High-performance liquid chromatographic assay of a new HIV-1 protease inhibitor, LB71350, in the plasma of dogs

被引:1
|
作者
Jeong, YN [1 ]
Seo, MK [1 ]
Choi, YJ [1 ]
Kim, IC [1 ]
Lee, YH [1 ]
机构
[1] LG Chem Biotech Res Inst, Taejon 305380, South Korea
来源
JOURNAL OF CHROMATOGRAPHY B | 1997年 / 703卷 / 1-2期
关键词
HIV-1 protease inhibitor; LB71350;
D O I
10.1016/S0378-4347(97)00395-2
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A reliable reversed-phase high-performance liquid chromatographic method has been developed for the determination of LB71350 in the plasma of dogs. The analyte was deproteinized with 1.5 volumes of methanol and 0.5 volumes of 10% zinc sulfate, and the supernatant was injected into a 5-mu m Capcell Pak C-18 column (150x4.6 mm I.D.). The mobile phase was a stepwise gradient mixture of acetonitrile and 0.2% triethylamine-HCl with a flow-rate of 1 ml/min and detection at UV 245 nm. The proportion of acetonitrile was kept at 52% for the first 6 min, increased to 100% for the next 0.5 min, kept at 100% for the next 2 min, decreased to 52% for the next 0.5 min, and finally kept at 52% for the next 7 min. The retention time of LB71350 was 6.9 min. The calibration was linear over the concentration range of 0.1-100 mg/l for dog plasma (r>0.997) and the limit of quantitation was 0.1 mg/l using 0.1 ml plasma. The quality control samples were reproducible with acceptable accuracy and precision at 0.1, 1, 10 and 100 mg/l concentrations. The within-day recover); (n=5) was 90.2-93.9%, the between-day recovery (n=5) was 89.5-93.5%, and the absolute between-day recovery (n=5) was 77-81%. The within-day precision (n=5) and between-day precision (n=5) were 2.59-5.82% and 3.17-4.55%, respectively. No interferences from endogenous substances were observed. Taken together, the above HPLC assay method by deproteinization and UV detection was suitable for the determination of LB71350 in the preclinical pharmacokinetics. (C) 1997 Elsevier Science B.V.
引用
收藏
页码:284 / 288
页数:5
相关论文
共 50 条
  • [31] HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC ASSAY OF IOMEPROL IN PLASMA AND URINE
    LORUSSO, V
    POGGESI, I
    DALFIUME, D
    TIRONE, P
    JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1990, 525 (02): : 401 - 409
  • [32] A simple and efficient high-performance liquid chromatographic assay for etomidate in plasma
    McIntosh, MP
    Rajewski, RA
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2001, 24 (04) : 689 - 694
  • [33] HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC ASSAY FOR AMILORIDE IN PLASMA AND URINE
    XU, DK
    ZHOU, JH
    YUAN, YS
    LIU, XQ
    HUANG, SK
    JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1991, 567 (02): : 451 - 458
  • [34] High-performance liquid chromatographic assay validation of Manumycin A in mouse plasma
    Gonzales, J
    Yeung, SCJ
    Smith, JA
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2002, 776 (02): : 177 - 182
  • [35] STEREOSPECIFIC HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC ASSAY OF SOTALOL IN PLASMA
    CARR, RA
    FOSTER, RT
    BHANJI, NH
    PHARMACEUTICAL RESEARCH, 1991, 8 (09) : 1195 - 1198
  • [36] Determination of indinavir, a HIV-1 protease inhibitor, in human plasma using ion-pair reversed-phase high-performance liquid chromatography
    Li, WZ
    Coombs, RW
    Collier, AC
    Raisys, VA
    THERAPEUTIC DRUG MONITORING, 1999, 21 (03) : 360 - 366
  • [37] A NEW RAPID AND SENSITIVE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC ASSAY FOR DICLOFENAC IN HUMAN PLASMA
    HANSES, A
    SPAHNLANGGUTH, H
    MUTSCHLER, E
    ARCHIV DER PHARMAZIE, 1995, 328 (03) : 257 - 260
  • [38] A high-performance liquid chromatographic assay for sparfloxacin
    Marona, HRN
    Schapoval, EES
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1999, 20 (03) : 413 - 417
  • [39] High-performance liquid chromatography assay for the determination of the HIV-protease inhibitor tipranavir in human plasma in combination with nine other antiretroviral medications
    Choi, Sun Ok
    Rezk, Naser L.
    Kashuba, Angela D. M.
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2007, 43 (04) : 1562 - 1567
  • [40] High-performance liquid chromatographic assay for minocycline in human plasma and parotid saliva
    Orti, V
    Audran, M
    Gibert, P
    Bougard, G
    Bressolle, F
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2000, 738 (02): : 357 - 365