The intra- and inter-instrument reliability of DXA based on ex vivo soft tissue measurements

被引:71
|
作者
Guo, Y [1 ]
Franks, PW [1 ]
Brookshire, T [1 ]
Tataranni, PA [1 ]
机构
[1] NIDDKD, Clin Diabet & Nutr Sect, NIH, Phoenix, AZ 85016 USA
来源
OBESITY RESEARCH | 2004年 / 12卷 / 12期
关键词
phantom; DXA; comparison; percentage of fat; reliability;
D O I
10.1038/oby.2004.241
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: Comparison of ex-vivo soft tissue measurements using the GE/Lunar pencil (DPX-L; GE/Lunar Co., Madison, WI) and fan beam (Prodigy dual-energy X-ray absorptiometers (DXA) GE/Lunar Co.). Research Methods and Procedures: Intra-instrument reliability was assessed by repeatedly scanning soft tissue phantoms for lean tissue (water) and fat tissue (methanol) using one DPX-L and two identical Prodigy DXAs at fast, medium, and slow scan modes. For each machine, 10 scans of each phantom were performed at each scan speed. The number of scans per instrument totaled 60. Data were analyzed using ANOVA to ascertain whether scan speed affected the intra-instrument reliability and to test whether soft tissue measurements differed among instruments. Percentage fat (phantom density) was the outcome variable. Results: Intra-instrument reliability, expressed as coefficient of variation, ranged between 0.7% and 5.2% for the DPX-L and 0.4% and 4.5% for the Prodigy, with the lowest coefficients of variation observed when scanning the fat tissue phantom. Scan speed also affected the intra-instrument reliability (p < 0.01). Furthermore, differences in the measurement of percentage body fat for both the lean and fat tissue phantoms were observed among all three absorptiometers (all p < 0.01). After adjusting for scan speed, differences persisted for all three instruments. Discussion: Intra- and inter-instrument reliability of DXA machines, even those from the same manufacturer, remains unpredictable. Thus, when measuring body composition using DXA, it is important to consider that even in the absence of measurement bias, the use of different DXA machines, particularly when using a variety of speed settings, will increase the residual error around the true value.
引用
收藏
页码:1925 / 1929
页数:5
相关论文
共 50 条
  • [21] Intra- and Inter-Rater Reliability of Processing Ultrasound Tissue Characterization Scans in Midportion Achilles Tendinopathy
    Paantjens, Marc Adriaan
    Helmhout, Pieter Herman
    Backx, Franciscus Jacobus Gerardus
    Martens, Meindert Thomas Annechien Willibrord
    van Dongen, Joeri Paulus Adrianus
    Bakker, Eric Wilhelmus Petrus
    TRANSLATIONAL SPORTS MEDICINE, 2022, 2022
  • [22] Inter-Instrument Reliability and Agreement of Fitbit Charge Measurements of Heart Rate and Activity at Rest, during the Modified Canadian Aerobic Fitness Test, and in Recovery
    Nazari, Goris
    MacDermid, Joy C.
    Sinden, Kathryn E.
    Richardson, Julie
    Tang, Ada
    PHYSIOTHERAPY CANADA, 2019, 71 (03) : 197 - 206
  • [23] INTRA- AND INTER-RATER RELIABILITY OF IN VIVO THREE-DIMENSIONAL MEASUREMENT OF THE SCAPULAR POSES USING A MARKER-BASED LOCATOR
    Kuo, Mei-Ying
    Hong, Shih-Wun
    Li, Jia-Da
    Lu, Tung-Wu
    Hsu, Horng-Chaung
    BIOMEDICAL ENGINEERING-APPLICATIONS BASIS COMMUNICATIONS, 2018, 30 (01):
  • [24] Intra- And Inter-rater Reliability Of Sonographic Cross-sectional Area Measurements In Asymptomatic Patella Tendons
    Reschke, Antje
    Cassel, Michael
    Intziegianni, Konstantina
    Mayer, Frank
    MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 2014, 46 (05): : 927 - 928
  • [25] THE INTRA- AND INTER-INDIVIDUAL VARIABILITY OF LAPAROSCOPIC SMALL BOWEL LENGTH MEASUREMENT IN BARIATRIC SURGERY: AN EX VIVO EXPERIMENT
    Slagter, N.
    Van Wilsum, M.
    de Heide, L.
    Jutte, E.
    Damen, S.
    van Beek, A.
    OBESITY SURGERY, 2022, 32 (SUPPL 2) : 546 - 546
  • [26] Intra- and inter-observer reliability of perineal body volume measurements with 3-dimensional transperineal ultrasound
    Rahim, A.
    Ahmad, A.
    Tailor, V
    El Shamy, T.
    Bhide, A.
    Digesu, A.
    Fernando, R.
    Khullar, V
    INTERNATIONAL UROGYNECOLOGY JOURNAL, 2020, 31 (SUPPL 1) : S103 - S103
  • [27] Intra- and inter-individual variability in the mechanical properties of the human skin from in vivo measurements on 20 volunteers
    Jacquet, E.
    Chambert, J.
    Pauchot, J.
    Sandoz, P.
    SKIN RESEARCH AND TECHNOLOGY, 2017, 23 (04) : 491 - 499
  • [28] Intra- and inter-individual variability in measurements of biomarkers for oxidative damage in vivo:: Nutrition and Breast Health Study
    Kato, I
    Ren, J
    Heilbrun, LK
    Djuric, Z
    BIOMARKERS, 2006, 11 (02) : 143 - 152
  • [29] Intra- and inter-observer reliability of ten major histological scoring systems used for the evaluation of in vivo cartilage repair
    Davide Edoardo Bonasia
    Antongiulio Marmotti
    Alessandro Domenico Felice Massa
    Andrea Ferro
    Davide Blonna
    Filippo Castoldi
    Roberto Rossi
    Knee Surgery, Sports Traumatology, Arthroscopy, 2015, 23 : 2484 - 2493
  • [30] Intra- and inter-observer reliability of ten major histological scoring systems used for the evaluation of in vivo cartilage repair
    Bonasia, Davide Edoardo
    Marmotti, Antongiulio
    Massa, Alessandro Domenico Felice
    Ferro, Andrea
    Blonna, Davide
    Castoldi, Filippo
    Rossi, Roberto
    KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2015, 23 (09) : 2484 - 2493