Comparison of dose response functions for EBT3 model GafChromic™ film dosimetry system

被引:28
|
作者
Aldelaijan, Saad [1 ,2 ,3 ,4 ]
Devic, Slobodan [1 ,2 ,5 ]
机构
[1] Jewish Gen Hosp, Radiat Oncol Dept, Montreal, PQ H3T 1E2, Canada
[2] McGill Univ, Med Phys Unit, Montreal, PQ H4A 3J1, Canada
[3] Montreal Neurol Inst, Biol & Biomed Engn Dept, Montreal, PQ H3A 2B4, Canada
[4] King Faisal Specialist Hosp & Res Ctr, Biomed Phys Dept, Riyadh, Saudi Arabia
[5] McGill Univ, Jewish Gen Hosp, Segal Canc Ctr, Montreal, PQ, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
Radiochromic film dosimetry system; Response function; PROTOCOL;
D O I
10.1016/j.ejmp.2018.05.014
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: Different dose response functions of EBT3 model GafChromic (TM) film dosimetry system have been compared in terms of sensitivity as well as uncertainty vs. error analysis. We also made an assessment of the necessity of scanning film pieces before and after irradiation. Methods: Pieces of EBT3 film model were irradiated to different dose values in Solid Water (SW) phantom. Based on images scanned in both reflection and transmission mode before and after irradiation, twelve different response functions were calculated. For every response function, a reference radiochromic film dosimetry system was established by generating calibration curve and by performing the error vs. uncertainty analysis. Results: Response functions using pixel values from the green channel demonstrated the highest sensitivity in both transmission and reflection mode. All functions were successfully fitted with rational functional form, and provided an overall one-sigma uncertainty of better than 2% for doses above 2 Gy. Use of pre-scanned images to calculate response functions resulted in negligible improvement in dose measurement accuracy. Conclusion: Although reflection scanning mode provides higher sensitivity and could lead to a more widespread use of radiochromic film dosimetry, it has fairly limited dose range and slightly increased uncertainty when compared to transmission scan based response functions. Double-scanning technique, either in transmission or reflection mode, shows negligible improvement in dose accuracy as well as a negligible increase in dose uncertainty. Normalized pixel value of the images scanned in transmission mode shows linear response in a dose range of up to 11 Gy.
引用
收藏
页码:112 / 118
页数:7
相关论文
共 50 条
  • [21] Characterization of Gafchromic EBT3 films for ultraviolet radiation dosimetry
    Aydarous, Abdulkadir
    Al-Omary, Enaam Ali
    El Ghazaly, Mahmoud
    RADIATION EFFECTS AND DEFECTS IN SOLIDS, 2014, 169 (03): : 249 - 255
  • [22] Linearization of EBT3 film dose response and virtual film dosimetry for SBRT quality assurance
    Cai, M.
    Archibald-Heeren, B.
    Wang, Y.
    Metcalfe, P.
    MICRO-MINI & NANO-DOSIMETRY & INNOVATIVE TECHNOLOGIES IN RADIATION THERAPY (MMND&ITRO2016), 2017, 777
  • [23] Effect of verification imaging on in vivo dosimetry results using Gafchromic EBT3 film
    Aland, Trent
    Moylan, Rebecca
    Kairn, Tanya
    Trapp, Jamie
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2016, 32 (11): : 1461 - 1465
  • [24] Dosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verification
    Borca, Valeria Casanova
    Pasquino, Massimo
    Russo, Giuliana
    Grosso, Pierangelo
    Cante, Domenico
    Sciacero, Piera
    Girelli, Giuseppe
    La Porta, Maria Rosa
    Tofani, Santi
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (02): : 158 - 171
  • [25] Absolute dose calibration of EBT3 Gafchromic films
    Campajola, L.
    Casolaro, P.
    Di Capua, F.
    JOURNAL OF INSTRUMENTATION, 2017, 12
  • [26] Evaluation of lens dose from anterior electron beams: comparison of Pinnacle and Gafchromic EBT3 film
    Sonier, Marcus
    Wronski, Matt
    Yeboah, Collins
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (02): : 304 - 314
  • [27] An adaptive method for recalibrating Gafchromic EBT3 film
    Chang, Liyun
    Yeh, Shyh-An
    Ho, Sheng-Yow
    Ding, Hueisch-Jy
    Chen, Pang-Yu
    Lee, Tsair-Fwu
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2021, 1007
  • [28] Evaluation of lens dose from anterior electron beams: comparison of Pinnacle and Gafchromic EBT3 film
    Sonier, M.
    Wronski, M.
    Yeboah, C.
    MEDICAL PHYSICS, 2014, 41 (08) : 12 - 12
  • [29] An Adaptive Method for Recalibrating Gafchromic EBT3 Film
    Chang, L.
    MEDICAL PHYSICS, 2020, 47 (06) : E695 - E695
  • [30] Utilizing GAFCHROMIC EBT3 film for precise skin dosimetry in head and neck cancer patients
    Kumar, Dinesh
    Pradhan, Anirudh
    Chauhan, Vishnu
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2024,