Adaptation, Commissioning, and Evaluation of a 3D Treatment Planning System for High-Resolution Small-Animal Irradiation

被引:5
|
作者
Jeong, Jeho [1 ]
Chen, Qing [1 ]
Febo, Robert [1 ]
Yang, Jie [1 ]
Hai Pham [1 ]
Xiong, Jian-Ping [1 ]
Zanzonico, Pat B. [1 ]
Deasy, Joseph O. [1 ]
Humm, John L. [1 ]
Mageras, Gig S. [1 ]
机构
[1] Mem Sloan Kettering Canc Ctr, Dept Med Phys, 1275 York Ave, New York, NY 10065 USA
关键词
small-animal irradiator; radiation therapy; commissioning; treatment planning system; 3D dose distribution; CONFORMAL RADIOTHERAPY; MONTE-CARLO; DOSIMETRY; CT; GUIDANCE; CANCER; ENERGY;
D O I
10.1177/1533034615584522
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Although spatially precise systems are now available for small-animal irradiations, there are currently limited software tools available for treatment planning for such irradiations. We report on the adaptation, commissioning, and evaluation of a 3-dimensional treatment planning system for use with a small-animal irradiation system. The 225-kV X-ray beam of the X-RAD 225Cx microirradiator (Precision X-Ray) was commissioned using both ion-chamber and radiochromic film for 10 different collimators ranging in field size from 1 mm in diameter to 40 x 40 mm(2). A clinical 3-dimensional treatment planning system (Metropolis) developed at our institution was adapted to small-animal irradiation by making it compatible with the dimensions of mice and rats, modeling the microirradiator beam orientations and collimators, and incorporating the measured beam data for dose calculation. Dose calculations in Metropolis were verified by comparison with measurements in phantoms. Treatment plans for irradiation of a tumor-bearing mouse were generated with both the Metropolis and the vendor-supplied software. The calculated beam-on times and the plan evaluation tools were compared. The dose rate at the central axis ranges from 74 to 365 cGy/min depending on the collimator size. Doses calculated with Metropolis agreed with phantom measurements within 3% for all collimators. The beam-on times calculated by Metropolis and the vendor-supplied software agreed within 1% at the isocenter. The modified 3-dimensional treatment planning system provides better visualization of the relationship between the X-ray beams and the small-animal anatomy as well as more complete dosimetric information on target tissues and organs at risk. It thereby enhances the potential of image-guided microirradiator systems for evaluation of dose-response relationships and for preclinical experimentation generally.
引用
收藏
页码:460 / 471
页数:12
相关论文
共 50 条
  • [1] High-resolution 3D Bayesian image reconstruction using the microPET small-animal scanner
    Qi, JY
    Leahy, RM
    Cherry, SR
    Chatziioannou, A
    Farquhar, TH
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (04): : 1001 - 1013
  • [2] High-resolution SPECT for small-animal imaging
    QI Yu-Jin (Shanghai Institute of Applied Physics
    [J]. Nuclear Science and Techniques, 2006, (03) : 164 - 169
  • [3] An evaluation of exact and approximate 3-D reconstruction algorithms for a high-resolution, small-animal PET scanner
    Farquhar, TH
    Chatziioannou, A
    Cherry, SR
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1998, 17 (06) : 1073 - 1080
  • [4] Development and characterization of a high-resolution MicroSPECT system for small-animal Imaging
    Qi, YJ
    Tsui, BMW
    Wang, YC
    Yoder, B
    Wojcik, R
    Majewski, S
    Weisenberger, AG
    [J]. Small Animal SPECT Imaging, 2005, : 259 - 266
  • [5] High-Resolution Small-Animal SPECT with Conventional Gamma Cameras
    Schramm, N. U.
    Hoppin, J. W.
    Lackas, C.
    Behe, M.
    Behr, T. M.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2005, 32 : S253 - S254
  • [6] Tomographic small-animal imaging using a high-resolution semiconductor camera
    Kastis, GA
    Wu, MC
    Balzer, SJ
    Wilson, DW
    Furenlid, LR
    Stevenson, G
    Barber, HB
    Barrett, HH
    Woolfenden, JM
    Kelly, P
    Appleby, M
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2002, 49 (01) : 172 - 175
  • [7] Methodology for commissioning a brachytherapy treatment planning system in the era of 3D planning
    Dempsey, Claire
    [J]. AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2010, 33 (04) : 373 - 373
  • [8] Methodology for commissioning a brachytherapy treatment planning system in the era of 3D planning
    Claire Dempsey
    [J]. Australasian Physical & Engineering Sciences in Medicine, 2010, 33 : 341 - 349
  • [9] Commissioning of a 3D MicroCT-Based Small Animal Radiotherapy System
    Rodriguez, M.
    Zhou, H.
    Jogani, R.
    Nelson, G.
    Keall, P.
    Graves, E.
    [J]. MEDICAL PHYSICS, 2009, 36 (06)
  • [10] Animals In Synchrotrons: Overcoming Challenges For High-Resolution, Live, Small-Animal Imaging
    Donnelley, Martin
    Parsons, David
    Morgan, Kaye
    Siu, Karen
    [J]. 6TH INTERNATIONAL CONFERENCE ON MEDICAL APPLICATIONS OF SYNCHROTRON RADIATION, 2010, 1266 : 30 - +