Dosimetric characterization of the 60Co BEBIG Co0.A86 high dose rate brachytherapy source using PENELOPE

被引:16
|
作者
Guerrero, Rafael [1 ]
Almansa, Julio F. [2 ]
Torres, Javier [3 ]
Lallena, Antonio M. [4 ]
机构
[1] Hosp Univ San Cecilio, Serv Radiofis & Protecc Radiol, E-18012 Granada, Spain
[2] Hosp Reg Univ Virgen de las Nieves, Serv Fis & Protecc Radiol, E-18014 Granada, Spain
[3] Hosp Gen Univ Ciudad Real, Serv Radiofis & Protecc Radiol, E-13005 Ciudad Real, Spain
[4] Univ Granada, Dept Fis Atom Mol & Nucl, E-18071 Granada, Spain
来源
关键词
Monte Carlo dosimetry; Co-60; HDR; PENELOPE; TECHNICAL NOTE; AAPM;
D O I
10.1016/j.ejmp.2014.06.039
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Co-60 sources are being used as an alternative to Ir-192 sources in high dose rate brachytherapy treatments. In a recent document from AAPM and ESTRO, a consensus dataset for the Co-60 BEBIG (model Co0.A86) high dose rate source was prepared by using results taken from different publications due to discrepancies observed among them. The aim of the present work is to provide a new calculation of the dosimetric characteristics of that Co-60 source according to the recommendations of the AAPM and ESTRO report. Radial dose function, anisotropy function, air-kerma strength, dose rate constant and absorbed dose rate in water have been calculated and compared to the results of previous works. Simulations using the two different geometries considered by other authors have been carried out and the effect of the cable density and length has been studied. (C) 2014 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:960 / 967
页数:8
相关论文
共 50 条
  • [41] Evaluating dose distributions of high dose rate 60Co brachytherapy in an asymmetric tumor: A comparison of different designs of vaginal cylindrical applicators
    Aloraibi, Hayder adil
    Siavashpour, Zahra
    Ghorbani, Mahdi
    Abtahi, Marzieh
    jassim, Hussam
    Gholami, Somayeh
    APPLIED RADIATION AND ISOTOPES, 2024, 204
  • [42] Modern dosimetric tools for 60Co irradiation at high containment laboratories
    Twardoski, Barri
    Feldmann, Heinz
    Bloom, Marshall E.
    Ward, Joe
    INTERNATIONAL JOURNAL OF RADIATION BIOLOGY, 2011, 87 (10) : 1039 - 1044
  • [43] Dosimetric characterization of a new directional low-dose rate brachytherapy source
    Aima, Manik
    DeWerd, Larry A.
    Mitch, Michael G.
    Hammer, Clifford G.
    Culberson, Wesley S.
    MEDICAL PHYSICS, 2018, 45 (08) : 3848 - 3860
  • [44] Dose to pelvic lymph nodes during brachytherapy of locally advanced cervical cancer with 60Co HDR source
    Siavashpour, Zahra
    Aghili, Mahdi
    Anjidani, Shabnam
    Zayeri, Farid
    Moghani, Mona Molekzadeh
    Maleki, Maedeh
    Jaberi, Ramin
    BRACHYTHERAPY, 2022, 21 (02) : 158 - 169
  • [45] Prediction of the bias currents induced by 60Co via dose and dose rate
    Liu Changshi
    RADIATION EFFECTS AND DEFECTS IN SOLIDS, 2012, 167 (04): : 275 - 280
  • [46] Quality assurance tool for determination of position and transit time of a Co-60 source in high dose rate brachytherapy
    Piriz, Gustavo H.
    Ortega-Spina, Henry G.
    Gonzalez-Sprinberg, Gabriel A.
    APPLIED RADIATION AND ISOTOPES, 2021, 178
  • [47] Quality Assurance Tool for Determination of Position and Transit Time of a Co-60 Source in High Dose Rate Brachytherapy
    Piriz, G.
    Ortega, H.
    Gonzalez-Sprinberg, G.
    MEDICAL PHYSICS, 2020, 47 (06) : E836 - E836
  • [48] Monte Carlo calculation of the ionization chamber response to 60Co beam using PENELOPE
    Yi, Chul-Young
    Hah, Suck-Ho
    Yeom, Min Sun
    MEDICAL PHYSICS, 2006, 33 (05) : 1213 - 1221
  • [49] Absolute dose measurements using radiochromic films in 192Ir and 60Co HDR brachytherapy
    Oliver Canamas, Laura
    Jose Rovira-Escutia, Juan
    Vijande-Asenjo, Javier
    Candela-Juan, Cristian
    Gimeno-Olmos, Jose
    Carmen Pujades-Claumarchirant, Ma
    Angel Martinez-Navarro, Jose
    Zaragoza-Carrillo, Antonio
    Garcia-Cases, Francisco
    Ballester-Pallares, Facundo
    Perez-Calatayud, Jose
    RADIOTHERAPY AND ONCOLOGY, 2024, 194 : S350 - S352
  • [50] A comparative assessment of inhomogeneity and finite patient dimension effects in 60Co and 192Ir high-dose-rate brachytherapy
    Fotina, Irina
    Zourari, Kyveli
    Lahanas, Vasileios
    Pantelis, Evaggelos
    Papagiannis, Panagiotis
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2018, 10 (01) : 73 - 84