Initial phantom studies for an office-based low-field MR system for prostate biopsy

被引:4
|
作者
Chiragzada, Selin [1 ]
Hellman, Eva [1 ]
Michael, Duncan [1 ]
Narayanan, Ram [1 ]
Nacev, Aleksandar [1 ]
Kumar, Dinesh [1 ]
机构
[1] Promaxo Inc, Oakland, CA 94607 USA
关键词
Targeted prostate biopsy; Low-field MRI; Office-based MRI; Navigation accuracy; Prostate cancer; MR-guided biopsy; ULTRASOUND; BRACHYTHERAPY; ACCURACY; GUIDANCE; FUSION; TIME;
D O I
10.1007/s11548-021-02364-7
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Purpose Prostate cancer is the second most prevalent cancer in US men, with about 192,000 new cases and 33,000 deaths predicted for 2020. With only a 31% 5-year survival rate for patients with an initial diagnosis of stage-four prostate cancer, the necessity for early screening and diagnosis is clear. In this paper, we present navigation accuracy results for Promaxo's MR system intended to be used in a physician's office for image-guided transperineal prostate biopsy. Methods The office-based low-field MR system was used to acquire images of prostate phantoms with needles inserted through a transperineal template. Coordinates of the estimated sample core locations in the office-based MR system were compared to ground truth needle coordinates identified in a 1.5T external reference scan. The error was measured as the distance between the planned target and the ground truth core center and as the shortest perpendicular distance between the planned target and the ground truth trajectory of the whole core. Results The average error between the planned target and the ground truth core center was 2.57 +/- 1.02 mm, [1.93-3.21] 95% CI. The average error between the planned target to the actual core segment was 2.05 +/- 1.24 mm, [1.53-2.56] 95% CI. Conclusion The average navigation errors were below the clinically significant threshold of 5 mm. The initial phantom results demonstrate the feasibility of the office-based system for prostate biopsy.
引用
收藏
页码:741 / 748
页数:8
相关论文
共 50 条
  • [31] MR-compatible Assistance System for Biopsy in a High-Field-Strength System: Initial Results in Patients with Suspicious Prostate Lesions
    Zangos, Stephan
    Melzer, Andreas
    Eichler, Katrin
    Sadighi, Cyrus
    Thalhammer, Axel
    Bodelle, Boris
    Wolf, Renate
    Gruber-Rouh, Tatjana
    Proschek, Dirk
    Hammerstingl, Renate
    Mueller, Cindy
    Mack, Martin G.
    Vogl, Thomas J.
    RADIOLOGY, 2011, 259 (03) : 903 - 910
  • [32] Retrospective 5-Year Analysis of MR-Guided Biopsies in a Low-Field MR System
    Zangos, S.
    Mueller, C.
    Mayer, F.
    Naguib, N. N.
    Nour-Eldin, N-E. A.
    Hansmann, M-L.
    Herzog, C.
    Hammerstingl, R. M.
    Thalhammer, A.
    Mack, M.
    Vogl, T. J.
    Eichler, K.
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2009, 181 (07): : 658 - 663
  • [33] Clinical evaluation of office-based MR-guided transperineal prostate biopsies without the use of antibiotic prophylaxis
    Boesen, L.
    Norgaard, N.
    Bisbjerg, R.
    Al-Hamadani, M. M. N.
    Sjolin, C. S.
    Logager, V. B.
    EUROPEAN UROLOGY, 2024, 85 : S2191 - S2191
  • [34] Comparison and Validation of Distortion Correction on a Low-Field MR-Guided Radiotherapy System
    Gao, Y.
    Pham, J.
    Kalbasi, A.
    Raldow, A.
    Hu, P.
    Yang, Y.
    MEDICAL PHYSICS, 2020, 47 (06) : E414 - E415
  • [35] Low-field MR arthrography of the shoulder: early results using an open 0.2 T MR system.
    Low, R
    Kreitner, KF
    Runkel, M
    Zollner, J
    Thelen, M
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 1998, 168 (04): : 316 - 322
  • [36] MR-guided transgluteal biopsies with an open low-field system in patients with clinically suspected prostate cancer: technique and preliminary results
    Zangos, S
    Eichler, K
    Engelmann, K
    Ahmed, M
    Dettmer, S
    Herzog, C
    Pegios, W
    Wetter, A
    Lehnert, T
    Mack, MG
    Vogl, TJ
    EUROPEAN RADIOLOGY, 2005, 15 (01) : 174 - 182
  • [37] MR-guided transgluteal biopsies with an open low-field system in patients with clinically suspected prostate cancer: technique and preliminary results
    Stephan Zangos
    Katrin Eichler
    Kerstin Engelmann
    Mukhtiar Ahmed
    Sebastian Dettmer
    Christopher Herzog
    Wasilios Pegios
    A. Wetter
    Thomas Lehnert
    Martin G. Mack
    Thomas J. Vogl
    European Radiology, 2005, 15 : 174 - 182
  • [38] Characterization of the Initial Hydration Process of Ordinary Portland Cement Based on Low-Field NMR
    Huo, Jin-hua
    Peng, Zhi-gang
    Ye, Zhongbin
    Feng, Qian
    Zheng, Yong
    Zhang, Jian
    APPLIED MAGNETIC RESONANCE, 2019, 50 (1-3) : 187 - 198
  • [39] Characterization of the Initial Hydration Process of Ordinary Portland Cement Based on Low-Field NMR
    Jin-hua Huo
    Zhi-gang Peng
    Zhongbin Ye
    Qian Feng
    Yong Zheng
    Jian Zhang
    Applied Magnetic Resonance, 2019, 50 : 187 - 198
  • [40] Setting up MR compatibility of a commercial stereo-localization system for low-field open MR interventional procedures
    Romain Viard
    Maximilien Vermandel
    Jean Rousseau
    International Journal of Computer Assisted Radiology and Surgery, 2009, 4 : 65 - 69