Influence of pupil dilation on retinal nerve fibre layer measurements with spectral domain OCT

被引:18
|
作者
Massa, G. C. [1 ]
Vidotti, V. G. [1 ]
Cremasco, F. [1 ]
Lupinacci, A. P. C. [1 ]
Costa, V. P. [1 ]
机构
[1] Univ Estadual Campinas, Dept Ophthalmol, BR-01244001 Sao Paulo, Brazil
基金
巴西圣保罗研究基金会;
关键词
OCT tomography; mydriasis; glaucoma; OPTICAL COHERENCE TOMOGRAPHY; THICKNESS MEASUREMENTS; ULTRAHIGH-RESOLUTION; MACULAR THICKNESS; STRATUS OCT; GLAUCOMA; REPRODUCIBILITY; CIRRUS; HEAD;
D O I
10.1038/eye.2010.72
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose To evaluate the influence of pupil dilation on retinal nerve fibre layer (RNFL) measurements with spectral domain optical coherence tomography (OCT). Patients and Methods In total, 29 healthy individuals and 26 glaucoma patients underwent RNFL measurements with the Cirrus HD-OCT Model 4000 (Carl Zeiss Meditec Inc.) before and 30-40 min after pupil dilation with one drop of tropicamide 1%. Average thickness, quadrant thickness, and clock-hour thickness measurements were compared with the paired Student's t-test. We also compared the quality scores of the images obtained pre- and post-pupil dilation. Results Mean ages in the glaucoma and control groups were 58.3 +/- 13.4 and 41.6 +/- 16.4 years respectively (P<0.001). Mean deviation values were -6.96+6.31 dB in the glaucoma group and -1.26+0.79 dB in the control group (P<0.001). Mean RNFL measurements obtained in the glaucoma group were significantly lower than those obtained in the control group (P<0.001). There were no statistically significant differences between mean quality scores obtained before and after dilation neither in the glaucoma group (7.73 +/- 0.92 vs 7.54 +/- 1.10, P=0.232) nor in the control group (8.14 +/- 0.88 vs 8.00 +/- 0.71, P-0.380). There was no statistically significant difference between mean RNFL measurements obtained pre- and post-pupil dilation neither among normal individuals (P>0.05), nor among glaucoma patients (P>0.05). Conclusion The results of this study indicate that RNFL measurements obtained with spectral domain OCT are not influenced by pupil size. Eye (2010) 24, 1498-1502; doi:10.1038/eye.2010.72; published online 28 May 2010
引用
收藏
页码:1498 / 1502
页数:5
相关论文
共 50 条
  • [11] Retinal nerve fibre layer measurements: a concern (Comparison of retinal nerve fibre layer measurements from time domain and spectral domain optical coherence tomography systems. Vol. 44[5])
    Wiwanitkit, Viroj
    [J]. CANADIAN JOURNAL OF OPHTHALMOLOGY-JOURNAL CANADIEN D OPHTALMOLOGIE, 2010, 45 (02): : 176 - 176
  • [12] Comparison of Influence of Cataract and Small Pupil on Retinal Nerve Fiber Layer Thickness Measurements Between Time Domain and Spectral Domain Optical Coherence Tomography
    Cheng, C.
    Yip, L. W.
    Natividad, M. G.
    Yong, V.
    Lim, B.
    Wong, H.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)
  • [13] VARIABLITY OF PERIPAPILLARY RETINAL NERVE FIBRE LAYER MEASUREMENTS IN FOUR SPECTRAL DOMAIN OPTICAL COHERENCE TOMOGRAPHS
    Pearce, John G.
    Pearce, Megan E.
    Maddess, Ted
    [J]. CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2013, 41 : 131 - 131
  • [14] Retinal Nerve Fiber Layer Thickness in Normals Measured by Spectral Domain OCT
    Bendschneider, Delia
    Tornow, Ralf P.
    Horn, Folkert K.
    Laemmer, Robert
    Roessler, Christopher W.
    Juenemann, Anselm G.
    Kruse, Friedrich E.
    Mardin, Christian Y.
    [J]. JOURNAL OF GLAUCOMA, 2010, 19 (07) : 475 - 482
  • [15] Comparison of retinal nerve fibre layer measurements from time domain and spectral domain optical coherence tomography systems
    Johnson, Davin E.
    El-Defrawy, Sherif R.
    Almeida, David R. P.
    Campbell, Robert J.
    [J]. CANADIAN JOURNAL OF OPHTHALMOLOGY-JOURNAL CANADIEN D OPHTALMOLOGIE, 2009, 44 (05): : 562 - 566
  • [16] Reproducibility of peripapillary retinal nerve fiber layer thickness measurements using Spectral Domain OCT in Brazilian patients
    Toscano, Daniela Araujo
    de Avila, Marcos Pereira
    Catai Chalita, Maria Regina
    [J]. ARQUIVOS BRASILEIROS DE OFTALMOLOGIA, 2012, 75 (05) : 320 - 323
  • [17] Influence of Cataract on Time Domain and Spectral Domain Optical Coherence Tomography Retinal Nerve Fiber Layer Measurements
    Kim, Na Rae
    Lee, Hun
    Lee, Eun Suk
    Kim, Ji Hyun
    Hong, Samin
    Seong, Gong Je
    Kim, Chan Yun
    [J]. JOURNAL OF GLAUCOMA, 2012, 21 (02) : 116 - 122
  • [18] The Effect of Pupillary Dilation on Retinal Nerve Fiber Layer Thickness Measurements Taken by Spectral Domain Optical Coherence Tomography
    Cam, Duygu
    Karabag, Revan Yildirim
    Arikan, Gul
    Gunenc, Uzeyir
    [J]. TURK OFTALMOLOJI DERGISI-TURKISH JOURNAL OF OPHTHALMOLOGY, 2014, 44 (06): : 454 - 457
  • [19] Correlation Between Spectral Domain OCT Retinal Nerve Fibre Layer Thickness and Multifocal Pattern Electroretinogram in Advanced Retinitis Pigmentosa
    Sliesoraityte, Ieva
    Troeger, Eric
    Bernd, Antje
    Kurtenbach, Anne
    Zrenner, Eberhart
    [J]. RETINAL DEGENERATIVE DISEASES, 2012, 723 : 471 - 478
  • [20] SPECTRAL DOMAIN OCT MEASURED RETINAL NERVE FIBRE LAYER THICKNESS AND OPTIC DISC PARAMETERS IN CHILDREN: NORMATIVE DATA AND ASSOCIATIONS
    Tariq, Yasser
    Li, Haitao
    Burlutsky, George
    Mitchell, Paul
    [J]. CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2011, 39 : 57 - 58