Magnetic Resonance Imaging in Nondependent Pacemaker Patients with Pacemakers and Defibrillators with a Nearly Depleted Battery

被引:6
|
作者
Okamura, Hideo [1 ]
Padmanabhan, Deepak [1 ]
Watson, Robert E., Jr. [2 ]
Dalzell, Connie [1 ]
Acker, Nancy [1 ]
Jondal, Mary [1 ]
Romme, Abby L. [1 ]
Cha, Yong-mei [1 ]
Asirvatham, Samuel J. [1 ]
Felmlee, Joel P. [2 ]
Friedman, Paul A. [1 ]
机构
[1] Mayo Clin, Div Cardiovasc Dis, Rochester, MN USA
[2] Mayo Clin, Div Diagnost Radiol, Rochester, MN USA
来源
关键词
pacemaker; ICD; MRI; ERI; depleted battery;
D O I
10.1111/pace.13042
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BackgroundMagnetic resonance imaging (MRI) in patients with non-MRI-conditional cardiac implantable electronic devices (CIEDs) has been shown to be safe when performed under closely monitored protocols. However, the safety of MRI in patients with devices with a nearly depleted battery has not been reported. MethodsProspective data were collected between January 2008 and May 2015 in patients with non-MRI-conditional CIEDs undergoing clinically indicated MRI under institutional protocol. Patients who were pacemaker dependent were excluded. Patients whose devices were at elective replacement indicator (ERI) at the time of MRI or close to ERI (ERI or replacement for battery depletion within 3 months of scan) were identified through database review and analyzed for clinical events. ResultsMRI scans (n = 569) were performed in 442 patients. Of these, we identified 13 scans performed with a nearly depleted battery in nine patients. All scans with implantable cardioverter defibrillators (ICDs, n = 9) were uneventful. However, two scans with pacemakers close to ERI resulted in a power-on-reset (PoR) event. One scan with a pacemaker close to ERI that was programmed to DOO mode reached ERI during MRI and automatically changed to VVI mode. Additionally, one scan with a pacemaker at ERI did not allow programming. All pacemakers with events were implanted before 2005. ConclusionPatients with pacemakers and ICDs with a nearly depleted battery can safely undergo MRI when patients are not pacemaker dependent. Attention should be paid because old devices can result in PoR or ERI during MRI, which may lead to oversensing and inhibition of pacing.
引用
收藏
页码:476 / 481
页数:6
相关论文
共 50 条
  • [21] Determining the Risks of Magnetic Resonance Imaging at 1.5 Tesla for Patients with Pacemakers and Implantable Cardioverter Defibrillators (The MagnaSafe Registry)
    Russo, Robert J.
    Costal, Heather S.
    Doud, Debra
    Uretsky, Seth
    Wolff, Steven D.
    Martin, Edward T.
    Higgins, Steven
    Tominaga, Gail
    Machado, Christian
    Florin, Todd
    Bloomgarden, Daniel C.
    Birgersdotter-Green, Ulrika
    Ponce, George
    Schaerf, Raymond
    Lampert, Rachel
    Chilson, Donald
    Shah, Dipan J.
    Arshad, Aysha
    Tonkin, Alison
    Anderson, Jeffrey
    CIRCULATION, 2012, 126 (21)
  • [22] Magnetic Resonance Imaging in Patients with Recently Implanted Pacemakers
    Friedman, Hannah L.
    Acker, Nancy
    Dalzell, Connie
    Shen, Win K.
    Asirvatham, Samuel J.
    Cha, Yong-Mei
    Hodge, David
    Felmlee, Joel
    Watson, Robert
    Friedman, Paul A.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2013, 36 (09): : 1090 - 1095
  • [23] Magnetic resonance imaging in patients with cardiac pacemakers - Reply
    Martin, ET
    Coman, JA
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2005, 46 (03) : 562 - 562
  • [24] Magnetic resonance imaging in patients with temporary external pacemakers
    Fyenbo, Daniel Benjamin
    Jensen, Morten Steen Kvistholm
    Kronborg, Mads Brix
    Kristensen, Jens
    Nielsen, Jens Cosedis
    Witt, Christoffer Tobias
    EUROPACE, 2022, 24 (12): : 1960 - 1966
  • [25] Magnetic resonance imaging, pacemakers and implantable cardioverter-defibrillators: current situation and clinical perspective
    Gotte, M. J. W.
    Russel, I. K.
    de Roest, G. J.
    Germans, T.
    Veldkamp, R. F.
    Knaapen, P.
    Allaart, C. P.
    van Rossum, A. C.
    NETHERLANDS HEART JOURNAL, 2010, 18 (01) : 31 - 37
  • [26] Recent innovations in the development of magnetic resonance imaging conditional pacemakers and implantable cardioverter-defibrillators
    Lobodzinski, S. Suave
    CARDIOLOGY JOURNAL, 2012, 19 (01) : 98 - 104
  • [27] A systematic review of the use of magnetic resonance imaging in non-conditional pacemakers and implantable defibrillators
    Murphy, R.
    Kerley, R. K.
    Kearney, P. K.
    EUROPEAN HEART JOURNAL, 2021, 42 : 229 - 229
  • [28] Evaluation of a safety protocol for clinically indicated magnetic resonance imaging of patients with permanent pacemakers and implantable defibrillators at 1.5 tesla
    Nazarjan, S
    Roguin, A
    Zviman, MM
    Lardo, AC
    Dickfeld, TL
    Berger, RD
    Bluemke, DA
    Halperin, HR
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2005, 45 (03) : 14A - 14A
  • [29] Clinical utility and safety of a protocol for magnetic resonance Imaging of patients with permanent pacemakers and implantable cardioverter defibrillators at 1.5 tesia
    Nazarian, S
    Roguin, A
    Zviman, MM
    Lardo, AC
    Dickfeld, TL
    Berger, RD
    Bluemke, DA
    Halperin, HR
    CIRCULATION, 2005, 112 (17) : U575 - U575
  • [30] Magnetic resonance imaging in patients with a pacemaker system designed for the magnetic resonance environment
    Wilkoff, Bruce L.
    Bello, David
    Taborsky, Milos
    Vymazal, Josef
    Kanal, Emanuel
    Heuer, Hubertus
    Hecking, Katrin
    Ben Johnson, W.
    Young, William
    Ramza, Brian
    Akhtar, Naveed
    Kuepper, Bernhard
    Hunold, Peter
    Luechinger, Roger
    Puererfellner, Helmut
    Duru, Firat
    Gotte, M. J. W.
    Sutton, Richard
    Sommer, Torsten
    HEART RHYTHM, 2011, 8 (01) : 65 - 73