sensor chip, all based on medical-grade stainless steel. The stent device forms an inductor-capacitor resonant tank that enables radio-frequency (RF) wireless pressure sensing in an operating frequency range of 30–80 MHz. With an overall length of 20 mm, the device is designed to be compatible with standard balloon catheters and necessary crimping process. The balloon-expanded devices are characterized in saline and blood to determine selective coating of passivation layer, Parylene C, with tailored thicknesses in order to maximize both RF and sensing abilities. In vitro testing of the devices reveals a frequency sensitivity up to 146 ppm/mmHg over a pressure range of 250 mmHg. Tests in pig models show wireless detection of device’s resonance and frequency response to variations in local blood pressure, the targeted function of the device.
引用
收藏
页码:745 / 759
页数:14
相关论文
共 46 条
  • [31] In Vitro and In Vivo characterization of wireless and passive micro system enabling gastrointestinal pressure monitoring
    Shi, Qiang
    Wang, Junbo
    Chen, Deyong
    Chen, Jian
    Li, Jing
    Bao, Kaikai
    [J]. BIOMEDICAL MICRODEVICES, 2014, 16 (06) : 859 - 868
  • [32] Intelligent Ureteral Stent Placeable via Standard Procedure for Kidney Pressure Telemetry: An Ex-Vivo Study
    Darestani, Mohammad Reza Yousefi
    Lange, Dirk
    Chew, Ben H.
    Takahata, Kenichi
    [J]. ANNALS OF BIOMEDICAL ENGINEERING, 2024,
  • [33] Wireless pressure sensor integrated with a 3D printed polymer stent for smart health monitoring
    Park, Jongsung
    Kim, Ji-Kwan
    Kim, Dong-Su
    Shanmugasundaram, Arunkumar
    Park, Su A.
    Kang, Sohi
    Kim, Sung-Ho
    Jeong, Myung Ho
    Lee, Dong-Weon
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2019, 280 : 201 - 209
  • [34] A wireless microsensor for monitoring flow and pressure in a blood vessel utilizing a dual-inductor antenna stent and two pressure sensors
    Takahata, K
    DeHennis, A
    Wise, KD
    Gianchandani, YB
    [J]. MEMS 2004: 17TH IEEE INTERNATIONAL CONFERENCE ON MICRO ELECTRO MECHANICAL SYSTEMS, TECHNICAL DIGEST, 2004, : 216 - 219
  • [35] Temperature Self-Compensating Intelligent Wireless Measuring Contact Lens for Quantitative Intraocular Pressure Monitoring
    Li X.
    Chen W.
    Li H.
    Shen B.
    He J.
    Gao H.
    Bin F.
    Li H.
    Xiao D.
    [J]. ACS Applied Materials and Interfaces, 2024, 16 (17): : 22522 - 22531
  • [36] Temperature Self-Compensating Intelligent Wireless Measuring Contact Lens for Quantitative Intraocular Pressure Monitoring
    Li, Xu
    Chen, Wei
    Li, Hongyang
    Shen, Biwen
    He, Jiangang
    Gao, Huanlin
    Bin, Fengjiao
    Li, Hui
    Xiao, Dengbao
    [J]. ACS APPLIED MATERIALS & INTERFACES, 2024, 16 (17) : 22522 - 22531
  • [37] Wireless Blood Pressure Monitoring with a Novel Implantable Device: Long-Term In Vivo Results
    Cleven, Nina J.
    Isfort, Peter
    Penzkofer, Tobias
    Woitok, Anna
    Hermanns-Sachweh, Benita
    Steinseifer, Ulrich
    Schmitz-Rode, Thomas
    [J]. CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2014, 37 (06) : 1580 - 1588
  • [38] Wireless Blood Pressure Monitoring with a Novel Implantable Device: Long-Term In Vivo Results
    Nina J. Cleven
    Peter Isfort
    Tobias Penzkofer
    Anna Woitok
    Benita Hermanns-Sachweh
    Ulrich Steinseifer
    Thomas Schmitz-Rode
    [J]. CardioVascular and Interventional Radiology, 2014, 37 : 1580 - 1588
  • [39] Design of remote control device using wireless sensor network and its use in intelligent monitoring of farmland information
    Jian Zhang
    Ying Wang
    [J]. EURASIP Journal on Wireless Communications and Networking, 2021
  • [40] Design of remote control device using wireless sensor network and its use in intelligent monitoring of farmland information
    Zhang, Jian
    Wang, Ying
    [J]. EURASIP JOURNAL ON WIRELESS COMMUNICATIONS AND NETWORKING, 2021, 2021 (01)