A graphite calorimeter for absolute measurements of absorbed dose to water: application in medium-energy x-ray filtered beams

被引:12
|
作者
Pinto, M. [1 ]
Pimpinella, M. [1 ]
Quini, M. [1 ]
D'Arienzo, M. [1 ]
Astefanoaei, I. [2 ]
Loreti, S. [3 ]
Guerra, A. S. [1 ]
机构
[1] Ctr Ric ENEA Casaccia, Ist Nazl Metrol Radiaz Ionizzanti, Via Anguillarese 301, I-00123 Rome, Italy
[2] Al I Cuza Iasi Univ, Fac Phys, Carpath Ctr, 1 Carol I Blvd, RO-700506 Iasi, Romania
[3] Ctr Ric ENEA Frascati, Dipartimento Fus & Tecnol Sicurezza Nucl, Lab Tecnol Nucl, Via E Fermi 45, I-00044 Frascati, Italy
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2016年 / 61卷 / 04期
关键词
graphite calorimetry; calorimetry; medium-energy x-rays; reference dosimetry; primary standard; x-ray dosimetry; HEAT DEFECT; IONIZATION CHAMBERS; LNE-LNHB; DOSIMETRY; KV; UNCERTAINTIES; ABSORPTION; RADIATION; STANDARDS;
D O I
10.1088/0031-9155/61/4/1738
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The Italian National Institute of Ionizing Radiation Metrology (ENEA-INMRI) has designed and built a graphite calorimeter that, in a water phantom, has allowed the determination of the absorbed dose to water in medium-energy x-rays with generating voltages from 180 to 250 kV. The new standard is a miniaturized three-bodies calorimeter, with a disc-shaped core of 21 mm diameter and 2 mm thickness weighing 1.134 g, sealed in a PMMA waterproof envelope with air-evacuated gaps. The measured absorbed dose to graphite is converted into absorbed dose to water by means of an energy-dependent conversion factor obtained from Monte Carlo simulations. Heat-transfer correction factors were determined by FEM calculations. At a source-to-detector distance of 100 cm, a depth in water of 2 g cm(-2), and at a dose rate of about 0.15 Gy min(-1), results of calorimetric measurements of absorbed dose to water, D-w, were compared to experimental determinations, D-wK, obtained via an ionization chamber calibrated in terms of air kerma, according to established dosimetry protocols. The combined standard uncertainty of D-w and D-wK were estimated as 1.9% and 1.7%, respectively. The two absorbed dose to water determinations were in agreement within 1%, well below the stated measurement uncertainties. Advancements are in progress to extend the measurement capability of the new in-water-phantom graphite calorimeter to other filtered medium-energy x-ray qualities and to reduce the D-w uncertainty to around 1%. The new calorimeter represents the first implementation of in-water-phantom graphite calorimetry in the kilovoltage range and, allowing independent determinations of D-w, it will contribute to establish a robust system of absorbed dose to water primary standards for medium-energy x-ray beams.
引用
收藏
页码:1738 / 1764
页数:27
相关论文
共 50 条
  • [21] DETERMINATION OF ABSORBED DOSE FROM X-RAY EXPOSURE MEASUREMENTS
    WORSNOP, RR
    PHYSICS IN MEDICINE AND BIOLOGY, 1965, 10 (02): : 289 - &
  • [22] Study of the formalism used to determine the absorbed dose for low-energy x-ray beams
    Chica, U.
    Anguiano, M.
    Lallena, A. M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (23): : 6963 - 6977
  • [23] New Ionization Chamber Dosimetry of Absorbed Dose to Water in Diagnostic KV X-Ray Beams
    Araki, F.
    Ohno, T.
    MEDICAL PHYSICS, 2016, 43 (06) : 3717 - 3717
  • [24] Monte Carlo studies on dose conversion factors from graphite to water for high energy X-ray beams
    Kim, Y. H.
    Yi, C. Y.
    Kim, I. J.
    Kim, B. C.
    Kim, J. H.
    Seong, Y. M.
    Kim, H. M.
    RADIATION PHYSICS AND CHEMISTRY, 2020, 171
  • [25] Thimble ionization chambers in medium-energy x-ray beams and the role of constructive details of the central electrode: Monte Carlo simulations and measurements
    Ubrich, F.
    Wulff, J.
    Kranzer, R.
    Zink, K.
    PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (18): : 4893 - 4906
  • [26] VARIATIONS IN SEMICONDUCTOR-DEVICE RESPONSE IN A MEDIUM-ENERGY X-RAY DOSE-ENHANCING ENVIRONMENT
    BEUTLER, DE
    FLEETWOOD, DM
    BEEZHOLD, W
    KNOTT, D
    LORENCE, LJ
    DRAPER, BL
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1987, 34 (06) : 1544 - 1550
  • [27] Absorbed dose measurements of a handheld 50 kVP X-ray source in water with thermoluminescence dosemeters
    Soares, Christopher
    Drupieski, Chris
    Wingert, Brian
    Pritchett, Garey
    Pagonis, Vasilis
    O'Brien, Michelle
    Sliski, Alan
    Bilski, Pawel
    Olko, Pawel
    RADIATION PROTECTION DOSIMETRY, 2006, 120 (1-4) : 78 - 82
  • [28] Quality indexes based on water measurements for low and medium energy x-ray beams: A theoretical study with PENELOPE
    Chica, U.
    Anguiano, M.
    Vilches, M.
    Lallena, A. M.
    MEDICAL PHYSICS, 2014, 41 (01)
  • [29] An alternative beam quality index for medium-energy x-ray dosimetry
    Rosser, KE
    PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (03): : 587 - 598
  • [30] Operating a graphite calorimeter in quasi-isothermal mode under high-energy x-ray beams
    Kim, In Jung
    Kim, Yun Ho
    Park, Jong In
    Kim, Byoung-Chul
    Yi, Chul-Young
    PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (23):