Dosimetric evaluation of a novel polymer gel dosimeter for proton therapy

被引:53
|
作者
Zeidan, O. A. [1 ]
Sriprisan, S. I. [1 ]
Lopatiuk-Tirpak, O. [1 ]
Kupelian, P. A. [1 ]
Meeks, S. L. [1 ]
Hsi, W. C. [2 ]
Li, Z. [2 ]
Palta, J. R. [2 ]
Maryanski, M. J. [3 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Orlando, FL 32806 USA
[2] Univ Florida, Proton Therapy Inst, Jacksonville, FL 32206 USA
[3] MGS Res Inc, Madison, CT 06443 USA
关键词
dosimeters; polymer gels; radiation therapy; RADIATION-DOSE DISTRIBUTIONS; BEAM; VERIFICATION; PERFORMANCE; SCANNER;
D O I
10.1118/1.3388869
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Methods: Using passive scattered modulated and nonmodulated proton beams, the dose response of the gel was assessed. A next-generation optical CT scanner is used as the readout mechanism of the radiation-induced absorbance in the gel medium. Comparison of relative dose profiles in the gel to ion chamber profiles in water is performed. A simple and easily reproducible calibration protocol is established for routine gel batch calibrations. Relative stopping power ratio measurement of the gel medium was performed to ensure accurate water-equivalent depth dose scaling. Measured dose distributions in the gel were compared to treatment planning system for benchmark irradiations and quality of agreement is assessed using clinically relevant gamma index criteria. Results: The dosimetric response of the gel was mapped up to 600 cGy using an electron-based calibration technique. Excellent dosimetric agreement is observed between ion chamber data and gel. The most notable result of this work is the fact that this gel has no observed dose quenching in the Bragg peak region. Quantitative dose distribution comparisons to treatment planning system calculations show that most (>97%) of the gel dose maps pass the 3%/3 mm gamma criterion. Conclusions: This study shows that the new proton-sensitive gel dosimeter is capable of reproducing ion chamber dose data for modulated and nonmodulated Bragg peak beams with different clinical beam energies. The findings suggest that the gel dosimeter can be used as QA tool for millimeter range verification of proton beam deliveries in the dosimeter medium.
引用
收藏
页码:2145 / 2152
页数:8
相关论文
共 50 条
  • [1] Dosimetric Evaluation of Polymer Gel Dosimeter Using Saccharide in Clinical Radiation Therapy System
    Cho, Y.
    Lee, D.
    Lee, Y.
    Park, J.
    Kim, K.
    Jung, H.
    Ji, Y.
    Chang, U.
    Kwon, S.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 90 : S936 - S936
  • [2] Dosimetric Accuracy of the CrystalBalla,, c: New Reusable Radiochromic Polymer Gel Dosimeter for Patient QA in Proton Therapy
    Avery, S.
    Kraus, J.
    Lin, L.
    Kassaee, A.
    Maryanski, M.
    MEDICAL PHYSICS, 2015, 42 (06) : 3581 - 3581
  • [3] The Basic Dosimetric Properties of NIPAM Polymer Gel Dosimeter
    Pak, F.
    Farajollahi, A.
    Miabi, Z.
    MEDICAL PHYSICS, 2012, 39 (06) : 3737 - 3737
  • [4] Dosimetric evaluation of a glass dosimeter for proton beam measurements
    Rah, Jeong-Eun
    Oh, Do Hoon
    Shin, Dongho
    Kim, Dae-Hyun
    Ji, Young Hoon
    Kim, Jong Won
    Park, Sung Yong
    APPLIED RADIATION AND ISOTOPES, 2012, 70 (08) : 1616 - 1623
  • [5] Dosimetric Evaluation of ArcCHECK and 3DVH System Using Customized Polymer Gel Dosimeter
    Ono, K.
    Fujimoto, S.
    Hayashi, S.
    Miyazawa, M.
    Akagi, Y.
    Hirokawa, Y.
    MEDICAL PHYSICS, 2015, 42 (06) : 3406 - 3406
  • [6] Dosimetric characteristics of PASSAG as a new polymer gel dosimeter with negligible toxicity
    Farhood, Bagher
    Abtahi, Seyed Mohammad Mandi
    Geraily, Ghazale
    Ghorbani, Mehdi
    Mandavi, Seied Rabi
    Zahmatkesh, Mohammad Hasan
    RADIATION PHYSICS AND CHEMISTRY, 2018, 147 : 91 - 100
  • [7] Dosimetric evaluation of PASSAG-U polymer gel dosimeter: Dependence of dose rate and photon energy
    Farhood, Bagher
    Mohammadi, Kamal A. S. L.
    Sarvizadeh, Mostafa
    Aliasgharzadeh, Akbar
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2020, 28 (04) : 641 - 658
  • [8] Dosimetric Evaluation of the Respiratory Interplay Effect During VMAT Delivery Using IPAGAT Polymer Gel Dosimeter
    Ono, K.
    Fujimoto, S.
    Hayashi, S.
    Hioki, K.
    Miyazawa, M.
    Akagi, Y.
    Hirokawa, Y.
    MEDICAL PHYSICS, 2016, 43 (06) : 3634 - 3635
  • [9] Dosimetric Validation of the Proton- Sensitive BANG® 3 PRO-2 Polymer Gel Dosimeter Irradiated by Single Field Therapeutic Proton Beams
    Tirpak, L.
    Hsi, W.
    Meeks, S.
    Maryanski, M.
    Kupelian, P.
    Zeidan, O.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [10] Dosimetric characteristics of a single use MOSFET dosimeter for in vivo dosimetry in proton therapy
    Cheng, Chee-Wai
    Wolanski, Mark
    Zhao, Q.
    Fanelli, Leia
    Gautam, Archana
    Pack, Daniel
    Das, Indra J.
    MEDICAL PHYSICS, 2010, 37 (08) : 4266 - 4273