The effects of pulse rate, artefact and pulse strength on oscillometric non-invasive blood pressure measurements.

被引:0
|
作者
Amoore, JN [1 ]
Geake, WB [1 ]
Scott, DHT [1 ]
机构
[1] Royal Infirm, Dept Med Phys & Med Engn, Edinburgh EH3 9YW, Midlothian, Scotland
关键词
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We examined the effects of pulse rate, pulse and artefact on 20 adult oscillometric non-invasive blood pressure measurements using NIBP test simulators. Over a range of pulse rates from 40 to 200 bpm the 20 NIBP monitors studied recorded little change in registered pressures. The monitors achieved consistent readings at low pulse rates by decreasing the deflation rate and hence increasing the determination time (from under 20 sec at 200 bpm to over 30 sec at 40 bpm). High levels of artefact produced artificially high systolic pressures in 12 of the 16 monitors that recorded pressures, 6 of them being over 30 mmHg above the target 120 mmHg; 4 of the monitors signalled failure to record due to excessive artefact. Pulse strengths of 25% of the nominal were coped with well by most monitors with a slight tendency for the monitors to record low systolic pressures.
引用
收藏
页码:1977 / 1978
页数:2
相关论文
共 50 条
  • [41] Non-invasive assessment of arterial stiffness using oscillometric blood pressure measurement
    Hidehiko Komine
    Yoshiyuki Asai
    Takashi Yokoi
    Mutsuko Yoshizawa
    BioMedical Engineering OnLine, 11
  • [42] Non-invasive assessment of arterial stiffness using oscillometric blood pressure measurement
    Komine, Hidehiko
    Asai, Yoshiyuki
    Yokoi, Takashi
    Yoshizawa, Mutsuko
    BIOMEDICAL ENGINEERING ONLINE, 2012, 11
  • [43] An automated system for the functional evaluation of oscillometric non-invasive blood pressure monitors
    Geake, WB
    Amoore, JN
    Scott, DHT
    JOURNAL OF MEDICAL ENGINEERING & TECHNOLOGY, 1995, 19 (05) : 162 - 176
  • [44] Non-invasive blood pressure monitoring with an oscillometric brachial cuff: impact of arrhythmia
    Lakhal, Karim
    Martin, Maelle
    Ehrmann, Stephan
    Faiz, Sofian
    Rozec, Bertrand
    Boulain, Thierry
    JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2018, 32 (04) : 707 - 715
  • [45] NON-INVASIVE VS INVASIVE BLOOD-PRESSURE MEASUREMENTS
    GRUNDY, BL
    PROCOPIO, P
    SIEBER, FE
    CRITICAL CARE MEDICINE, 1981, 9 (03) : 267 - 267
  • [46] Non-invasive blood pressure monitoring with an oscillometric brachial cuff: impact of arrhythmia
    Karim Lakhal
    Maëlle Martin
    Stephan Ehrmann
    Sofian Faiz
    Bertrand Rozec
    Thierry Boulain
    Journal of Clinical Monitoring and Computing, 2018, 32 : 707 - 715
  • [47] Non-Invasive Estimation of Pulse Pressure in HeartMate 3 Patients
    Gaudig, A.
    Pinsino, A.
    Braghieri, L.
    Wang, Y.
    Mondellini, G. M.
    Hastie, J. M.
    Panzer, O.
    Takeda, K.
    Naka, Y.
    Sayer, G. T.
    Uriel, N.
    Yuzefpolskaya, M.
    Colombo, P. C.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2021, 40 (04): : S99 - S100
  • [48] Blood pressure in the neonate: Three non-invasive oscillometric pressure monitors compared with invasively measured blood pressure
    Dannevig, I
    Dale, HC
    Liestol, K
    Lindemann, R
    ACTA PAEDIATRICA, 2005, 94 (02) : 191 - 196
  • [49] Differences in oscillometric non-invasive blood pressure measurements recorded by different revisions of the Philips Component Monitoring System
    Murray, IC
    Amoore, JN
    Scott, DH
    BLOOD PRESSURE MONITORING, 2005, 10 (04) : 215 - 222
  • [50] CONTINUOUS AND NON-INVASIVE BLOOD PRESSURE MEASUREMENT BASED ON PULSE TRANSIT TIME COMPARED TO AMBULATORY BLOOD PRESSURE MEASUREMENT
    Haberl, R.
    Wesely, M.
    Thimel, K.
    Gehring, J.
    Kuechler, G.
    JOURNAL OF HYPERTENSION, 2016, 34 : E287 - E287