Magnetically actuated glaucoma drainage device for regulating intraocular pressure after implantation

被引:0
|
作者
Inês C. F. Pereira
Ralph J. S. van Mechelen
Hans M. Wyss
Leonard Pinchuk
Henny J. M. Beckers
Jaap M. J. den Toonder
机构
[1] Eindhoven University of Technology,Microsystems, Department of Mechanical Engineering
[2] Eindhoven University of Technology,Institute for Complex Molecular Systems (ICMS)
[3] University Eye Clinic Maastricht,Ophthalmic Biophysics Center, Bascom Palmer Eye Institute
[4] Maastricht University Medical Centre+ (MUMC+),undefined
[5] InnFocus,undefined
[6] Inc.,undefined
[7] a Santen Company,undefined
[8] University of Miami Miller School of Medicine,undefined
关键词
D O I
暂无
中图分类号
学科分类号
摘要
The key risk factor for glaucoma is increased intraocular pressure (IOP). Glaucoma drainage devices implanted in the eye can reduce IOP and thus stop disease progression. However, most devices currently used in clinical practice are passive and do not allow for postsurgical IOP control, which may result in serious complications such as hypotony (i.e., excessively low IOP). To enable noninvasive IOP control, we demonstrate a novel, miniature glaucoma implant that will enable the repeated adjustment of the hydrodynamic resistance after implantation. This is achieved by integrating a magnetic microvalve containing a micropencil-shaped plug that is moved using an external magnet, thereby opening or closing fluidic channels. The microplug is made from biocompatible poly(styrene-block-isobutylene-block-styrene) (SIBS) containing iron microparticles. The complete implant consists of an SIBS drainage tube and a housing element containing the microvalve and fabricated with hot embossing using femtosecond laser-machined glass molds. Using in vitro and ex vivo microfluidic experiments, we demonstrate that when the microvalve is closed, it can provide sufficient hydrodynamic resistance to overcome hypotony. Valve function is repeatable and stable over time. Due to its small size, our implant is a promising, safe, easy-to-implant, minimally invasive glaucoma surgery device.
引用
收藏
相关论文
共 50 条
  • [1] Magnetically actuated glaucoma drainage device for regulating intraocular pressure after implantation
    Beckers, H. J. M.
    Pereira, I. C. F.
    van Mechelen, R. J. S.
    Wyss, H. M.
    Pinchuk, L.
    den Toonder, J. M. J.
    [J]. ACTA OPHTHALMOLOGICA, 2023, 101 : 17 - 17
  • [2] Magnetically actuated glaucoma drainage device for regulating intraocular pressure after implantation
    Pereira, Ines C. F.
    Wyss, Hans M.
    Beckers, Henny J.
    den Toonder, Jaap M. J.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2022, 63 (07)
  • [3] Magnetically actuated glaucoma drainage device for regulating intraocular pressure after implantation
    Pereira, Ines C. F.
    van Mechelen, Ralph J. S.
    Wyss, Hans M.
    Pinchuk, Leonard
    Beckers, Henny J. M.
    den Toonder, Jaap M. J.
    [J]. MICROSYSTEMS & NANOENGINEERING, 2023, 9 (01)
  • [4] Continuous intraocular pressure (IOP) measurement during glaucoma drainage device implantation
    Gouws, Pieter
    Moss, Edward B.
    Trope, Graham E.
    Ethier, C. Ross
    [J]. JOURNAL OF GLAUCOMA, 2007, 16 (03) : 329 - 333
  • [5] Simultaneous Implantation of an Ahmed and Baerveldt Glaucoma Drainage Device for Uncontrolled Intraocular Pressure in Advanced Glaucoma
    Rao, Veena S.
    Christenbury, Joseph
    Lee, Paul
    Allingham, Rand
    Herndon, Leon
    Challa, Pratap
    [J]. JOURNAL OF GLAUCOMA, 2017, 26 (02) : 101 - 106
  • [6] Corneal graft survival and intraocular pressure control after penetrating keratoplasty and glaucoma drainage device implantation
    Arroyave, CP
    Scott, IU
    Fantes, FE
    Feuer, WJ
    Murray, TG
    [J]. OPHTHALMOLOGY, 2001, 108 (11) : 1978 - 1985
  • [7] Simultaneous Implantation of an Ahmed and Baerveldt Glaucoma Drainage Device for Uncontrolled Intraocular Pressure in Advanced Glaucoma: Comment
    Gupta, Richa
    Gupta, Surabhi
    [J]. JOURNAL OF GLAUCOMA, 2018, 27 (01) : E24 - E24
  • [8] Diplopia after glaucoma drainage device implantation
    Abdelaziz, Amany
    Capo, Hilda
    Banitt, Michael R.
    Schiffman, Joyce
    Feuer, William J.
    McKeown, Craig A.
    Spencer, Nehemiah E.
    Parrish, Richard K.
    [J]. JOURNAL OF AAPOS, 2013, 17 (02): : 192 - 196
  • [9] The effect of trabeculectomy and glaucoma drainage device implantation on postural intraocular pressure changes in glaucomatous eyes
    Belkin, Avner
    Dar, Nimrod
    Pillar, Shani
    Tempelhof, Ortal F.
    Barkana, Yaniv
    Sheiman, Vadim
    Naftali Ben Haim, Liron
    Geffen, Noa
    [J]. ACTA OPHTHALMOLOGICA, 2021, 99 (07) : E1112 - E1117
  • [10] Author Response, Regarding "Simultaneous Implantation of an Ahmed and Baerveldt Glaucoma Drainage Device for Uncontrolled Intraocular Pressure in Advanced Glaucoma"
    Rao, Veena
    Challa, Pratap
    [J]. JOURNAL OF GLAUCOMA, 2018, 27 (01) : E24 - E25