The intra- and inter-instrument reliability of DXA based on ex vivo soft tissue measurements

被引:71
|
作者
Guo, Y [1 ]
Franks, PW [1 ]
Brookshire, T [1 ]
Tataranni, PA [1 ]
机构
[1] NIDDKD, Clin Diabet & Nutr Sect, NIH, Phoenix, AZ 85016 USA
来源
OBESITY RESEARCH | 2004年 / 12卷 / 12期
关键词
phantom; DXA; comparison; percentage of fat; reliability;
D O I
10.1038/oby.2004.241
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: Comparison of ex-vivo soft tissue measurements using the GE/Lunar pencil (DPX-L; GE/Lunar Co., Madison, WI) and fan beam (Prodigy dual-energy X-ray absorptiometers (DXA) GE/Lunar Co.). Research Methods and Procedures: Intra-instrument reliability was assessed by repeatedly scanning soft tissue phantoms for lean tissue (water) and fat tissue (methanol) using one DPX-L and two identical Prodigy DXAs at fast, medium, and slow scan modes. For each machine, 10 scans of each phantom were performed at each scan speed. The number of scans per instrument totaled 60. Data were analyzed using ANOVA to ascertain whether scan speed affected the intra-instrument reliability and to test whether soft tissue measurements differed among instruments. Percentage fat (phantom density) was the outcome variable. Results: Intra-instrument reliability, expressed as coefficient of variation, ranged between 0.7% and 5.2% for the DPX-L and 0.4% and 4.5% for the Prodigy, with the lowest coefficients of variation observed when scanning the fat tissue phantom. Scan speed also affected the intra-instrument reliability (p < 0.01). Furthermore, differences in the measurement of percentage body fat for both the lean and fat tissue phantoms were observed among all three absorptiometers (all p < 0.01). After adjusting for scan speed, differences persisted for all three instruments. Discussion: Intra- and inter-instrument reliability of DXA machines, even those from the same manufacturer, remains unpredictable. Thus, when measuring body composition using DXA, it is important to consider that even in the absence of measurement bias, the use of different DXA machines, particularly when using a variety of speed settings, will increase the residual error around the true value.
引用
收藏
页码:1925 / 1929
页数:5
相关论文
共 50 条
  • [31] Intra- and inter-rater reliability of muscle and fat thickness measurements obtained using portable ultrasonography in older adults
    de Oliveira, Gabriela Benatti
    Fernandes, Lara Vilar
    Chen, Xiayu Summer
    Andrade, Flavia Cristina Drumond
    Costa, Luciana Scarlazzari
    Vasques, Ana Carolina Junqueira
    Corona, Ligiana Pires
    CLINICAL NUTRITION ESPEN, 2024, 60 : 65 - 72
  • [32] Intra- and Inter-Rater Reliability of Three Measurements for Assessing Tactile Acuity in Individuals with Chronic Low Back Pain
    Wang, Juan
    Chen, Changcheng
    Peng, Mengsi
    Wang, Yizu
    Wu, Bao
    Zheng, Yili
    Wang, Xueqiang
    EVIDENCE-BASED COMPLEMENTARY AND ALTERNATIVE MEDICINE, 2020, 2020
  • [33] Intra- and inter-rater reliability of the Assessment of Children's Hand Skills based on video recordings
    Chien, Chi-Wen
    Scanlon, Clare
    Rodger, Sylvia
    Copley, Jodie
    BRITISH JOURNAL OF OCCUPATIONAL THERAPY, 2014, 77 (02) : 82 - 90
  • [34] In-vivo assessment of tissue parameters with intra- and inter-patient variation using dual-energy CT
    Peters, N.
    Kieslich, A.
    Wohlfahrt, P.
    Hofmann, C.
    Richter, C.
    RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S587 - S588
  • [35] Intra- and inter-operator reliability assessment of a novel extramedullary accelerometer-based smart cutting guide for total knee arthroplasty: an in vivo study
    Giulio Maria Marcheggiani Muccioli
    Domenico Alesi
    Arcangelo Russo
    Mirco Lo Presti
    Iacopo Sassoli
    Matteo La Verde
    Stefano Zaffagnini
    International Orthopaedics, 2023, 47 : 83 - 87
  • [36] Intra- and inter-operator reliability assessment of a novel extramedullary accelerometer-based smart cutting guide for total knee arthroplasty: an in vivo study
    Muccioli, Giulio Maria Marcheggiani
    Alesi, Domenico
    Russo, Arcangelo
    Lo Presti, Mirco
    Sassoli, Iacopo
    La Verde, Matteo
    Zaffagnini, Stefano
    INTERNATIONAL ORTHOPAEDICS, 2023, 47 (01) : 83 - 87
  • [37] Assessment of needle-tissue force models based on ex vivo measurements
    Traczynski, Marek
    Patalas, Adam
    Roslan, Katarzyna
    Suszynski, Marcin
    Talar, Rafa l
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2024, 150
  • [38] Inter- and Intra-Rater Reliability of Soft Tissue Palpation Scoring in the Equine Thoracic Epaxial Region
    Merrifield-Jones, Megan
    Tabor, Gillian
    Williams, Jane
    JOURNAL OF EQUINE VETERINARY SCIENCE, 2019, 83
  • [39] Reliability of Ultrasound Diameter Measurements in Patients with a Small Asymptomatic Popliteal Artery Aneurysm: An Intra- and Inter-observer Agreement Study
    Zwiers, I.
    Hoogland, C. M. T.
    Mackaay, A. J. C.
    EUROPEAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2016, 51 (03) : 410 - 414
  • [40] Intra- and inter-session reliability of traditional and entropy-based variables describing stance on a wobble board
    Bizovska, Lucia
    Janura, Miroslav
    Svoboda, Zdenek
    Cerny, Martin
    Krohova, Jana
    Smondrk, Maros
    MEDICAL ENGINEERING & PHYSICS, 2017, 50 : 29 - 34