Comparison of Visual Assessment Techniques for Wear and Corrosion in Modular Hip Replacement Systems

被引:5
|
作者
Di Prima, Matthew A. [1 ]
Vesnovsky, Oleg [1 ]
Kovacs, Paul
Hopper, Robert H., Jr. [2 ]
Ho, Henry [2 ]
Engh, Charles A., Jr. [2 ]
Topoleski, L. D. Timmie [1 ,3 ]
机构
[1] US FDA, Ctr Device & Radiol Hlth, 10903 New Hampshire Ave, Silver Spring, MD 20993 USA
[2] Anderson Orthopaed Res Inst, Alexandria, VA 22306 USA
[3] Univ Maryland Baltimore Cty, Dept Mech Engn, Baltimore, MD 21250 USA
关键词
metal-on-metal; corrosion; orthopaedic; wear; modular; hip arthroplasty; FRETTING CORROSION; TAPER JUNCTION; INTERFACE; FAILURE; HEAD; ARTHROPLASTY; RETRIEVAL; CREVICE;
D O I
10.1520/STP159120140152
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Wear and corrosion of implant metals have been associated with adverse tissue reactions and failures of total hips. However, the factors leading to component wear and corrosion, as well as the clinical effects of wear and corrosion, are not well understood. A visual inspection standard for taper corrosion and fretting, the Goldberg method, currently is used to quantify the amount of apparent corrosion damage present on an implant using a subjective score. We developed an approach to quantify the damage by creating digital microscopic images of the entire taper and trunnion surfaces, which enables an objective quantification of detailed wear and corrosion damage of the contacting surfaces, which we call the digital mosaic method (DMM). We compared the results of this new methodology to the Goldberg method and Anderson method (an expanded version of the Goldberg method). We examined the modular surfaces of ten retrieved metal-on-metal hip replacement systems. Each set of taper and trunnion surfaces were imaged and stitched into a composite image of the entire taper-trunnion surface. Three independent observers measured the surface area fraction of damage (wear or corrosion) for each surface, which in turn was compared with visual scores performed by independent observers on the same retrieved joints. To determine a measure of the intraobserver repeatability of the new method, a single reviewer measured the surface fraction of damage for the same set of explanted surfaces across different days, blinded to any previous specimen identification, with the image order randomized. The results of this study show that the DMM damage quantification correlates to the current visual scoring approaches (Goldberg and Anderson). The potential advantages to using this approach include better damage quantification, detailed analysis of specific locations, ability to identify matching damage locations on the two surfaces, and higher resolution of the contact area.
引用
收藏
页码:147 / 163
页数:17
相关论文
共 50 条
  • [21] A comparison of two different navigated hip replacement techniques on leg length discrepancy
    Jennison, T. N.
    Craig, P.
    Davis, Edward D.
    [J]. JOURNAL OF ORTHOPAEDICS, 2018, 15 (03) : 765 - 767
  • [22] The Computational Approach to Predicting Wear: Comparison of Wear Performance of CFR-PEEK and XLPE Liners in Total Hip Replacement
    Alpkaya, Alican Tuncay
    Mihcin, Senay
    [J]. TRIBOLOGY TRANSACTIONS, 2023, 66 (01) : 59 - 72
  • [23] Fretting corrosion testing of acetabular modular tapers for total hip replacements: A comparison of two designs
    Mace, Annsley
    Goodwin, Charley M.
    Gilbert, Jeremy L.
    [J]. JOURNAL OF ORTHOPAEDIC RESEARCH, 2023, 41 (08) : 1738 - 1745
  • [24] Total prosthetic hip replacement: an assessment of new techniques for transfusion and prophylaxis of venous thrombosis
    Grouille, D
    Orsel, I
    Ledan, C
    Benrhaiem, M
    Vandroux, D
    [J]. ANNALES FRANCAISES D ANESTHESIE ET DE REANIMATION, 1999, 18 (07): : 812 - 813
  • [25] Comparison of osteotomy level with three different canine total hip replacement systems
    Franklin, Samuel P.
    Franklin, Ashley L.
    Feyerabend, Nathan
    Hayashi, Kei
    Miller, Nathan A.
    Cook, James L.
    [J]. OPEN VETERINARY JOURNAL, 2022, 12 (06) : 975 - 979
  • [26] ASSESSING THE RESULTS OF HIP-REPLACEMENT - A COMPARISON OF 5 DIFFERENT RATING SYSTEMS
    CALLAGHAN, JJ
    DYSART, SH
    SAVORY, CF
    HOPKINSON, WJ
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1990, 72 (06): : 1008 - 1009
  • [27] Wear particles induce a new macrophage phenotype with the potential to accelerate material corrosion within total hip replacement interfaces
    Bijukumar, Divya Rani
    Salunkhe, Shruti
    Zheng, Guoxing
    Barba, Mark
    Halle, Deborah J.
    Pourzal, Robin
    Mathew, Mathew T.
    [J]. ACTA BIOMATERIALIA, 2020, 101 : 586 - 597
  • [28] The determination of linear and angular penetration of the femoral head into the acetabular component as an assessment of wear in total hip replacement - A comparison of four computer-assisted methods
    Geerdink, C. H.
    Grimm, B.
    Vencken, W.
    Heyligers, I. C.
    Tonino, A. J.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2008, 90B (07): : 839 - 846
  • [29] AIR-EMBOLISM DURING TOTAL HIP-REPLACEMENT - COMPARISON OF 2 SURGICAL TECHNIQUES
    EVANS, RD
    PALAZZO, MGA
    ACKERS, JWL
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 1989, 62 (03) : 243 - 247
  • [30] BODY-FAT ASSESSMENT - A COMPARISON OF VISUAL ESTIMATION AND SKINFOLD TECHNIQUES
    STERNER, MTG
    BURKE, EJ
    [J]. PHYSICIAN AND SPORTSMEDICINE, 1986, 14 (04): : 101 - &