Three-Dimensional Magnetic Resonance Imaging Quantification of Glenoid Bone Loss Is Equivalent to 3-Dimensional Computed Tomography Quantification: Cadaveric Study

被引:33
|
作者
Yanke, Adam B. [1 ]
Shin, Jason J. [2 ]
Pearson, Ian [3 ]
Bach, Bernard R., Jr. [1 ]
Romeo, Anthony A. [1 ]
Cole, Brian J. [1 ]
Verma, Nikhil N. [1 ]
机构
[1] Rush Univ, Dept Orthopaed, Sports Med Div, Chicago, IL 60612 USA
[2] Univ Saskatchewan, Royal Univ Hosp, Dept Orthopaed, Saskatoon, SK, Canada
[3] Rush Univ, Med Sch, Chicago, IL 60612 USA
关键词
SHOULDER INSTABILITY; ARTICULAR-CARTILAGE; RADIATION-EXPOSURE; GLENOHUMERAL JOINT; MR ARTHROGRAPHY; BANKART REPAIR; DEFECT; CT; DISLOCATION; RECURRENCE;
D O I
10.1016/j.arthro.2016.08.025
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose: To assess the ability of 3-dimensional (3D) magnetic resonance imaging (MRI, 1.5 and 3 tesla [ T]) to quantify glenoid bone loss in a cadaveric model compared with the current gold standard, 3D computed tomography (CT). Methods: Six cadaveric shoulders were used to create a bone loss model, leaving the surrounding soft tissues intact. The anteroposterior (AP) dimension of the glenoid was measured at the glenoid equator and after soft tissue layer closure the specimen underwent scanning (CT, 1.5-T MRI, and 3-T MRI) with the following methods (0%, 10%, and 25% defect by area). Raw axial data from the scans were segmented using manual mask manipulation for bone and reconstructed using Mimics software to obtain a 3D en face glenoid view. Using calibrated Digital Imaging and Communications in Medicine images, the diameter of the glenoid at the equator and the area of the glenoid defect was measured on all imaging modalities. Results: In specimens with 10% or 25% defects, no difference was detected between imaging modalities when comparing the measured defect size (10% defect P = .27, 25% defect P = .73). All 3 modalities demonstrated a strong correlation with the actual defect size (CT, rho =.97; 1.5-T MRI, rho = .93; 3-T MRI, rho = .92, P <.0001). When looking at the absolute difference between the actual and measured defect area, no significance was noted between imaging modalities (10% defect P = .34, 25% defect P =.47). The error of 3-T 3D MRI increased with increasing defect size (P = .02). Conclusions: Both 1.5- and 3-Tebased 3D MRI reconstructions of glenoid bone loss correlate with measurements from 3D CT scan data and actual defect size in a cadaveric model. Regardless of imaging modality, the error in bone loss measurement tends to increase with increased defect size. Use of 3D MRI in the setting of shoulder instability could obviate the need for CT scans. Clinical Relevance: The goal of our work was to develop a reproducible method of determining glenoid bone loss from 3D MRI data and hence eliminate the need for CT scans in this setting. This will lead to decreased cost of care as well as decreased radiation exposure to patients. The long-term goal is a fully automated system that is as approachable for clinicians as current 3D CT technology.
引用
收藏
页码:709 / 715
页数:7
相关论文
共 50 条
  • [31] Recurrent anterior glenohumeral instability: the quantification of glenoid bone loss using magnetic resonance imaging
    Patrícia Martins e Souza
    Bruno Lobo Brandão
    Eduardo Brown
    Geraldo Motta
    Martim Monteiro
    Edson Marchiori
    [J]. Skeletal Radiology, 2014, 43 : 1085 - 1092
  • [32] Recurrent anterior glenohumeral instability: the quantification of glenoid bone loss using magnetic resonance imaging
    Martins e Souza, Patricia
    Brandao, Bruno Lobo
    Brown, Eduardo
    Motta, Geraldo
    Monteiro, Martim
    Marchiori, Edson
    [J]. SKELETAL RADIOLOGY, 2014, 43 (08) : 1085 - 1092
  • [33] Three-dimensional quantification of myocardial perfusion during regadenoson stress computed tomography
    Mor-Avi, Victor
    Kachenoura, Nadjia
    Maffessanti, Francesco
    Bhave, Nicole M.
    Port, Steven
    Lodato, Joseph A.
    Chandra, Sonal
    Freed, Benjamin H.
    Lang, Roberto M.
    Patel, Amit R.
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2016, 85 (05) : 885 - 892
  • [34] Editorial Commentary: Both 3-Dimensional Magnetic Resonance Imaging and Computed Tomography Are Valuable for Determination of Glenoid and Humeral Bone Loss in Patients With On- and Off-Track Shoulder Instability
    Whalen, Ryan J.
    Provencher, Matthew T.
    [J]. ARTHROSCOPY-THE JOURNAL OF ARTHROSCOPIC AND RELATED SURGERY, 2024, 40 (06): : 1789 - 1792
  • [35] Distal Clavicle Autograft Versus Traditional and Congruent Arc Latarjet Procedures: A Comparison of Surface Area and Glenoid Apposition With 3-Dimensional Computed Tomography and 3-Dimensional Magnetic Resonance Imaging
    Kim, Billy I.
    Hudson, Caroline P.
    Taylor, Dean C.
    Anakwenze, Oke A.
    Dickens, Jonathan F.
    Lau, Brian C.
    [J]. AMERICAN JOURNAL OF SPORTS MEDICINE, 2023, 51 (05): : 1295 - 1302
  • [36] QUANTIFICATION OF THORACIC VOLUMES BY 3-DIMENSIONAL IMAGING
    KRAYER, S
    REHDER, K
    BECK, KC
    CAMERON, PD
    DIDIER, EP
    HOFFMAN, EA
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 1987, 62 (02) : 591 - 598
  • [37] Evaluation of ossification of the posterior longitudinal ligament by three-dimensional computed tomography and magnetic resonance imaging
    Kawaguchi, Yoshiharu
    Urushisaki, Aya
    Seki, Shoji
    Hori, Takeshi
    Asanuma, Yumiko
    Kimura, Tomoatsu
    [J]. SPINE JOURNAL, 2011, 11 (10): : 927 - 932
  • [38] Three-dimensional cerebral vasculature of the CBA mouse brain: A magnetic resonance imaging and micro computed tomography study
    Dorr, A.
    Sled, J. G.
    Kabani, N.
    [J]. NEUROIMAGE, 2007, 35 (04) : 1409 - 1423
  • [40] Quantifying Glenoid Bone Loss in Anterior Shoulder Instability Reliability and Accuracy of 2-Dimensional and 3-Dimensional Computed Tomography Measurement Techniques
    Bois, Aaron J.
    Fening, Stephen D.
    Polster, Josh
    Jones, Morgan H.
    Miniaci, Anthony
    [J]. AMERICAN JOURNAL OF SPORTS MEDICINE, 2012, 40 (11): : 2569 - 2577