Evaluation criteria for film based intensity modulated radiation therapy quality assurance

被引:7
|
作者
Anjum, Muhammad Naeem [1 ,2 ]
Parker, William [2 ]
Ruo, Russell [2 ]
Afzal, Muhammad [1 ,3 ]
机构
[1] Islamia Univ Bahawalpur, Dept Phys, Bahawalpur, Pakistan
[2] McGill Univ, Ctr Hlth, Montreal Gen Hosp, Dept Med Phys, Montreal, PQ H3C 1A4, Canada
[3] Univ Dundee, Ninewells Hosp & Med Sch, Dept Med Phys, Dundee DD1 9SY, Scotland
来源
关键词
IMRT; Gafchromic (TM) EBT; gamma-Histograms; Film dosimetry; EBT FILM; RADIOTHERAPY; DOSIMETRY; CARCINOMA; CANCER; IMRT; IMPLEMENTATION; VERIFICATION; SENSITIVITY; DELIVERY;
D O I
10.1016/j.ejmp.2009.06.002
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this study was to use different gamma histogram criteria for the comparison of planned dose with irradiated dose distribution and find that what percent of pixels passing a certain criteria imitate a good quality plan. The dose was calculated for 156 patients by inverse planning optimization using the Corvus treatment planning system. Gaf-3 chromic films in combination with 2571 0.6 cm(3) Farmer type ionization chamber and Farmer 2570/1 electrometer from NE Technology were used to measure the delivered dose in solid water phantom. All the measurements were performed on Varian CL21EX linear accelerator (Varian Medical. Systems, Palo Alto, CA) fitted with a Millennium 120 leaf collimator. In this study the mean value of the percent of passing pixels within the region of interest under the criterion of 3% DD and 3 mm DTA is 90.2 +/- 7.1% for head and neck cases and 92.2 +/- 5.8% for non-head and neck cases. If we choose the criteria of 3% DD and 3 mm DTA then 96.3% head and neck plans have the percent of passing pixels >= 75% and 95.1% non-head and neck plans have the percent of passing pixels >= 80%. It is evident from the results of this study that the criterion of 5% DD and 3 mm DTA with the percent of passing pixels >= 90 for non-head and neck cases while the percent of passing pixels >= 85 for head and neck cases endorse that a plan is good. The results of this study may be useful for other institutions which use verification software and EBT films for patient specific IMRT QA. (C) 2009 Associazione Itatiana di Fisica Medica. Published by Elsevier Ltd. All. rights reserved.
引用
收藏
页码:38 / 43
页数:6
相关论文
共 50 条
  • [21] Statistical process control analysis for patient quality assurance of intensity modulated radiation therapy
    Rena Lee
    Kyubo Kim
    Samju Cho
    Sangwook Lim
    Suk Lee
    Jang Bo Shim
    Hyun Do Huh
    Sang Hoon Lee
    Sohyun Ahn
    [J]. Journal of the Korean Physical Society, 2017, 71 : 717 - 721
  • [22] Investigating the Importance of Phantom Geometry in Intensity-Modulated Radiation Therapy Quality Assurance
    Hamilton, B.
    Langner, U.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [23] Intensity-modulated radiation therapy -: Quality assurance with the mutual information index (MI)
    Brendemuehl, Astrid
    Ostrowitzki, Stefan
    Altenburger, Uwe
    Haller, J. Christoph
    Ruhlmann, Juegen
    Buzug, Thorsten M.
    [J]. Advances in Medical Engineering, 2007, 114 : 205 - 210
  • [24] Comparison of Error Range of Delivery Quality Assurance by Type of Intensity Modulated Radiation Therapy
    Kim, Jeong Ho
    Yoo, Se Jong
    [J]. JOURNAL OF MAGNETICS, 2018, 23 (04) : 587 - 595
  • [25] Statistical process control analysis for patient quality assurance of intensity modulated radiation therapy
    Lee, Rena
    Kim, Kyubo
    Cho, Samju
    Lim, Sangwook
    Lee, Suk
    Shim, Jang Bo
    Huh, Hyun Do
    Lee, Sang Hoon
    Ahn, Sohyun
    [J]. JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2017, 71 (10) : 717 - 721
  • [26] The impacts of dose rate in sliding window intensity modulated radiation therapy quality assurance
    Gul, O., V
    Inan, G.
    [J]. INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2021, 19 (03): : 625 - 632
  • [27] Film dosimetry for intensity modulated radiation therapy
    Benites-Rengifo, J
    Martínez-Dávalos, A
    Celis, M
    Lárraga, J
    [J]. MEDICAL PHYSICS, 2004, 724 : 244 - 248
  • [28] Optimization and quality assurance of an image-guided radiation therapy system for intensity-modulated radiation therapy radiotherapy
    Tsai, Jen-San
    Micaily, Bizhan
    Miyamoto, Curtis
    [J]. MEDICAL DOSIMETRY, 2012, 37 (03) : 321 - 333
  • [29] Dosimetric verification and quality assurance for intensity-modulated radiation therapy using Gafchromic (R) EBT3 film
    Iqbal, Khalid
    Iqbal, Muhammad Mazhar
    Akram, Muhammad
    Altaf, Saima
    Buzdar, Saeed Ahmad
    [J]. JOURNAL OF RADIOTHERAPY IN PRACTICE, 2018, 17 (01) : 85 - 95
  • [30] Design and implementation of an anthropomorphic quality assurance phantom for intensity-modulated radiation therapy for the radiation therapy oncology group
    Molineu, A
    Followill, DS
    Balter, PA
    Hanson, WF
    Gillin, MT
    Huq, MS
    Eisbruch, A
    Ibbott, GS
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 63 (02): : 577 - 583