Closed-loop glucose control in critically ill patients using continuous glucose monitoring system (CGMS) in real time

被引:81
|
作者
Chee, F [1 ]
Fernando, T
van Heerden, PV
机构
[1] Univ Western Australia, Sch Elect Elect & Comp Engn, Crawley, WA 6009, Australia
[2] Sir Charles Gairdner Hosp, Intens Care Unit, Nedlands, WA 6009, Australia
关键词
artificial pancreas; closed-loop glucose control; intravenous insulin infusion; sliding scale; subcutaneous blood glucose monitoring;
D O I
10.1109/TITB.2003.808509
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
A study was conducted to determine if continuous subcutaneous glucose monitoring (from MiniMed CGMS) could be used in real-time to control blood sugar level (BSL) in patients with critical illness. A closed-loop control system was constructed to use CGMS in a real-time manner, coupled with. a proportional integral (PI) control algorithm based on a sliding scale approach, for automatic intravenous infusion of insulin to patients. A total of five subjects with high BSL (> 10 mmol/L) participated in formal studies of the closed-loop control system. Subjects were recruited from critically ill patients in the intensive care unit (ICU) after informed consent was obtained. Error grid analysis showed that 64.6% of the BSL readings as determined in real time using CGMS sensor, when compared to conventional BSL measurements on blood drawn from an arterial line, was clinically accurate (i.e., < 20% deviation from glucometer value). In the five patients who underwent closed-loop control, the controller managed to control only one patient's glycaemia without any manual intervention. Manual intervention was required due to the real-time sensor reading deviating more than 20% from the glucometer value, and also as a safety mechanism. Test on equality of mean and variance for BSL attained prior to, during, and post trial showed that the controller's performance was comparable to manual control. We conclude that the automatic sliding scale approach of closed-loop BSL control is feasible in patients in intensive care. More work is needed in the refinement of the algorithm and the improvement of real-time sensor accuracy.
引用
收藏
页码:43 / 53
页数:11
相关论文
共 50 条
  • [31] Glucose control during early insulinization with glargine as assessed by the continuous glucose monitoring system (CGMS)
    Yale, JF
    Gerstein, H
    Harris, SB
    Issa, M
    Granger, E
    Stewart, J
    Dempsey, E
    DIABETOLOGIA, 2005, 48 : A309 - A309
  • [32] REAL-TIME EATING RECOGNITION USING GOOGLE GLASS TO IMPROVE CLOSED-LOOP GLUCOSE CONTROL
    Kleinberg, S.
    Merck, C.
    Rahman, S. A.
    Huang, Y.
    DIABETES TECHNOLOGY & THERAPEUTICS, 2015, 17 : A45 - A45
  • [33] ASSESSMENT OF THE ACCURACY OF REAL-TIME CONTINUOUS GLUCOSE MONITORING AND ITS CORRELATED FACTORS IN CRITICALLY ILL PATIENTS
    MZ Lu
    Y Kang
    Intensive Care Medicine Experimental, 3 (Suppl 1)
  • [34] Development of a real-time closed-loop dual wavelength optical polarimeter for glucose monitoring
    Malik, Bilal H.
    Cote, Gerard L.
    OPTICAL DIAGNOSTICS AND SENSING X: TOWARD POINT-OF-CARE DIAGNOSTICS, 2010, 7572
  • [35] Real-time, closed-loop dual-wavelength optical polarimetry for glucose monitoring
    Malik, Bilal H.
    Cote, Gerard L.
    JOURNAL OF BIOMEDICAL OPTICS, 2010, 15 (01)
  • [36] Accuracy of real-time glucose values using continuous glucose monitoring:: the Guardian™ II continuous glucose monitoring system
    Schwartz, SL
    Buchea, O
    Shin, J
    Hoss, U
    Shah, R
    DIABETOLOGIA, 2004, 47 : A315 - A315
  • [37] Accuracy and reliability of a subcutaneous continuous glucose-monitoring system in critically ill patients
    Brunner, Richard
    Kitzberger, Reinhard
    Miehsler, Wolfgang
    Herkner, Harald
    Madl, Christian
    Holzinger, Ulrike
    CRITICAL CARE MEDICINE, 2011, 39 (04) : 659 - 664
  • [38] CLOSED-LOOP CONTROL AND QUALITY OF CONTINUOUS GLUCOSE MONITOR DATA BASED ON PROSPECTIVE CLOSED-LOOP CLINICAL TRIAL
    Huyett, L. M.
    Harvey, R. A.
    Bevier, W.
    Zisser, H.
    Jovanovic, L.
    Dassau, E.
    Doyle, F. J., III
    DIABETES TECHNOLOGY & THERAPEUTICS, 2013, 15 : A11 - A11
  • [39] The Future of Continuous Glucose Monitoring: Closed Loop
    Hovorka, Roman
    CURRENT DIABETES REVIEWS, 2008, 4 (03) : 269 - 279
  • [40] FEASIBILITY OF CONTINUOUS GLUCOSE MONITORING IN CRITICALLY ILL EMERGENCY DEPARTMENT PATIENTS
    Lee, Jae Hyuk
    Kim, Kyuseok
    Jo, You Hwan
    Rhee, Joong Eui
    Lee, Jung Chan
    Kim, Kyung Su
    Kwon, Woon Yong
    Suh, Gil Joon
    Kim, Hee Chan
    Yoon, Ho Il
    Park, Sang Heon
    JOURNAL OF EMERGENCY MEDICINE, 2012, 43 (02): : 251 - 257