Energy filtering with X-ray lenses: optimization for photon-counting mammography

被引:0
|
作者
Fredenberg, Erik [1 ]
Cederstrom, Bjorn [1 ]
Danielsson, Mats [1 ]
机构
[1] Royal Inst Technol, Dept Phys, SE-10691 Stockholm, Sweden
基金
瑞典研究理事会;
关键词
D O I
10.1093/rpd/ncq049
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Chromatic properties of the multi-prism and prism-array X-ray lenses (MPL and PAL) can potentially be utilized for efficient energy filtering and dose reduction in mammography. The line-shaped foci of the lenses are optimal for coupling to photon-counting silicon strip detectors in a scanning system. A theoretical model was developed and used to investigate the benefit of two lenses compared with an absorption-filtered reference system. The dose reduction of the MPL filter was similar to 15% compared with the reference system at matching scan time, and the spatial resolution was higher. The dose of the PAL-filtered system was found to be similar to 20% lower than for the reference system at equal scan time and resolution, and only similar to 20% higher than for a monochromatic beam. An investigation of some practical issues remains, including the feasibility of brilliant-enough X-ray sources and manufacturing of a polymer PAL.
引用
收藏
页码:339 / 342
页数:4
相关论文
共 50 条
  • [1] Photon-counting x-ray detectors for CT
    Danielsson, Mats
    Persson, Mats
    Sjolin, Martin
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (03):
  • [2] An X-Ray Fluorescence Technique for Energy Calibration of Photon-Counting Detectors
    Ding, H.
    Cho, H.
    Barber, W.
    Iwanczyk, J.
    Molloi, S.
    MEDICAL PHYSICS, 2014, 41 (06) : 476 - 476
  • [3] Evaluation of a photon-counting X-ray imaging system
    Lundqvist, M
    Cederström, B
    Chmill, V
    Danielsson, M
    Hasegawa, B
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2001, 48 (04) : 1530 - 1536
  • [4] MicroComputed tomography with a photon-counting X-ray detector
    Frey, Ec.
    Taguchi, K.
    Kapusta, M.
    Xu, J.
    Orskaug, T.
    Ninive, I.
    Wagenaar, D.
    Patt, B.
    Tsui, B. M. W.
    MEDICAL IMAGING 2007: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2007, 6510
  • [5] Modeling Spectral Distortions in Energy Resolved Photon-counting X-ray Detector
    Wang, Xiaolan
    Meier, Dirk
    Hugg, James
    Chowdhury, Samir
    Wagenaar, Douglas
    Patt, Bradley
    Frey, Eric
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 3054 - 3057
  • [6] Projection-based energy weighting on photon-counting X-ray images in digital subtraction mammography: a feasibility study
    Choi, Sung-Hoon
    Lee, Seung-Wan
    Choi, Yu-Na
    Lee, Young-Jin
    Kim, Hee-Joung
    MEDICAL IMAGING 2014: PHYSICS OF MEDICAL IMAGING, 2014, 9033
  • [7] Material separation in x-ray CT with energy resolved photon-counting detectors
    Wang, Xiaolan
    Meier, Dirk
    Taguchi, Katsuyuki
    Wagenaar, Douglas J.
    Patt, Bradley E.
    Frey, Eric C.
    MEDICAL PHYSICS, 2011, 38 (03) : 1534 - 1546
  • [8] Material Separation in X-ray CT with Energy Resolved Photon-Counting Detectors
    Wang, X.
    Meier, D.
    Taguchi, K.
    Wagenaar, D. J.
    Patt, B. E.
    Frey, E. C.
    MEDICAL IMAGING 2011: PHYSICS OF MEDICAL IMAGING, 2011, 7961
  • [9] Characterization of energy response for photon-counting detectors using x-ray fluorescence
    Ding, Huanjun
    Cho, Hyo-Min
    Barber, William C.
    Iwanczyk, Jan S.
    Molloi, Sabee
    MEDICAL PHYSICS, 2014, 41 (12)
  • [10] Noise in energy-discriminating photon-counting x-ray imaging detectors
    Tanguay, Jesse
    Kim, Ho Kyung
    Cunningham, Ian A.
    MEDICAL IMAGING 2010: PHYSICS OF MEDICAL IMAGING, 2010, 7622