Test methods to determine magnetic resonance (MR) safety and MR image compatibility of implants/devices

被引:0
|
作者
Schaefers, Gregor [1 ,2 ]
机构
[1] MR Comp GmbH, Pruflab MR Sicherheit & Kompatibilitat, Buschgrundstr 23, D-45894 Gelsenkirchen, Germany
[2] MRI STaR Magnet Resonance Inst Safety Technol & R, Gelsenkirchen, Germany
来源
RADIOLOGE | 2019年 / 59卷 / 10期
关键词
MR safety; MR testing methods; MRI; RF heating; Implants;
D O I
10.1007/s00117-019-00588-y
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Methodical innovations In the present article, interactions associated with magnetic resonance (MR) procedures and MR test procedures for implants/devices are examined. Performance Since 2012, many interactions of items with MR procedures have been physically described and translated into standardized ASTM and ISO testing procedures. Despite the standardized procedures, the determination of the test method to use is an important decision. The MR user is also responsible for the transfer and interpretation of the individual technical parameters despite the MR Conditional labelling and therefore relatively unambiguous instruction. This includes the total MR examination duration, which often has no clinical practical duration, but is derived from the 15 & x202f;min of the ASTM radiofrequency (RF) heating test. Achievements There has been an increasing standardization of the test methods as well as the MR labeling requirements and the advantageous transfer of the parameters to suitable input masks on the MR systems. Practical recommendations The current use of standardized MR test methods and MR marking represents the best possible state of the art from the point of view of the approval of medical devices as well as from a liability point of view for the manufacturers of implants-and for MR users in clinical practice. However, off-label decisions (i.e., deviations from the manufacturer's official MR marking) in everyday clinical practice can be medically justified.
引用
收藏
页码:875 / 884
页数:10
相关论文
共 50 条
  • [21] Magnetic Resonance(MR) Image Reconstruction using Compressive Sensing
    Waseem, Mona
    Anwar, Syed
    INTERNATIONAL JOURNAL OF COMPUTER SCIENCE AND NETWORK SECURITY, 2015, 15 (01): : 66 - 70
  • [22] In vitro investigation of the electromagnetic interference between MR imaging and cochlear implants: Compatibility and safety aspects
    Christian, T
    Kremser, C
    Hochmair, ES
    HochmairDesoyer, IJ
    RADIOLOGY, 1997, 205 : 945 - 945
  • [23] Cochlear implants: In vitro investigation of electromagnetic interference at MR imaging-compatibility and safety aspects
    Teissl, C
    Kremser, C
    Hochmair, ES
    Hochmair-Desoyer, IJ
    RADIOLOGY, 1998, 208 (03) : 700 - 708
  • [24] Market potential of MR-compatible guidewires and assessment of MR safety and compatibility
    Uihlein, G.
    Schaefers, G.
    BIOMEDICAL ENGINEERING-BIOMEDIZINISCHE TECHNIK, 2014, 59 : S460 - S460
  • [25] A Research On The Compatibility Between MR/VR Devices And Exergaming
    Taguchi, Taiyo
    Ishikawa, Tomokazu
    2022 NICOGRAPH INTERNATIONAL, NICOINT 2022, 2022, : 17 - 20
  • [26] CLINICAL APPLICATION OF CARDIOVASCULAR MAGNETIC RESONANCE IN PATIENTS WITH MR-CONDITIONAL DEVICES: SAFETY, FEASIBILITY AND CLINICAL IMPACT
    Baritussio, A.
    De Garate, E.
    Dastidar, A. Ghosh
    Ahmed, N.
    Scatteia, A.
    Rodrigues, J.
    Lawton, C.
    Nisbet, A.
    Duncan, E.
    Cripps, T.
    Diab, I.
    Thomas, G.
    Bucciarelli-Ducci, C.
    HEART, 2016, 102 : A1 - A2
  • [27] MR Labeling Information for Implants and Devices: Explanation of Terminology
    Shellock, Frank G.
    Woods, Terry O.
    Crues, John V., III
    RADIOLOGY, 2009, 253 (01) : 26 - 30
  • [28] Comprehensive analysis of RF Test Lab System for MR Safety Testing of Medical Implants
    Kozlov, M.
    Schaefers, G.
    2014 IEEE ANTENNAS AND PROPAGATION SOCIETY INTERNATIONAL SYMPOSIUM (APSURSI), 2014, : 1194 - 1195
  • [29] Evaluation of PET and MR compatibility of dedicated radiotherapy positioning devices
    Zorz, A.
    Giannelli, M.
    Scaggion, A.
    Turco, P.
    Capotosto, A.
    Rossato, M.
    Zucchetta, P.
    Bui, F.
    Paiusco, M.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2018, 45 : S725 - S725
  • [30] Improving MR Image Quality in Patients with Metallic Implants
    Lee, Elizabeth M.
    Ibrahim, El-Sayed H.
    Dudek, Nancy
    Lu, Jimmy C.
    Kalia, Vivek
    Runge, Mason
    Srinivasan, Ashok
    Stojanovska, Jadranka
    Agarwal, Prachi P.
    RADIOGRAPHICS, 2021, 41 (04) : E126 - E137