Closed-loop glucose control in critically ill patients using continuous glucose monitoring system (CGMS) in real time

被引:81
|
作者
Chee, F [1 ]
Fernando, T
van Heerden, PV
机构
[1] Univ Western Australia, Sch Elect Elect & Comp Engn, Crawley, WA 6009, Australia
[2] Sir Charles Gairdner Hosp, Intens Care Unit, Nedlands, WA 6009, Australia
关键词
artificial pancreas; closed-loop glucose control; intravenous insulin infusion; sliding scale; subcutaneous blood glucose monitoring;
D O I
10.1109/TITB.2003.808509
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
A study was conducted to determine if continuous subcutaneous glucose monitoring (from MiniMed CGMS) could be used in real-time to control blood sugar level (BSL) in patients with critical illness. A closed-loop control system was constructed to use CGMS in a real-time manner, coupled with. a proportional integral (PI) control algorithm based on a sliding scale approach, for automatic intravenous infusion of insulin to patients. A total of five subjects with high BSL (> 10 mmol/L) participated in formal studies of the closed-loop control system. Subjects were recruited from critically ill patients in the intensive care unit (ICU) after informed consent was obtained. Error grid analysis showed that 64.6% of the BSL readings as determined in real time using CGMS sensor, when compared to conventional BSL measurements on blood drawn from an arterial line, was clinically accurate (i.e., < 20% deviation from glucometer value). In the five patients who underwent closed-loop control, the controller managed to control only one patient's glycaemia without any manual intervention. Manual intervention was required due to the real-time sensor reading deviating more than 20% from the glucometer value, and also as a safety mechanism. Test on equality of mean and variance for BSL attained prior to, during, and post trial showed that the controller's performance was comparable to manual control. We conclude that the automatic sliding scale approach of closed-loop BSL control is feasible in patients in intensive care. More work is needed in the refinement of the algorithm and the improvement of real-time sensor accuracy.
引用
收藏
页码:43 / 53
页数:11
相关论文
共 50 条
  • [21] Improved metabolic control in diabetic adolescents using the continuous glucose monitoring system (CGMS)
    Schaepelynck-Bélicar, P
    Vague, P
    Simonin, G
    Lassmann-Vague, V
    DIABETES & METABOLISM, 2003, 29 (06) : 608 - 612
  • [22] Evaluation of a continuous glucose monitoring system in critically ill adult patients.
    Newman, P.
    Muscedere, J.
    Heyland, D.
    Drover, J.
    Houlden, R.
    Day, A.
    Su, Y.
    CRITICAL CARE MEDICINE, 2006, 34 (12) : A66 - A66
  • [23] Glycemic control in diabetic CAPD patients assessed by continuous glucose monitoring system (CGMS)
    Marshall, J
    Jennings, P
    Scott, A
    Fluck, RJ
    McIntyre, CW
    KIDNEY INTERNATIONAL, 2003, 64 (04) : 1480 - 1486
  • [24] Characteristics of glucose metabolism indexes and continuous glucose monitoring system (CGMS) in patients with insulinoma
    Weijun Gu
    Yixin Liu
    Hongyan Liu
    Guoqing Yang
    Qinghua Guo
    Jin Du
    Nan Jin
    Li Zang
    Zhaohui Lv
    Jianming Ba
    Yiming Mu
    Jingtao Dou
    Diabetology & Metabolic Syndrome, 9
  • [25] What is a normal glucose concentration using the minimed continuous glucose monitoring system (CGMS)®?
    Ford, T
    Parrish, D
    O'Grady, D
    Lane, J
    Mack-Shipman, L
    Colling, C
    Larsen, J
    DIABETES, 2002, 51 : A11 - A11
  • [26] Use of continuous glucose monitoring and hybrid closed-loop therapy in pregnancy
    Benhalima, Katrien
    Yamamoto, Jennifer M.
    DIABETES OBESITY & METABOLISM, 2024,
  • [27] Accuracy and reliability of continuous glucose monitoring in critically ill patients
    Holzinger, Ulrike
    Warszawska, Joanna
    Kitzberger, Reinhaid
    Fuhrmann, Valentin
    Schenk, Peter
    Herkner, Harald
    Madl, Christian
    CRITICAL CARE MEDICINE, 2007, 35 (12) : A137 - A137
  • [28] Accuracy and limitations of continuous glucose monitoring using spectroscopy in critically ill patients
    van Hooijdonk, Roosmarijn T. M.
    Winters, Tineke
    Fischer, Johan C.
    van Dongen-Lases, Edmee C.
    Krinsley, James S.
    Preiser, Jean-Charles
    Schultz, Marcus J.
    ANNALS OF INTENSIVE CARE, 2014, 4 : 1 - 8
  • [29] Accuracy and limitations of continuous glucose monitoring using spectroscopy in critically ill patients
    Roosmarijn TM van Hooijdonk
    Tineke Winters
    Johan C Fischer
    Edmée C van Dongen-Lases
    James S Krinsley
    Jean-Charles Preiser
    Marcus J Schultz
    Annals of Intensive Care, 4
  • [30] Feasibilitv of continuous glucose monitoring in the critically ill
    Borja, E
    Schein, R
    Quartin, A
    Kett, D
    Cely, CM
    Fridman, C
    Meneghini, LF
    DIABETES, 2005, 54 : A491 - A491