Classification of physiologically significant pumping states in an implantable rotary blood pump: Effects of cardiac rhythm disturbances

被引:10
|
作者
Karantonis, Dean M. [1 ]
Lovell, Nigel H.
Ayre, Peter J.
Mason, David G.
Cloherty, Shaun L.
机构
[1] Univ New S Wales, Grad Sch Biomed Engn, Sydney, NSW 2052, Australia
[2] Univ New S Wales, Sch Elect Engn & Telecommun, Sydney, NSW 2052, Australia
[3] Ventracor Ltd, Sydney, NSW, Australia
[4] Natl Informat & Commun Technol, Eveleigh, NSW, Australia
关键词
implantable rotary blood pump; pumping states; control strategy; left ventricular assist device; arrhythmias;
D O I
10.1111/j.1525-1594.2007.00409.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Methods of speed control for implantable rotary blood pumps (iRBPs) are vital for providing implant recipients with sufficient blood flow to cater for their physiological requirements. The detection of pumping states that reflect the physiological state of the native heart forms a major component of such a control method. Employing data from a number of acute animal experiments, five such pumping states have been previously identified: regurgitant pump flow, ventricular ejection (VE), nonopening of the aortic valve (ANO), and partial collapse (intermittent [PVC-I] and continuous [PVC-C]) of the ventricle wall. An automated approach that noninvasively detects such pumping states, employing a classification and regression tree (CART), has also been developed. An extension to this technique, involving an investigation into the effects of cardiac rhythm disturbances on the state detection process, is discussed. When incorporating animal data containing arrhythmic events into the CART model, the strategy showed a marked improvement in detecting pumping states as compared to the model devoid of arrhythmic data: state VE-57.4/91.7% (sensitivity/specificity) improved to 97.1/100.0%; state PVC-I-66.7/83.1% improved to 100.0/88.3%, and state PVC-C-11.1/66.2% changed to 0.0/100%. With a simplified binary scheme differentiating suction (PVC-I, PVC-C) and nonsuction (VE, ANO) states, suction was initially detected with 100/98.5% sensitivity/specificity, whereas with the subsequent improved model, both these states were detected with 100% sensitivity. The accuracy achieved demonstrates the robustness of the technique presented, and substantiates its inclusion into any iRBP control methodology.
引用
收藏
页码:476 / 479
页数:4
相关论文
共 7 条
  • [1] Identification and classification of physiologically significant pumping states in an implantable rotary blood pump
    Karantonis, Dean M.
    Lovell, Nigel H.
    Ayre, Peter J.
    Mason, David G.
    Cloherty, Shaun L.
    [J]. ARTIFICIAL ORGANS, 2006, 30 (09) : 671 - 679
  • [2] Classification of physiologically significant pumping states in an Implantable rotary blood pump: Patient trial results
    Karantonis, Dean M.
    Mason, David G.
    Salamonsen, Robert F.
    Ayre, Peter J.
    Cloherty, Shaun L.
    Lovell, Nigel H.
    [J]. ASAIO JOURNAL, 2007, 53 (05) : 617 - 622
  • [3] Identifying physiologically significant pumping states in implantable rotary blood pumps using non-invasive system observers
    Ayre, PJ
    Lovell, NH
    [J]. PROCEEDINGS OF THE 25TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4: A NEW BEGINNING FOR HUMAN HEALTH, 2003, 25 : 439 - 442
  • [4] Classification of Implantable Rotary Blood Pump States With Class Noise
    Ooi, Hui-Lee
    Seera, Manjeevan
    Ng, Siew-Cheok
    Lim, Chee Peng
    Loo, Chu Kiong
    Lovell, Nigel H.
    Redmond, Stephen J.
    Lim, Einly
    [J]. IEEE JOURNAL OF BIOMEDICAL AND HEALTH INFORMATICS, 2016, 20 (03) : 829 - 837
  • [5] Automated non-invasive detection of pumping states in an implantable rotary blood pump
    Karantonis, Dean M.
    Cloherty, Shaun L.
    Mason, David G.
    Salarnonsen, Robert F.
    Ayre, Peter J.
    Lovell, Nigel H.
    [J]. 2006 28TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-15, 2006, : 4216 - +
  • [6] Evaluation of Suction Detection During Different Pumping States in an Implantable Rotary Blood Pump
    Ng, Siew-Cheok
    Lim, Einly
    Mason, David G.
    Avolio, Alberto P.
    Lovell, Nigel H.
    [J]. ARTIFICIAL ORGANS, 2013, 37 (08) : E145 - E154
  • [7] Identifying physiologically significant pumping state transitions in implantable rotary blood pumps used as left ventricular assist devices:: an in-vivo study
    Ayre, PJ
    Lovell, NH
    Morris, RW
    Wilson, ML
    Woodard, JC
    [J]. PROCEEDINGS OF THE 23RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4: BUILDING NEW BRIDGES AT THE FRONTIERS OF ENGINEERING AND MEDICINE, 2001, 23 : 445 - 448