Left Ventricular Assist Device Flow Pattern Analysis Using a Novel Model Incorporating Left Ventricular Pulsatility

被引:14
|
作者
Grinstein, Jonathan [1 ]
Torii, Ryo [2 ]
Bourantas, Christos V. [3 ]
Garcia-Garcia, Hector M. [4 ]
机构
[1] Univ Chicago, Dept Med, Cardiol Sect, Chicago, IL 60637 USA
[2] UCL, Dept Mech Engn, London, England
[3] Barts Heart Ctr, Dept Cardiol, London, England
[4] MedStar Cardiovasc Res Network, Washington, DC USA
关键词
computational fluid dynamics; LVAD flow; Programable flow algorithms; shear stress; shear rate; COMPUTATIONAL FLUID-DYNAMICS; HEARTMATE II; BLOOD-FLOW; SUPPORT; THROMBOSIS; STRESS; SYSTEM;
D O I
10.1097/MAT.0000000000001341
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Our current understanding of flow through the circuit of left ventricular assist device (LVAD), left ventricle and ascending aorta remains incompletely understood. Computational fluid dynamics, which allow for analysis of flow in the cardiovascular system, have been used for this purpose, although current simulation models have failed to fully incorporate the interplay between the pulsatile left ventricle and continuous-flow generated by the LVAD. Flow-through the LVAD is dependent on the interaction between device and patient-specific factors with suboptimal flow patterns evoking increased risk of LVAD-related complications. Computational fluid dynamics can be used to analyze how different pump and patient factors affect flow patterns in the left ventricle and the aorta. Computational fluid dynamics simulations were carried out on a patient with a HeartMate II. Simulations were also conducted for theoretical scenarios substituting HeartWare HVAD, HeartMate 3 (HM3) in continuous mode and HM3 with Artificial Pulse. An anatomical model of the patient was reconstructed from computed tomography (CT) images, and the LVAD outflow was used as the inflow boundary condition. The LVAD outflow was calculated separately using a lumped-parameter-model of the systemic circulation, which was calibrated to the patient based on the patient-specific ventricular volume change reconstructed from 4 dimensional computed tomography and pulmonary capillary wedge pressure tracings. The LVADs were implemented in the lumped-parameter-model via published pressure head versus flow (H-Q) curves. To quantify the flushing effect, virtual contrast agent was released in the ascending aorta and its flushing over the cycles was quantified. Shear stress acting on the aortic endothelium and shear rate in the bloodstream were also quantified as indicators of normal/abnormal blood flow, especially the latter being a biomarker of platelet activation and hemolysis. LVAD speeds for the HVAD and HM3 were selected to match flow rates for the patient's HMII (9,000 RPM for HMII, 5,500 RPM for HM3, and 2,200 RPM for HVAD), the cardiac outputs were 5.81 L/min, 5.83 L/min, and 5.92 L/min, respectively. The velocity of blood flow in the outflow cannula was higher in the HVAD than in the two HeartMate pumps with a cycle average (range) of 0.92 m/s (0.78-1.19 m/s), 0.91 m/s (0.86-1.00 m/s), and 1.74 m/s (1.40-2.24 m/s) for HMII, HM3, and HVAD, respectively. Artificial pulse increased the peak flow rate to 9.84 L/min for the HM3 but the overall cardiac output was 5.96 L/min, which was similar to the continuous mode. Artificial pulse markedly decreased blood stagnation in the ascending aorta; after six cardiac cycles, 48% of the blood was flushed out from the ascending aorta under the continuous operation mode while 60% was flushed under artificial pulse. Shear stress and shear rate in the aortic arch were higher with the HVAD compared to the HMII and HM3, respectively (shear stress: 1.76 vs. 1.33 vs. 1.33 Pa, shear rate: 136 vs. 91.5 vs. 89.4 s(-1)). Pump-specific factors such as LVAD type and programmed flow algorithms lead to unique flow patterns which influence blood stagnation, shear stress, and platelet activation. The pump-patient interaction can be studied using a novel computational fluid dynamics model to better understand and potentially mitigate the risk of downstream LVAD complications.
引用
收藏
页码:724 / 732
页数:9
相关论文
共 50 条
  • [41] Endocarditis in left ventricular assist device
    Thyagarajan, Braghadheeswar
    Kumar, Monisha Priyadarshini
    Sikachi, Rutuja R.
    Agrawal, Abhinav
    INTRACTABLE & RARE DISEASES RESEARCH, 2016, 5 (03) : 177 - 184
  • [42] Design, analysis and validation of left ventricular assist device
    Antony, Martin
    Tom, Melvin
    Varghese, Jacob
    Joseph, Sebin
    Polachan, Sheen
    MATERIALS TODAY-PROCEEDINGS, 2021, 47 : 5406 - 5409
  • [43] Control of Left Ventricular Load and Pulsatility by a Continuous-flow Left Ventricular Assist Device With a Native Heart Load Control System in Chronic Awake Phase
    Date, Kazuma
    Nishimura, Takashi
    Kishimoto, Satoru
    Takewa, Yoshiaki
    Ono, Minoru
    Tatsumi, Eisuke
    CIRCULATION, 2014, 130
  • [44] Device Management and Flow Optimization on Left Ventricular Assist Device Support
    Tchoukina, Inna
    Smallfield, Melissa C.
    Shah, Keyur B.
    CRITICAL CARE CLINICS, 2018, 34 (03) : 453 - +
  • [45] Temporary extracorporeal left ventricular assist device support for implantable left ventricular assist device replacement cases
    Schechter, Matthew A.
    Patel, Chetan B.
    Rogers, Joseph G.
    Milano, Carmelo A.
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2014, 147 (04): : E46 - E48
  • [46] Medical Management of Patients With a Left Ventricular Assist Device for the Non-Left Ventricular Assist Device Specialist
    DeVore, Adam D.
    Patel, Priyesh A.
    Patel, Chetan B.
    JACC-HEART FAILURE, 2017, 5 (09) : 621 - 631
  • [47] Extending the perspective on left ventricular assist device pump thrombosis to left ventricular assist device system thrombosis
    Kormos, Robert L.
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2018, 155 (01): : E37 - E38
  • [48] Initial Experience of Conversion of Toyobo Paracorporeal Left Ventricular Assist Device to DuraHeart Left Ventricular Assist Device
    Yoshioka, Daisuke
    Sakaguchi, Taichi
    Saito, Shunsuke
    Miyagawa, Shigeru
    Nishi, Hiroyuki
    Yoshikawa, Yasushi
    Fukushima, Satsuki
    Ueno, Takayoshi
    Kuratani, Toru
    Sawa, Yoshiki
    CIRCULATION JOURNAL, 2012, 76 (02) : 372 - 376
  • [49] Development of a direct mechanical left ventricular assist device for left ventricular failure
    Hotei, Hiroshi
    Koura, Yoshihiko
    Orihashi, Kazumasa
    Sueda, Taijiro
    Fukunaga, Sintaro
    Matsuura, Yuichiro
    1997, (21)
  • [50] A Novel Toroidal-Flow Left Ventricular Assist Device Minimizes Blood Trauma: Implications of Improved Ventricular Assist Device Hemocompatibility
    Bartoli, Carlo R.
    Hennessy-Strahs, Samson
    Gohean, Jeff
    Villeda, Maryann
    Larson, Erik
    Longoria, Raul
    Kurusz, Mark
    Acker, Michael A.
    Smalling, Richard
    ANNALS OF THORACIC SURGERY, 2019, 107 (06): : 1761 - 1767