Three-dimensional versus four-dimensional dose calculation for volumetric modulated arc therapy of hypofractionated treatments

被引:23
|
作者
Ehrbar, Stefanie [1 ]
Lang, Stephanie
Stieb, Sonja
Riesterer, Oliver
Stark, Luisa Sabrina
Guckenberger, Matthias
Kloeck, Stephan
机构
[1] Univ Zurich Hosp, Dept Radiat Oncol, Ramistr 100, CH-8091 Zurich, Switzerland
来源
ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK | 2016年 / 26卷 / 01期
基金
瑞士国家科学基金会;
关键词
VMAT; respiratory motion; interplay effect; SBRT; DOSIMETRIC IMPACT; BREATHING MOTION; TARGET VOLUME; ORGAN MOTION; DELIVERY; RADIOTHERAPY; SIMULATION;
D O I
10.1016/j.zemedi.2015.06.010
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Respiratory motion is a non-negligible source of uncertainty in radiotherapy. A common approach is to delineate the target volume in all respiratory phases (ITV) and to calculate a treatment plan using the average reconstruction of the four-dimensional computed tomography (4DCT) scans. In this study the extent of the interplay effect caused by interaction between dynamic dose delivery and respiratory tumor motion, as well as other motion effects were investigated. These effects are often ignored when the ITV concept is used. Methods and Materials: Nine previously treated patients with in ten abdominal or thoracic cancer lesions (3 liver, 3 adrenal glands and 4 lung lesions) were selected for this planning study. For all patients, phase-sorted respiration-correlated 4DCT scans were taken, and volumetric modulated arc therapy (VMAT) treatments were planned using the ITV concept. Margins from ITV to planning target volume (PTV) of 3-10 mm were used. Plans were optimized and dose distributions were calculated on the average reconstruction of the 4DCT. 4D dose distributions were calculated to evaluate motion effects, caused by the interference of dynamic treatment delivery with respiratory tumor motion and inhomogeneously planned target dose. These calculations were performed on the phase sorted CT series with a respiration-correlated assignment of the treatment plan's monitor units (MU) to the respiration phases of the 4DCT. The 4D dose was accumulated with rigid as well as deformable registrations of the CT series and compared to the original 3D dose distribution. Maximum, minimum and mean doses to ITV and PTV, and maximum or mean doses to organs at risk (OAR), were compared after rigid accumulation. The dose variation in the gross tumor volume (GTV) was compared after deformable registration. Results: Using rigid registrations, variations in the investigated dose parameters between 3D and 4D dose calculations were found to be within -2.1% to 1.4% for all target volumes and within -0.8% to 1.7% in OAR. Using deformable registrations, dose differences in the GTV were below 3.8% for dose accumulation of lung and adrenal gland patients. For liver patients the used deformable registrations were not considered to be robust enough. It was also shown that a major part of the dose differences originates from the Hounsfield unit differences between 3D and 4D calculations, regardless of the interplay effect. Conclusion: The evaluated motion effects during VMAT treatments resulted in negligible dose variability. Therefore, the approximation of calculating the dose on the average reconstruction of the 4DCT (3D dose calculation), instead of calculating on the respiration-correlated phase CTs (4D dose calculation) with assignment of the corresponding MUs, gives acceptable results.
引用
收藏
页码:45 / 53
页数:9
相关论文
共 50 条
  • [31] Fetal neurobehavior assessed by three-dimensional/four-dimensional sonography
    Kurjak, Asim
    Luetic, Ana Tikvica
    ZDRAVNISKI VESTNIK-SLOVENIAN MEDICAL JOURNAL, 2010, 79 (11): : 790 - 799
  • [32] Three-Dimensional Conformal Radiation Therapy, Volumetric Intensity Modulated Arc Therapy, and Helical Tomotherapy for Pediatric Craniospinal Irradiation: Dosimetric Analysis of Integral Dose
    Achari, R. Basu
    Gowardhanan, D.
    Arun, B.
    Dharmendran, P.
    Prasath, S.
    Chatterjee, S.
    Mallick, I.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 87 (02): : S591 - S592
  • [33] VOLUMETRIC MODULATED ARC THERAPY FOR DELIVERY OF PROSTATE RADIOTHERAPY: COMPARISON WITH INTENSITY-MODULATED RADIOTHERAPY AND THREE-DIMENSIONAL CONFORMAL RADIOTHERAPY
    Palma, David
    Vollans, Emily
    James, Kerry
    Nakano, Sandy
    Moiseenko, Vitali
    Shaffer, Richard
    Mckenzie, Michael
    Morris, James
    Otto, Karl
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 72 (04): : 996 - 1001
  • [34] VOLUMETRIC MODULATED ARC THERAPY FOR PEDIATRIC CRANIOSPINAL IRRADIATION: A COMPARSION WITH HELICAL TOMOTHERAPY AND THREE-DIMENSIONAL CONFORMAL RADIOTHERAPY
    Patel, S.
    Warkentin, H.
    Jacques, A.
    Syme, A.
    Warkentin, B.
    Fallone, G.
    RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S64 - S64
  • [35] Effect of arc length on skin dose from hypofractionated volumetric modulated arc radiotherapy treatments of the lung and spine
    Rijken, James
    Kairn, Tanya
    Crowe, Scott
    Trapp, Jamie
    MEDICAL DOSIMETRY, 2019, 44 (04) : 309 - 314
  • [36] Comparative Planning Analysis and Integral Dose of Volumetric Modulated Arc Therapy, Helical Tomotherapy, and Three-dimensional Conformal Craniospinal Irradiation for Pediatric Medulloblastoma
    Patel, Samir
    Drodge, Suzanne
    Jacques, Amanda
    Warkentin, Heather
    Powell, Kent
    Chafe, Susan
    JOURNAL OF MEDICAL IMAGING AND RADIATION SCIENCES, 2015, 46 (02) : 134 - 140
  • [37] Continuous aperture dose calculation and optimization for volumetric modulated arc therapy
    Christiansen, Eric
    Heath, Emily
    Xu, Tong
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (21):
  • [38] Prediction of Three Dimensional Dose Distribution Using a Convolutional Neural Network for Prostate Radiation Volumetric Modulated Arc Therapy
    Yuan, J.
    Kashani, R.
    Zaorsky, N.
    Spratt, D.
    Zheng, Y.
    MEDICAL PHYSICS, 2022, 49 (06) : E409 - E409
  • [39] Four-dimensional CT-based evaluation of volumetric modulated arc therapy for abdominal lymph node metastasis from hepatocellular carcinoma
    Zhang, Li
    Xi, Mian
    Deng, Xiao-Wu
    Li, Qiao-Qiao
    Huang, Xiao-Yan
    Liu, Meng-Zhong
    JOURNAL OF RADIATION RESEARCH, 2012, 53 (05) : 769 - 776
  • [40] The three-dimensional Kepler problem and the four-dimensional isotropic harmonic oscillator
    del Castillo, GFT
    Ortega, AL