Biomechanical design of Less Invasive Stabilization System femoral plates: Computational evaluation of the fracture environment

被引:8
|
作者
Reina-Romo, Esther [1 ]
Giraldez-Sanchez, M. A. [2 ]
Mora-Macias, J. [1 ]
Cano-Luis, P. [2 ]
Dominguez, Jaime [1 ]
机构
[1] Univ Seville, Dept Mech Engn, Seville 41092, Spain
[2] Hosp Univ Virgen del Rocio, Dept Traumatol, Seville, Spain
关键词
Less Invasive Stabilization System; femoral plate; biomechanics; finite element; femoral fracture; TROCHANTERIC GAMMA-NAIL; FINITE-ELEMENT-ANALYSIS; LOCKING PLATE; MECHANICAL CONDITIONS; KNEE ARTHROPLASTY; INTERNAL-FIXATION; FEMUR FRACTURES; SCREW PLACEMENT; LISS; BONE;
D O I
10.1177/0954411914554634
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Less Invasive Stabilization System femoral plates are currently accepted as a suitable fixation technique for supra-intercondylar femoral fractures. However, general agreement does not exist regarding the optimum design of this fixator type. Therefore, the aim of this article is to reduce the intrinsic Less Invasive Stabilization System complications by clarifying, from a biomechanical point of view, how the number of screws, the screw connection type (unicortical or bicortical), or the structured position of the screws can influence the outcome of the fracture site. These studies include a specific finite element analysis that determines how several biomechanical variables, such as the movement at the fracture site, are influenced by the preconditions of bone healing. The results of this study show that the screw type affects the mechanical stabilization of the femur to a greater extent than the material type of the Less Invasive Stabilization System femoral plates. The most significant differences among all the analyzed configurations are observed in the shear interfragmentary strain between screw types. Values are approximately 50% higher with unicortical screws than with bicortical ones.
引用
收藏
页码:1043 / 1052
页数:10
相关论文
共 50 条
  • [31] Less invasive stabilising system plate versus retrograde intramedullary nail for osteoporotic distal femoral fracture
    Mukherjee, Pavel
    Bovell, Christopher
    Davis, James
    EUROPEAN JOURNAL OF ORTHOPAEDIC SURGERY AND TRAUMATOLOGY, 2010, 20 (04): : 281 - 284
  • [32] Novel slide compression anatomic plates of the femoral neck for treating unstable femoral neck fracture: A biomechanical study
    Li, Jia
    Yin, Pengbin
    Li, Jiantao
    Zhao, Zhe
    Zhao, Jingxin
    Cui, Xiang
    Lyu, Houchen
    Zhang, Licheng
    Tang, Peifu
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2023, 41 (05) : 1088 - 1096
  • [33] Thoracolumbar fracture stabilization: comparative biomechanical evaluation of a new video-assisted implantable system
    Schultheiss, M
    Hartwig, E
    Kinzl, L
    Claes, L
    Wilke, HJ
    EUROPEAN SPINE JOURNAL, 2004, 13 (02) : 93 - 100
  • [34] Thoracolumbar fracture stabilization: comparative biomechanical evaluation of a new video-assisted implantable system
    M. Schultheiss
    E. Hartwig
    L. Kinzl
    L. Claes
    H.-J. Wilke
    European Spine Journal, 2004, 13 : 93 - 100
  • [35] A biomechanical comparison of proximal femoral nails and locking proximal anatomic femoral plates in femoral fracture fixation A study on synthetic bones
    Ozkan, Korhan
    Turkmen, Ismail
    Sahin, Adem
    Yildiz, Yavuz
    Erturk, Selim
    Soylemez, Mehmet Salih
    INDIAN JOURNAL OF ORTHOPAEDICS, 2015, 49 (03) : 347 - 351
  • [36] Less invasive stabilization of complex tibial plateau fractures -: A biomechanical evaluation of a unilateral locked screw plate and double plating
    Gösling, T
    Schandelmaier, P
    Marti, A
    Hufner, T
    Partenheimer, A
    Krettek, C
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2004, 18 (08) : 546 - 551
  • [37] A biomechanical comparison of proximal femoral nails and locking proximal anatomic femoral plates in femoral fracture fixationA study on synthetic bones
    Korhan Ozkan
    İsmail Türkmen
    Adem Sahin
    Yavuz Yildiz
    Selim Erturk
    Mehmet Salih Soylemez
    Indian Journal of Orthopaedics, 2015, 49 : 347 - 351
  • [38] Fixation of distal femoral fractures above total knee arthroplasty utilizing the Less Invasive Stabilization System (LISS)
    Kregor, PJ
    Hughes, JL
    Cole, PA
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2001, 32 : 64 - 75
  • [39] A New System for Periprosthetic Fracture Stabilization-A Biomechanical Comparison
    Rau, Daniel
    Russow, Gabriele
    Heyland, Mark
    Wulsten, Dag
    Koesters, Clemens
    Schmoelz, Werner
    Maerdian, Sven
    JOURNAL OF CLINICAL MEDICINE, 2022, 11 (03)
  • [40] Biomechanical Evaluation of Dual Plate Configurations for Femoral Shaft Fracture Fixation
    El Beaino, Marc
    Morris, Randal P.
    Lindsey, Ronald W.
    Gugala, Zbigniew
    BIOMED RESEARCH INTERNATIONAL, 2019, 2019