Simplified clinical quality assurance for helical tomotherapy

被引:0
|
作者
Grant, W.
McGary, J.
Childress, N.
CravenS, R.
Krumbach, T.
Olivera, G.
机构
[1] Baylor Coll Med, Methodist Hosp, Houston, TX 77030 USA
[2] Tomotherapy Inc, Madison, WI USA
[3] Med Radiat Res Ctr, Madison, WI USA
关键词
D O I
10.1118/1.2761395
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To evaluate a Web‐based Quality Assurance tool for maintaining the clinical integrity of the Tomotherapy machine. Method and Materials: This tool takes advantage of the output of the two distinct ion chamber systems on the Tomotherapy machine. One is the standard sealed ion chamber that resides in the head of a machine near the primary collimators and is used to calculate Monitor Units. The other is a set of pressurized Xenon detectors that are used for imaging. Both systems monitor ionizations independently and are rigidly attached so that their geometry is constant regardless of the orientation of the beam on the gantry which makes possible testing for both static and rotational modes. The collected data can be reviewed via a Web page and a downloadable Report in PDF format is available within seconds after collection of the data. Results: As an example, one performs a 200 second rotational delivery with the couch removed from the beam. The data are analyzed in both an integrated and pulse‐by pulse methodology to determine consistency between the two chamber systems, changes in output and changes in energy. The collected information is evaluated relative to a standard delivery created during machine commissioning. Comparison to standard ion chamber measurements in phantom under static conditions indicate that the analyses are consistent for calibration and more sensitive to energy changes than percentage depth dose measurements made in phantom. Additional information about the machine's performance is identified easily. Conclusion: New technology offers an opportunity to perform standard Quality Assurance tests in a new manner. The simple test described here is more sensitive and performed more quickly than traditional methodology. Other tests are possible. Conflict of Interest: Three authors are employees of Tomotherapy, Inc. © 2007, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:2561 / 2561
页数:1
相关论文
共 50 条
  • [1] Helical tomotherapy quality assurance
    Balog, John
    Soisson, Emilie
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2008, 71 (01): : S113 - S117
  • [2] Helical tomotherapy quality assurance with ArcCHECK
    Chapman, David
    Barnett, Rob
    Yartsev, Slav
    [J]. MEDICAL DOSIMETRY, 2014, 39 (02) : 159 - 162
  • [3] Helical tomotherapy dynamic quality assurance
    Balog, John
    Holmes, Tim
    Vaden, Richard
    [J]. MEDICAL PHYSICS, 2006, 33 (10) : 3939 - 3950
  • [4] Quality Assurance Procedures for Helical Tomotherapy
    Haaren, P.
    Althof, V.
    Kramer, D.
    Westendorp, H.
    Minken, A.
    [J]. MEDICAL PHYSICS, 2008, 35 (06)
  • [5] Quality assurance of a helical tomotherapy machine
    Fenwick, JD
    Tomé, WA
    Jaradat, HA
    Hui, SK
    James, JA
    Balog, JP
    DeSouza, CN
    Lucas, DB
    Olivera, GH
    Mackie, TR
    Paliwal, BR
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (13): : 2933 - 2953
  • [6] Mechanical and dosimetric quality assurance for helical tomotherapy
    Ramsey, C. R.
    Gibbons, J.
    Holmes, T.
    Jeong, K.
    Langen, K.
    Low, D.
    Meeks, S.
    Papanikolaou, N.
    Schubert, K.
    Tomsej, M.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 69 (03): : S699 - S699
  • [7] Enhanced efficacy in helical tomotherapy quality assurance
    Goddu, S.
    Mutic, S.
    Chaudhari, S.
    Grigsby, J.
    Low, D.
    [J]. MEDICAL PHYSICS, 2006, 33 (06) : 2248 - 2248
  • [8] Dose reconstruction quality assurance for helical tomotherapy
    Chase, D.
    Ramsey, C.
    Seibert, R.
    [J]. MEDICAL PHYSICS, 2006, 33 (06) : 2256 - 2256
  • [9] Computed radiography for helical tomotherapy quality assurance
    Ramsey, C.
    Harris, C.
    [J]. MEDICAL PHYSICS, 2006, 33 (06) : 2193 - 2193
  • [10] Patient specific quality assurance for helical tomotherapy
    Thomas, S
    MacKenzie, M
    Field, G
    Syme, A
    Fallone, B
    [J]. MEDICAL PHYSICS, 2005, 32 (07) : 2414 - 2414