The effects of pulse rate, artefact and pulse strength on oscillometric non-invasive blood pressure measurements.

被引:0
|
作者
Amoore, JN [1 ]
Geake, WB [1 ]
Scott, DHT [1 ]
机构
[1] Royal Infirm, Dept Med Phys & Med Engn, Edinburgh EH3 9YW, Midlothian, Scotland
关键词
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We examined the effects of pulse rate, pulse and artefact on 20 adult oscillometric non-invasive blood pressure measurements using NIBP test simulators. Over a range of pulse rates from 40 to 200 bpm the 20 NIBP monitors studied recorded little change in registered pressures. The monitors achieved consistent readings at low pulse rates by decreasing the deflation rate and hence increasing the determination time (from under 20 sec at 200 bpm to over 30 sec at 40 bpm). High levels of artefact produced artificially high systolic pressures in 12 of the 16 monitors that recorded pressures, 6 of them being over 30 mmHg above the target 120 mmHg; 4 of the monitors signalled failure to record due to excessive artefact. Pulse strengths of 25% of the nominal were coped with well by most monitors with a slight tendency for the monitors to record low systolic pressures.
引用
收藏
页码:1977 / 1978
页数:2
相关论文
共 50 条
  • [1] Enhancement of oscillometric index in non-invasive blood pressure measurements
    Jazbinsek, V.
    Luznik, J.
    Trontelj, Z.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 4: IMAGE PROCESSING, BIOSIGNAL PROCESSING, MODELLING AND SIMULATION, BIOMECHANICS, 2010, 25 : 1357 - 1360
  • [2] Acquiring the oscillometric envelopes for non-invasive blood pressure measurements
    Geršak, Gregor
    Elektrotehniski Vestnik/Electrotechnical Review, 2009, 76 (03): : 97 - 102
  • [3] Acquiring the oscillometric envelopes for non-invasive blood pressure measurements
    Gersak, Gregor
    ELEKTROTEHNISKI VESTNIK-ELECTROCHEMICAL REVIEW, 2009, 76 (03): : 97 - 102
  • [4] Comparison of non-invasive central blood pressure measurements using applanation tonometry and automated oscillometric radial pulse wave analysis
    Trinkmann, F.
    Benck, U.
    Halder, J.
    Saur, J.
    Borggrefe, M.
    Akin, I.
    Kaden, J. J.
    EUROPEAN HEART JOURNAL, 2017, 38 : 1160 - 1161
  • [5] Morphological Changes of Pressure Pulses in Oscillometric Non-Invasive Blood Pressure Measurements
    Avbelj, V.
    2014 37TH INTERNATIONAL CONVENTION ON INFORMATION AND COMMUNICATION TECHNOLOGY, ELECTRONICS AND MICROELECTRONICS (MIPRO), 2014, : 245 - 248
  • [6] Wavelet Estimation of Pulse Rate Variability from Oscillometric Blood Pressure Measurements
    Ahmad, Saif
    Bolic, Miodrag
    Dajani, Hilmi
    Groza, Voicu
    MEMEA: 2009 IEEE INTERNATIONAL WORKSHOP ON MEDICAL MEASUREMENTS AND APPLICATIONS, 2009, : 37 - 40
  • [7] Oscillometric non-invasive blood pressure measurements: The influence of the make of instrument on readings?
    Amoore, JN
    Geake, WB
    Scott, DHT
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1997, 35 (02) : 131 - 134
  • [8] Oscillometric non-invasive blood pressure measurements: the influence of the make of instrument on readings?
    J. N. Amoore
    W. B. Geake
    D. H. T. Scott
    Medical and Biological Engineering and Computing, 1997, 35 : 131 - 134
  • [9] Oscillometric non-invasive blood pressure measurements: The influence of the make of instrument on readings?
    Amoore, J.N.
    Geake, W.B.
    Scott, D.H.T.
    Medical and Biological Engineering and Computing, 1997, 35 (02): : 131 - 134
  • [10] NON-INVASIVE PRESSURE PULSE RECORDER
    JAFFRIN, MY
    RAO, S
    RAINES, JK
    PHYSICS IN MEDICINE AND BIOLOGY, 1972, 17 (05): : 721 - &