Analysis of the concurrent validity and reliability of five common clinical goniometric devices

被引:0
|
作者
Sirirat Kiatkulanusorn
Nongnuch Luangpon
Wirasinee Srijunto
Sarawoot Watechagit
Krittipat Pitchayadejanant
Sireetorn Kuharat
O. Anwar Bég
Bhornluck Paepetch Suato
机构
[1] Burapha University,Department of Physical Therapy, Faculty of Allied Health Sciences
[2] Mahidol University,Faculty of Engineering
[3] Burapha University International College,Corrosion Lab, Multi
[4] Burapha University,Physical Engineering Sciences Group (MPESG), Department of Mechanical/Aeronautical Engineering
[5] Salford University,undefined
来源
关键词
D O I
暂无
中图分类号
学科分类号
摘要
Measurement errors play an important role in the development of goniometric equipment, devices used to measure range of motion. Reasonable validity and reliability are critical for both the device and examiner before and after to testing in human subjects. The objective is to evaluate the concurrent validity and reliability of five different clinical goniometric devices for the purpose of establishing an acceptable measurement error margin for a novel device. We explored the validity and inter- and intrarater reliability scores of five goniometric devices namely (i) the universal goniometer (UG), a two-armed hand-held goniometer, (ii) the inclinometer (IC), featuring a single base, fluid level, and gravity-weighted inclinometer, (iii) the digital inclinometer (DI), functioning as both a DI and dynamometer, (iv) the smartphone application (SA), employing gyroscope-based technology within a smartphone platform application and (v) the modified inclinometer (MI), a gravity pendulum-based inclinometer equipped with a specialized fixing apparatus. Measurements were obtained at 12 standard angles and 8 human shoulder flexion angles ranging from 0° to 180°. Over two testing sessions, 120 standardized angle measurements and 160 shoulder angle measurements from 20 shoulders were repetitively taken by three examiners for each device. The intraclass correlation coefficient (ICC), standard error of measurement (SEM), and minimal detectable change (MDC) were calculated to assess reliability and validity. Concurrent validity was also evaluated through the execution of the 95% limit of agreement (95% LOA) and Bland–Altman plots, with comparisons made to the UG. The concurrent validity for all device pairs was excellent in both study phases (ICC > 0.99, 95% LOA − 4.11° to 4.04° for standard angles, and − 10.98° to 11.36° for human joint angles). Inter- and intrarater reliability scores for standard angles were excellent across all devices (ICC > 0.98, SEM 0.59°–1.75°, MDC 1°–4°), with DI showing superior reliability. For human joint angles, device reliability ranged from moderate to excellent (ICC 0.697–0.975, SEM 1.93°–4.64°, MDC 5°–11° for inter-rater reliability; ICC 0.660–0.996, SEM 0.77°–4.06°, MDC 2°–9° for intra-rater reliability), with SA demonstrating superior reliability. Wider angle measurement however resulted in reduced device reliability. In conclusion, our study demonstrates that it is essential to assess measurement errors independently for standard and human joint angles. The DI is the preferred reference for standard angle testing, while the SA is recommended for human joint angle testing. Separate evaluations across the complete 0°–180° range offer valuable insights.
引用
收藏
相关论文
共 50 条
  • [22] Reliability and concurrent validity of a new iPhone® goniometric application for measuring active wrist range of motion: a cross-sectional study in asymptomatic subjects
    Pourahmadi, Mohammad Reza
    Takamjani, Ismail Ebrahimi
    Sarrafzadeh, Javad
    Bahramian, Mehrdad
    Mohseni-Bandpei, Mohammad Ali
    Rajabzadeh, Fatemeh
    Taghipour, Morteza
    JOURNAL OF ANATOMY, 2017, 230 (03) : 484 - 495
  • [23] STATISTICAL METHODOLOGY FOR THE CONCURRENT ASSESSMENT OF INTERRATER AND INTRARATER RELIABILITY - USING GONIOMETRIC MEASUREMENTS AS AN EXAMPLE
    ELIASZIW, M
    YOUNG, SL
    WOODBURY, MG
    FRYDAYFIELD, K
    PHYSICAL THERAPY, 1994, 74 (08): : 777 - 788
  • [24] Reliability and concurrent validity of Futrex and bioelectrical impedance
    Vehrs, P
    Morrow, JR
    Butte, N
    INTERNATIONAL JOURNAL OF SPORTS MEDICINE, 1998, 19 (08) : 560 - 566
  • [25] Reliability and concurrent validity of the Ccomputer Workstation Checklist
    Baker, Nancy A.
    Livengood, Heather
    Jacobs, Karen
    WORK-A JOURNAL OF PREVENTION ASSESSMENT & REHABILITATION, 2013, 45 (02): : 213 - 221
  • [26] Reliability and concurrent validity of the Infant Motor Profile
    Heineman, Kirsten R.
    Middelburg, Karin J.
    Bos, Arend F.
    Eidhof, Lieke
    La Bastide-Van Gemert, Sacha
    Van den Heuvel, Edwin R.
    Hadders-Algra, Mijna
    DEVELOPMENTAL MEDICINE AND CHILD NEUROLOGY, 2013, 55 (06): : 539 - 545
  • [27] THE STUDENT GOALS EXPLORATION - RELIABILITY AND CONCURRENT VALIDITY
    STARK, JS
    BENTLEY, RJ
    LOWTHER, MA
    SHAW, KM
    EDUCATIONAL AND PSYCHOLOGICAL MEASUREMENT, 1991, 51 (02) : 413 - 422
  • [28] Concurrent Validity and Reliability for the Communication Complexity Scale
    Brady, Nancy C.
    Fleming, Kandace
    Romine, Rebecca Swinburne
    Holbrook, Alison
    Muller, Kristen
    Kasari, Connie
    AMERICAN JOURNAL OF SPEECH-LANGUAGE PATHOLOGY, 2018, 27 (01) : 237 - 246
  • [29] RELIABILITY AND CONCURRENT VALIDITY OF HEADACHE QUESTIONNAIRE DATA
    ANDRASIK, F
    HOLROYD, KA
    HEADACHE, 1980, 20 (01): : 44 - 46
  • [30] RELIABILITY AND CONCURRENT VALIDITY OF LIGHTS RETENTION SCALE
    SANDOVAL, J
    PSYCHOLOGY IN THE SCHOOLS, 1980, 17 (04) : 442 - 445