With respect to rotary blood pumps used as left ventricular assist devices (LVADs), it is clinically important to control pump flow to avoid complications associated with overor under-pumping of the native heart. By employing only the non-invasive observer of instantaneous pump impeller speed to assess flow dynamics, a number of physiologically significant pumping states may be detected. Based on a number of acute animal experiments, five such states were identified: regurgitant pump flow (PR), ventricular ejection (VE), nonopening of the aortic valve (ANO), and partial collapse (intermittent and continuous) of the ventricle wall (PVC-I and PVC-C). Two broader states, normal (corresponding to VE, ANO) and suction (corresponding to PVC-I, PVC-C) were readily discernable in clinical data from human patients implanted with LVADs. Based on data from both the animal experiments (N = 6) and the human patients (N = 10), a strategy for the automated non-invasive detection of significant pumping states has been developed and validated. Employing a classification and regression tree (CART), this system detects pumping states with a high degree of accuracy: state VE 87.5/100.0% (sensitivity/specificity); state ANO - 98.1/92.5%; state PVC-1 - 90.0/90.2%; state PVC-C - 61.2/98.0%. With a simplified binary scheme differentiating suction and normal states, both states were detected without error in data from the animal experiments, and with a sensitivity/specificity, for detecting suction, of 99.2/98.3% in the human patient data.
引用
收藏
页码:4216 / +
页数:2
相关论文
共 50 条
  • [1] Evaluation of Suction Detection During Different Pumping States in an Implantable Rotary Blood Pump
    Ng, Siew-Cheok
    Lim, Einly
    Mason, David G.
    Avolio, Alberto P.
    Lovell, Nigel H.
    [J]. ARTIFICIAL ORGANS, 2013, 37 (08) : E145 - E154
  • [2] Identifying physiologically significant pumping states in implantable rotary blood pumps using non-invasive system observers
    Ayre, PJ
    Lovell, NH
    [J]. PROCEEDINGS OF THE 25TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4: A NEW BEGINNING FOR HUMAN HEALTH, 2003, 25 : 439 - 442
  • [3] Identification and classification of physiologically significant pumping states in an implantable rotary blood pump
    Karantonis, Dean M.
    Lovell, Nigel H.
    Ayre, Peter J.
    Mason, David G.
    Cloherty, Shaun L.
    [J]. ARTIFICIAL ORGANS, 2006, 30 (09) : 671 - 679
  • [4] Non-invasive estimation and control of inlet pressure in an implantable rotary blood pump for heart failure patients
    AlOmari, A. H.
    Savkin, A. V.
    Ayre, P. J.
    Lim, E.
    Mason, D. G.
    Salamonsen, R. F.
    Fraser, J. F.
    Lovell, N. H.
    [J]. PHYSIOLOGICAL MEASUREMENT, 2011, 32 (08) : 1035 - 1060
  • [5] Non-invasive flow estimation in an implantable rotary blood pump: a study considering non-pulsatile and pulsatile flows
    Ayre, PJ
    Lovell, NH
    Woodard, JC
    [J]. PHYSIOLOGICAL MEASUREMENT, 2003, 24 (01) : 179 - 189
  • [6] Non-invasive estimation of pulsatile flow and differential pressure in an implantable rotary blood pump for heart failure patients
    AlOmari, A. H.
    Savkin, A. V.
    Karantonis, D. M.
    Lim, E.
    Lovell, N. H.
    [J]. PHYSIOLOGICAL MEASUREMENT, 2009, 30 (04) : 371 - 386
  • [7] Classification of physiologically significant pumping states in an Implantable rotary blood pump: Patient trial results
    Karantonis, Dean M.
    Mason, David G.
    Salamonsen, Robert F.
    Ayre, Peter J.
    Cloherty, Shaun L.
    Lovell, Nigel H.
    [J]. ASAIO JOURNAL, 2007, 53 (05) : 617 - 622
  • [8] Classification of physiologically significant pumping states in an implantable rotary blood pump: Effects of cardiac rhythm disturbances
    Karantonis, Dean M.
    Lovell, Nigel H.
    Ayre, Peter J.
    Mason, David G.
    Cloherty, Shaun L.
    [J]. ARTIFICIAL ORGANS, 2007, 31 (06) : 476 - 479
  • [9] Classification of Implantable Rotary Blood Pump States With Class Noise
    Ooi, Hui-Lee
    Seera, Manjeevan
    Ng, Siew-Cheok
    Lim, Chee Peng
    Loo, Chu Kiong
    Lovell, Nigel H.
    Redmond, Stephen J.
    Lim, Einly
    [J]. IEEE JOURNAL OF BIOMEDICAL AND HEALTH INFORMATICS, 2016, 20 (03) : 829 - 837
  • [10] State transitions in an implantable rotary blood pump
    Lovell, NH
    Ayre, PJ
    [J]. PROCEEDINGS OF THE SECOND IASTED INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING, 2004, : 69 - 72