Comparison of Visual Assessment Techniques for Wear and Corrosion in Modular Hip Replacement Systems

被引:5
|
作者
Di Prima, Matthew A. [1 ]
Vesnovsky, Oleg [1 ]
Kovacs, Paul
Hopper, Robert H., Jr. [2 ]
Ho, Henry [2 ]
Engh, Charles A., Jr. [2 ]
Topoleski, L. D. Timmie [1 ,3 ]
机构
[1] US FDA, Ctr Device & Radiol Hlth, 10903 New Hampshire Ave, Silver Spring, MD 20993 USA
[2] Anderson Orthopaed Res Inst, Alexandria, VA 22306 USA
[3] Univ Maryland Baltimore Cty, Dept Mech Engn, Baltimore, MD 21250 USA
关键词
metal-on-metal; corrosion; orthopaedic; wear; modular; hip arthroplasty; FRETTING CORROSION; TAPER JUNCTION; INTERFACE; FAILURE; HEAD; ARTHROPLASTY; RETRIEVAL; CREVICE;
D O I
10.1520/STP159120140152
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Wear and corrosion of implant metals have been associated with adverse tissue reactions and failures of total hips. However, the factors leading to component wear and corrosion, as well as the clinical effects of wear and corrosion, are not well understood. A visual inspection standard for taper corrosion and fretting, the Goldberg method, currently is used to quantify the amount of apparent corrosion damage present on an implant using a subjective score. We developed an approach to quantify the damage by creating digital microscopic images of the entire taper and trunnion surfaces, which enables an objective quantification of detailed wear and corrosion damage of the contacting surfaces, which we call the digital mosaic method (DMM). We compared the results of this new methodology to the Goldberg method and Anderson method (an expanded version of the Goldberg method). We examined the modular surfaces of ten retrieved metal-on-metal hip replacement systems. Each set of taper and trunnion surfaces were imaged and stitched into a composite image of the entire taper-trunnion surface. Three independent observers measured the surface area fraction of damage (wear or corrosion) for each surface, which in turn was compared with visual scores performed by independent observers on the same retrieved joints. To determine a measure of the intraobserver repeatability of the new method, a single reviewer measured the surface fraction of damage for the same set of explanted surfaces across different days, blinded to any previous specimen identification, with the image order randomized. The results of this study show that the DMM damage quantification correlates to the current visual scoring approaches (Goldberg and Anderson). The potential advantages to using this approach include better damage quantification, detailed analysis of specific locations, ability to identify matching damage locations on the two surfaces, and higher resolution of the contact area.
引用
收藏
页码:147 / 163
页数:17
相关论文
共 50 条
  • [1] WEAR AND CORROSION OF MODULAR INTERFACES IN TOTAL HIP REPLACEMENTS
    COOK, SD
    BARRACK, RL
    BAFFES, GC
    CLEMOW, AJT
    SEREKIAN, P
    DONG, N
    KESTER, MA
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1994, (298) : 80 - 88
  • [2] Enhanced Wear and Corrosion in Modular Tapers in Total Hip Replacement Is Associated With the Contact Area and Surface Topography
    Panagiotidou, Anna
    Meswania, Jay
    Hua, Jia
    Muirhead-Allwood, Sarah
    Hart, Alister
    Blunn, Gordon
    [J]. JOURNAL OF ORTHOPAEDIC RESEARCH, 2013, 31 (12) : 2032 - 2039
  • [3] Assessment of polyethylene wear in total hip replacement
    Devane, PA
    Horne, JG
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1999, (369) : 59 - 72
  • [4] Corrosion of modular titanium alloy stems in cementless hip replacement
    Urban, Robert M.
    Gilbert, Jeremy L.
    Jacobs, Joshua J.
    [J]. TITANIUM, NIOBIUM, ZIRCONIUM, AND TANTALUM FOR MEDICAL AND SURGICAL APPLICATIONS, 2006, 1471 : 215 - +
  • [5] Total Hip Wear Assessment: A Comparison Between Computational and In Vitro Wear Assessment Techniques Using ISO 14242 Loading and Kinematics
    Matsoukas, George
    Willing, Ryan
    Kim, Il Yong
    [J]. JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2009, 131 (04):
  • [6] Radiographic assessment of cup migration and wear after hip replacement
    Ilchmann, T
    [J]. ACTA ORTHOPAEDICA SCANDINAVICA, 1997, 68 : 87 - 87
  • [7] Radiographic assessment of cup migration and wear after hip replacement
    Ilchmann, T
    [J]. ACTA ORTHOPAEDICA SCANDINAVICA, 1997, 68 : 2 - +
  • [8] Early Corrosion-Related Failure of the Rejuvenate Modular Total Hip Replacement
    Meftah, Morteza
    Haleem, Amgad M.
    Burn, Matthew B.
    Smith, Kevin M.
    Incavo, Stephen J.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2014, 96A (06): : 481 - 487
  • [9] Assessment of wear on the cones of modular stainless steel Exeter hip stems
    Chaplin, RPS
    Lee, AJC
    Hooper, RM
    [J]. JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 2004, 15 (09) : 977 - 990
  • [10] Assessment of wear on the cones of modular stainless steel Exeter hip stems
    R. P. S. Chaplin
    A. J. C. Lee
    R. M. Hooper
    [J]. Journal of Materials Science: Materials in Medicine, 2004, 15 : 977 - 990