Influence of pupil dilation on retinal nerve fibre layer measurements with spectral domain OCT

被引:18
|
作者
Massa, G. C. [1 ]
Vidotti, V. G. [1 ]
Cremasco, F. [1 ]
Lupinacci, A. P. C. [1 ]
Costa, V. P. [1 ]
机构
[1] Univ Estadual Campinas, Dept Ophthalmol, BR-01244001 Sao Paulo, Brazil
基金
巴西圣保罗研究基金会;
关键词
OCT tomography; mydriasis; glaucoma; OPTICAL COHERENCE TOMOGRAPHY; THICKNESS MEASUREMENTS; ULTRAHIGH-RESOLUTION; MACULAR THICKNESS; STRATUS OCT; GLAUCOMA; REPRODUCIBILITY; CIRRUS; HEAD;
D O I
10.1038/eye.2010.72
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose To evaluate the influence of pupil dilation on retinal nerve fibre layer (RNFL) measurements with spectral domain optical coherence tomography (OCT). Patients and Methods In total, 29 healthy individuals and 26 glaucoma patients underwent RNFL measurements with the Cirrus HD-OCT Model 4000 (Carl Zeiss Meditec Inc.) before and 30-40 min after pupil dilation with one drop of tropicamide 1%. Average thickness, quadrant thickness, and clock-hour thickness measurements were compared with the paired Student's t-test. We also compared the quality scores of the images obtained pre- and post-pupil dilation. Results Mean ages in the glaucoma and control groups were 58.3 +/- 13.4 and 41.6 +/- 16.4 years respectively (P<0.001). Mean deviation values were -6.96+6.31 dB in the glaucoma group and -1.26+0.79 dB in the control group (P<0.001). Mean RNFL measurements obtained in the glaucoma group were significantly lower than those obtained in the control group (P<0.001). There were no statistically significant differences between mean quality scores obtained before and after dilation neither in the glaucoma group (7.73 +/- 0.92 vs 7.54 +/- 1.10, P=0.232) nor in the control group (8.14 +/- 0.88 vs 8.00 +/- 0.71, P-0.380). There was no statistically significant difference between mean RNFL measurements obtained pre- and post-pupil dilation neither among normal individuals (P>0.05), nor among glaucoma patients (P>0.05). Conclusion The results of this study indicate that RNFL measurements obtained with spectral domain OCT are not influenced by pupil size. Eye (2010) 24, 1498-1502; doi:10.1038/eye.2010.72; published online 28 May 2010
引用
收藏
页码:1498 / 1502
页数:5
相关论文
共 50 条
  • [1] Influence of pupil dilation on retinal nerve fibre layer measurements with spectral domain OCT
    G C Massa
    V G Vidotti
    F Cremasco
    A P C Lupinacci
    V P Costa
    [J]. Eye, 2010, 24 : 1498 - 1502
  • [2] Effect of pupil dilation on retinal nerve fibre layer thickness measurements and their repeatability with Cirrus HD-OCT
    G Savini
    M Carbonelli
    V Parisi
    P Barboni
    [J]. Eye, 2010, 24 : 1503 - 1508
  • [3] Effect of pupil dilation on retinal nerve fibre layer thickness measurements and their repeatability with Cirrus HD-OCT
    Savini, G.
    Carbonelli, M.
    Parisi, V.
    Barboni, P.
    [J]. EYE, 2010, 24 (09) : 1503 - 1508
  • [4] Spectral-domain OCT peripapillary retinal nerve fibre layer thickness measurements in patients with stargardt disease
    Genead, Mohamed A.
    Fishman, Gerald A.
    Anastasakis, Anastasios
    [J]. BRITISH JOURNAL OF OPHTHALMOLOGY, 2011, 95 (05) : 689 - 693
  • [5] The influence of axial length on retinal nerve fibre layer thickness and optic-disc size measurements by spectral-domain OCT
    Savini, Giacomo
    Barboni, Piero
    Parisi, Vincenzo
    Carbonelli, Michele
    [J]. BRITISH JOURNAL OF OPHTHALMOLOGY, 2012, 96 (01) : 57 - 61
  • [6] Influence of Pupil Dilation on the Measurement of Retinal Nerve Fiber Layer Thickness by Cirrus HD-OCT
    Silva, A. F.
    Queiroz, I. C.
    Melo, L. A. S., Jr.
    Tavares, I. M.
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2010, 51 (13)
  • [7] Comparison of the influence of cataract and pupil size on retinal nerve fibre layer thickness measurements with time-domain and spectral-domain optical coherence tomography
    Cheng, Clarissa S. M.
    Natividad, Maricel G.
    Earnest, Arul
    Yong, Vernon
    Lim, Boon A.
    Wong, Hon T.
    Yip, Leonard W.
    [J]. CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2011, 39 (03): : 215 - 221
  • [8] Influence of pupil size and cataract on retinal nerve fiber layer thickness measurements by stratus OCT
    Savini, Giacomo
    Zanini, Maurizio
    Barboni, Piero
    [J]. JOURNAL OF GLAUCOMA, 2006, 15 (04) : 336 - 340
  • [9] Effect of pupil dilation on retinal nerve fiber layer measurements in patients with glaucoma
    Liu, Ji
    Ryg, Peter
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2017, 58 (08)
  • [10] Intrasession, intersession, and interexaminer variabilities of retinal nerve fiber layer measurements with spectral-domain OCT
    Cremasco, Fernanda
    Massa, Graziela
    Vidotti, Vanessa Goncalves
    de Carvalho Lupinacci, Alvaro Pedroso
    Costa, Vital Paulino
    [J]. EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2011, 21 (03) : 264 - 270