Total Hip Wear Assessment: A Comparison Between Computational and In Vitro Wear Assessment Techniques Using ISO 14242 Loading and Kinematics

被引:20
|
作者
Matsoukas, George [1 ]
Willing, Ryan [1 ]
Kim, Il Yong [1 ]
机构
[1] Queens Univ, Dept Mech & Mat Engn, Kingston, ON K7L 3N6, Canada
关键词
biomechanics; biomedical materials; digital simulation; kinematics; medical computing; polymers; prosthetics; stress-strain relations; wear testing; MOLECULAR-WEIGHT POLYETHYLENE; FINITE-ELEMENT-ANALYSIS; ACETABULAR COMPONENTS; ARTHROPLASTY; SIMULATION; IMPLANTS; UHMWPE; CREEP; PREDICTION; OSTEOLYSIS;
D O I
10.1115/1.3049477
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
In the present study a direct comparison was made between in vitro total hip wear testing and a computational analysis considering the effects of time and a nonlinear stress-strain relationship for ultrahigh molecular weight polyethylene (UHMWPE) at 37 degrees C. The computational simulation was made correct through calibration to experimental volumetric wear results, and the predicted damage layout on the acetabular liner surface was compared with results estimated from laser scanning of the actual worn specimens. The wear rates for the testing specimens were found to be 17.14 +/- 1.23 mg/10(6) cycles and 19.39 +/- 0.79 mg/10(6) cycles, and the cumulative volumetric wear values after 3x10(6) cycles were 63.70 mm(3) and 64.02 mm(3) for specimens 1 and 2, respectively. The value of the calibrated wear coefficient was found to be 5.32(10(-10)) mm(3)/N mm for both specimens. The major difference between the computational and experimental wear results was the existence of two damage vectors in the experimental case. The actual location of damage was virtually the same in both cases, and the maximum damage depth of the computational model agreed well with the experiment. The existence of multiple wear vectors may indicate the need for computational approaches to account for multidirectional sliding or strain hardening of UHMWPE. Despite the limitation in terms of describing the overall damage layout, the present computational model shows that simulation can mimic some of the behavior of in vitro wear.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Assessment of polyethylene wear in total hip replacement
    Devane, PA
    Horne, JG
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1999, (369) : 59 - 72
  • [2] In-vitro Wear Test and Computational Wear Prediction of Total Hip Replacement in China
    Yang, Shu
    Cui, Wen
    Zhang, Xiaogang
    Zhang, Yali
    Xie, Fengbao
    Ren, Xuhui
    Chen, Zhenxian
    Jin, Zhongmin
    [J]. Mocaxue Xuebao/Tribology, 2021, 41 (06): : 1004 - 1018
  • [3] Comparative analysis of surface wear from total hip prostheses tested on a mechanical simulator according to standards ISO 14242-1 and ISO 14242-3
    Oliveira, A. L. L.
    Lima, R. G.
    Cueva, E. G.
    Queiroz, R. D.
    [J]. WEAR, 2011, 271 (9-10) : 2340 - 2345
  • [4] Comparison of Visual Assessment Techniques for Wear and Corrosion in Modular Hip Replacement Systems
    Di Prima, Matthew A.
    Vesnovsky, Oleg
    Kovacs, Paul
    Hopper, Robert H., Jr.
    Ho, Henry
    Engh, Charles A., Jr.
    Topoleski, L. D. Timmie
    [J]. MODULARITY AND TAPERS IN TOTAL JOINT REPLACEMENT DEVICES, 2015, 1591 : 147 - 163
  • [5] Modification of a Simplified Hip Joint Simulator Into an ISO 14242-1 Compliant Design and a Comparison of Wear Test Results
    Saikko, Vesa
    Morad, Omar
    Viitala, Raine
    [J]. JOURNAL OF TRIBOLOGY-TRANSACTIONS OF THE ASME, 2022, 144 (05):
  • [6] TECHNIQUES FOR ASSESSMENT OF WEAR BETWEEN HUMAN CARTILAGE SURFACES
    Verberne, Gabi
    Merkher, Yulia
    Halperin, Gregory
    Maroudas, Alice
    Etsion, Izhak
    [J]. PROCEEDINGS OF THE 9TH BIENNIAL CONFERENCE ON ENGINEERING SYSTEMS DESIGN AND ANALYSIS - 2008, VOL 3, 2009, : 399 - 403
  • [7] Techniques for assessment of wear between human cartilage surfaces
    Verberne, G.
    Merkher, Y.
    Halperin, G.
    Maroudas, A.
    Etsion, I.
    [J]. WEAR, 2009, 266 (11-12) : 1216 - 1223
  • [8] Arthroscopic assessment of total hip replacement and polyethylene wear: a case report
    Fontana, A
    Zecca, M
    Sala, C
    [J]. KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2000, 8 (04) : 244 - 245
  • [9] A radiological and laboratory assessment of polyethylene wear in failed total hip replacement
    Bisbinas, I
    Tripsianis, G
    Cunningham, JL
    Learmonth, ID
    [J]. HIP INTERNATIONAL, 2006, 16 (01) : 39 - 46
  • [10] Arthroscopic assessment of total hip replacement and polyethylene wear: a case report
    A. Fontana
    M. Zecca
    C. Sala
    [J]. Knee Surgery, Sports Traumatology, Arthroscopy, 2000, 8 : 244 - 245