Right ventricular assist system feedback flow control parameter for a rotary blood pump

被引:5
|
作者
Yoshikawa, M
Nakata, K
Nonaka, K
Linneweber, J
Kawahito, S
Takano, T
Shulte-Eistrup, S
Maeda, T
Glueck, J
Schima, H
Wolner, E
Nosé, Y
机构
[1] Baylor Coll Med, Houston, TX 77030 USA
[2] Univ Vienna, Vienna, Austria
关键词
right ventricular assist system; implantable right ventricular assist device; Gyro pump; flow control; feedback parameter; exercise test;
D O I
10.1046/j.1525-1594.2000.06579.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
At least 25-30% of patients with a permanent implantable left ventricular assist device (LVAD) experience right ventricular failure; therefore, an implantable biventricular assist system (BiVAS) with small centrifugal pumps is being developed. Many institutions are focusing and developing a control system for a left ventricular assist system (LVAS) with rotary blood pumps. These authors feel that the right ventricular assist system (RVAS) with rotary blood pumps should be developed simultaneously. A literature search indicated no recent reports on the effect of hemodynamics and exercise with this type of nonpulsatile implantable RVAS. In this study, a calf with an implantable right ventricular assist system (RVAS) was subjected to 30 min of exercise on a treadmill at 1.5 mph, resulting in excellent hemodynamics. The input voltage remained unchanged. Hemodynamic recordings were taken every 5 min throughout the testing period, and blood gas analysis was done every 10 min. Oxygen uptake (VO2), oxygen delivery (DO2), and oxygen extraction (O2ER) were calculated and analyzed. Two different pump flows were investigated: Group 1 low assist (<3.5 L/min) and Group 2 high assist (>3.5 L/min). In both groups, the RVAS flow rates were unchanged while the pulmonary artery (PA) flow increased during exercise; also, the heart rate and right atrial pressure (RAP) increased during exercise. There were no significant differences in the 2 groups. The PA flow correlates to the heart rate during exercise. In all of the tests, the VO2 and DO2 increased during exercise. Regarding VO2, no changes were observed during the different flow conditions; however, the DO2 of Group 2 was higher than that of Group 1. Because the implantable RVAS did not have pump flow changes during the test conditions, it was necessary to incorporate a flow control system for the implantable RVAS. During exercise with an implantable RVAS rotary blood pump, incorporating the heart rate and VO2 as feedback parameters is feasible for controlling the flow rate.
引用
收藏
页码:659 / 666
页数:8
相关论文
共 50 条
  • [31] ANALYTICAL DERIVATION AND NUMERICAL SIMULATION OF A PUMP FLOW DERIVED ESTIMATION OF VENTRICULAR RELAXATION IN ROTARY BLOOD PUMP RECIPIENTS
    Moscato, F.
    Naiyanetr, P.
    Wieselthaler, G.
    Schima, H.
    [J]. INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2010, 33 (07): : 425 - 425
  • [32] Evaluation of Left Ventricular Relaxation in Rotary Blood Pump Recipients Using the Pump Flow Waveform: A Simulation Study
    Moscato, Francesco
    Granegger, Marcus
    Naiyanetr, Phornphop
    Wieselthaler, Georg
    Schima, Heinrich
    [J]. ARTIFICIAL ORGANS, 2012, 36 (05) : 470 - 478
  • [33] Computational flow study of the continuous flow ventricular assist device, prototype number 3 blood pump
    Anderson, JB
    Wood, HG
    Allaire, PE
    Bearnson, G
    Khanwilkar, P
    [J]. ARTIFICIAL ORGANS, 2000, 24 (05) : 377 - 385
  • [34] Non-Invasive Flow Measurement of a Rotary Pump Ventricular Assist Device Using Quantitative Contrast Echocardiography
    Schwarz, Karl Q.
    Parikh, Sachin S.
    Chen, Xucai
    Farrar, David J.
    Steinmetz, Sherry
    Ramamurthi, Sridevi
    Hallinan, William
    Massey, H. Todd
    Chen, Leway
    [J]. JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2010, 23 (03) : 324 - 329
  • [35] Left ventricular assist device therapy with the HeartMate II continuous flow rotary pump as a bridge to heart transplantation
    Pagani, Francis D.
    Miller, Leslie W.
    Russell, Stuart D.
    John, Ranjit
    Boyle, Andrew J.
    Aaronson, Keith D.
    Conte, John V.
    Naka, Yoshifumi
    Mancini, Donna
    Delgado, Reynolds M.
    MacGillivray, Thomas E.
    Farrar, David J.
    Frazier, O. H.
    [J]. CIRCULATION, 2007, 116 (16) : 540 - 541
  • [36] Flow through an outlet cannula of a rotary ventricular assist device
    Manning, KB
    Miller, GE
    [J]. ARTIFICIAL ORGANS, 2002, 26 (08) : 714 - 723
  • [37] In Vivo Evaluation of Active and Passive Physiological Control Systems for Rotary Left and Right Ventricular Assist Devices
    Gregory, Shaun D.
    Stevens, Michael C.
    Pauls, Jo P.
    Schummy, Emma
    Diab, Sara
    Thomson, Bruce
    Anderson, Ben
    Tansley, Geoff
    Salamonsen, Robert
    Fraser, John F.
    Timms, Daniel
    [J]. ARTIFICIAL ORGANS, 2016, 40 (09) : 894 - 903
  • [38] Parameter optimization of rotary valve in the flow control type ECHPS system
    Academy of Armored Forces Engineering, Beijing 100072, China
    不详
    [J]. Nongye Jixie Xuebao/Transactions of the Chinese Society of Agricultural Machinery, 2008, 39 (11): : 26 - 29
  • [39] Automatic control of assist flow in totally implantable left ventricular assist device using variable pump rate
    Choi, J
    Choi, J
    Jo, Y
    Kim, W
    Min, B
    [J]. PROCEEDINGS OF THE 20TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOL 20, PTS 1-6: BIOMEDICAL ENGINEERING TOWARDS THE YEAR 2000 AND BEYOND, 1998, 20 : 462 - 465
  • [40] Impact of the Rotational Pump Speed of the Implantable Left Ventricular Assist Device on Right Ventricular Pump Function
    Sakamoto, Takafumi
    Fujino, Takeo
    Higo, Taiki
    Ide, Tomomi
    Tanoue, Yoshihisa
    Shiose, Akira
    Tsutsui, Hiroyuki
    [J]. JOURNAL OF CARDIAC FAILURE, 2016, 22 (09) : S164 - S164