Left Atrial Appendage Closure Using the Amulet Device: An Initial Experience with the Second Generation Amplatzer Cardiac Plug

被引:21
|
作者
Lam, Simon Cheung Chi [1 ]
Bertog, Stefan [1 ]
Gafoor, Sameer [1 ]
Vaskelyte, Laura [1 ]
Boehm, Patrick [1 ]
Ho, Raymond Wei Jian [1 ]
Franke, Jennifer [1 ]
Hofmann, Ilona [1 ]
Sievert, Horst [1 ]
机构
[1] CardioVasc Ctr Frankfurt, D-60389 Frankfurt, Germany
关键词
left atrial appendage closure; structural heart intervention; new device; TRANSCATHETER OCCLUSION; FIBRILLATION; STROKE; RISK;
D O I
10.1002/ccd.25644
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
ObjectivesAim of this study was to demonstrate the feasibility, safety, and short-term outcome of left atrial appendage (LAA) closure with a new generation LAA closure device. BackgroundThe Amulet device (AGA, St Jude Medical, Minneapolis, MN) is a new generation of the amplatzer cardiac plug (ACP), specifically designed for LAA closure. This new version is designed to facilitate the implantation process and minimize procedural or device-related complications. MethodsThe device was implanted in 17 patients with nonvalvular atrial fibrillation (AF). Clinical data were obtained at baseline, during the procedure, at discharge, at 30 and 90 days. ResultsAll devices were implanted successfully. Device sizes ranged from 20 mm to 31 mm. A 12 French (Fr) or 14 Fr delivery sheath was used depending on the selected device size. Full and partial recapture was performed in 1 case and 3 cases, respectively. There was 1 procedure-related pericardial effusion successfully managed with pericardiocentesis. There was no device embolization. The mean length of stay was 2.10.3 days. At 90 days, there were no deaths, strokes, systemic thromboembolism, or bleeding complications. There was no device-related thrombus or pericardial effusion at 90-day TEE. In 2 of the 17 patients minimal peridevice flow (smaller than 2 mm) was present. ConclusionsThe Amulet device, which has new novel features as compared with the first generation ACP, is a feasible option for LAA closure. From our initial experience, implantation of the Amulet is associated with high success rate and good short-term outcome. (c) 2014 Wiley Periodicals, Inc.
引用
收藏
页码:297 / 303
页数:7
相关论文
共 50 条
  • [21] LEFT ATRIAL APPENDAGE CLOSURE WITH A SECOND GENERATION DEVICE
    Abualsaud, Ali O.
    Freixa, Xavier
    Nosair, Mohamed
    Chan, Jason Leung Kwai
    Tzikas, Apostolos
    Basmadjian, Arsene
    Garceau, Patrick
    Ibrahim, Reda
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2013, 62 (18) : B210 - B210
  • [22] Tools and Techniques Clinical: Transcatheter closure of left atrial appendage using the Amplatzer™ Cardiac Plug
    Berti, Sergio
    Santoro, Gennaro
    Palmieri, Cataldo
    Meucci, Francesco
    EUROINTERVENTION, 2013, 9 (04) : 524 - 526
  • [23] Closure of Left Atrial Appendage With Persistent Distal Thrombus Using an Amplatzer Amulet Occluder
    Lange, Mathias
    Bueltel, Helmut
    Weglage, Heinrich
    Loeffeld, Patrick
    Wichter, Thomas
    JOURNAL OF INVASIVE CARDIOLOGY, 2016, 28 (09): : E75 - E76
  • [24] Left atrial appendage occlusion with the Amplatzer Amulet: update on device sizing
    Freixa, Xavier
    Aminian, Adel
    Tzikas, Apostolos
    Saw, Jacqueline
    Nielsen-Kudsk, Jens-Erick
    Ghanem, Alexander
    Schmidt, Boris
    Hildick-Smith, David
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2020, 59 (01) : 71 - 78
  • [25] Extending percutaneous left atrial appendage closure indications using the AMPLATZER™ Cardiac Plug device in patients with persistent left atrial appendage thrombus: The thrombus trapping technique
    Jalal, Zakaria
    Iriart, Xavier
    Dinet, Marie-Lou
    Selly, Jean-Bernard
    Tafer, Nadir
    Renou, Pauline
    Sibon, Igor
    Thambo, Jean-Benoit
    ARCHIVES OF CARDIOVASCULAR DISEASES, 2016, 109 (12) : 659 - 666
  • [26] Left atrial appendage occlusion with the Amplatzer Amulet: update on device sizing
    Xavier Freixa
    Adel Aminian
    Apostolos Tzikas
    Jacqueline Saw
    Jens-Erick Nielsen-Kudsk
    Alexander Ghanem
    Boris Schmidt
    David Hildick-Smith
    Journal of Interventional Cardiac Electrophysiology, 2020, 59 : 71 - 78
  • [27] Thrombus formation after left atrial appendage exclusion using an amplatzer cardiac plug device
    Cruz-Gonzalez, Ignacio
    Martin Moreiras, Javier
    Garcia, Eulogio
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2011, 78 (06) : 970 - 973
  • [28] LEFT ATRIAL APPENDAGE PERIDEVICE LEAK CLOSURE WITH AMPLATZER PLUG
    Qasim, Muhammad
    Tariq, Rana
    Bang, Vigyan
    Hornsby, Kyle
    Mattson, Scott
    Zawacki, Kevin
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2021, 77 (18) : 2505 - 2505
  • [29] Comparison of transesophageal and intracardiac echocardiography in guiding percutaneous left atrial appendage closure with an Amplatzer Amulet device
    Streb, Witold
    Mitrega, Katarzyna
    Podolecki, Tomasz
    Morawski, Stanislaw
    Szulik, Mariola
    Jedrzejczyk-Patej, Ewa
    Kukulski, Tomasz
    Kalarus, Zbigniew
    POSTEPY W KARDIOLOGII INTERWENCYJNEJ, 2019, 15 (04): : 446 - 454
  • [30] Late presentation of left atrial appendage erosion and perforation by an Amplatzer™ Amulet™ closure device: a case report
    Ben Wilkins
    Carranza, Christian L.
    Sondergaard, Lars
    De Backer, Ole
    EUROPEAN HEART JOURNAL-CASE REPORTS, 2020, 4 (03)