Validity and reliability of the ankle-brachial index by oscillometric blood pressure and automated ankle-brachial index

被引:8
|
作者
Chongthawonsatid, Sukanya [1 ]
Dutsadeevettakul, Somchai [2 ]
机构
[1] Mahidol Univ, Fac Social Sci & Humanities, Dept Social Sci, Salaya 73170, Nakhon Pathom, Thailand
[2] Mahidol Univ, Dept Med, Golden Jubilee Med Ctr, Salaya 73170, Nakhon Pathom, Thailand
关键词
Automated ankle-brachial index; oscillometric blood pressure; peripheral arterial disease; reliability; validity; PERIPHERAL ARTERIAL-DISEASE; VASCULAR-DISEASE; DOPPLER METHOD; DEVICE; STANDARDS;
D O I
10.4103/jrms.JRMS_728_16
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: This study was conducted to assess the validity and reliability of ankle-brachial index (ABI) by oscillometric blood pressure (BP) measurement as compared with an automated ABI as a gold standard. Materials and Methods: This study was conducted at Golden Jubilee Medical Center, Thailand. All the data were collected from 303 patients at risk of peripheral arterial disease (PAD) who were 45 years of age or above and who underwent treatment at the outpatient medical clinic between June and December 2015. Patients who were followed up at the medical clinic had both oscillometric BP measurement (Terumo, ES-P600) and an automated ABI (oscillometric method) at the same time. Sensitivity, specificity, positive predictive value, and negative predictive value of the oscillometric BP measurements to predict an abnormal ABI (< 0.90) were determined using the automated ABI as the gold standard. Results: ABI values were similar between the two methods. The oscillometric BP method for determining an ABI (cutoff point < 0.90) on the right side had a sensitivity of 88.89%, specificity of 99.32%, an accuracy of 99.01%, a positive predictive value of 80%, and a negative predictive value of 99.32% whereas ABI on the left side had a sensitivity of 69.23%, a specificity of 99.66%, an accuracy of 98.35%, a positive predictive value of 90%, and a negative predictive value of 98.63%. Reliability of the oscillometric BP method by Kappa statistics was 0.84 on the right side and 0.77 on the left side (P = 0.000). Conclusion: The oscillometric BP method is a reliable and useful alternative to conventional automated ABI determination in patients with no severe arterial occlusion for estimation of the prevalence and screening of PAD in primary health-care settings.
引用
收藏
页数:5
相关论文
共 50 条
  • [21] Ankle-brachial index in HIV infection
    Julián Olalla
    Daniel Salas
    Javier de la Torre
    Alfonso del Arco
    José Luis Prada
    Francisco Martos
    Emilio Perea-Milla
    Javier García-Alegría
    [J]. AIDS Research and Therapy, 6 (1)
  • [22] Ankle-brachial index in dementia.
    Laurin, D.
    Masaki, K.
    White, L. R.
    Launer, L. J.
    [J]. AMERICAN JOURNAL OF EPIDEMIOLOGY, 2006, 163 (11) : S239 - S239
  • [23] Chinese Ankle-Brachial Index Response
    Xing, Jin
    [J]. ANGIOLOGY, 2013, 64 (03) : 243 - 243
  • [24] Association of incident cardiovascular events with unilateral low ankle-brachial index and bilateral low ankle-brachial index
    Zhao, Song
    Xu, Chong
    Xiong, Jing
    Xu, Yawei
    Zhang, Yi
    [J]. EUROPEAN JOURNAL OF INTERNAL MEDICINE, 2024, 120 : 142 - 144
  • [25] Ankle-brachial Index as a Clinical Tool
    Ghannam, Michael
    Rodriguez, Yasser
    Ehrmann, Daniel
    Grossman, Paul M.
    [J]. AMERICAN JOURNAL OF MEDICINE, 2012, 125 (12): : 1171 - 1172
  • [26] Postexercise Ankle-Brachial Index Testing
    Mehta, Anurag
    Sperling, Laurence S.
    Wells, Bryan J.
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2020, 324 (08): : 796 - 797
  • [27] Diagnostic accuracy of automated oscillometric determination of the ankle-brachial index in peripheral artery disease
    Hageman, David
    van den Houten, Marijn M. L.
    Pesser, Niels
    Gommans, Lindy N. M.
    Scheltinga, Marc R. M.
    Teijink, Joep A. W.
    [J]. JOURNAL OF VASCULAR SURGERY, 2021, 73 (02) : 652 - 660
  • [28] Automated oscillometric determination of the ankle-brachial index: a systematic review and meta-analysis
    Verberk, Willem J.
    Kollias, Anastasios
    Stergiou, George S.
    [J]. HYPERTENSION RESEARCH, 2012, 35 (09) : 883 - 891
  • [29] Interpretation and significance of ankle-brachial systolic pressure index
    Ko, Sae Hee
    Bandyk, Dennis F.
    [J]. SEMINARS IN VASCULAR SURGERY, 2013, 26 (2-3) : 86 - 94
  • [30] Automated oscillometric determination of the ankle-brachial index: a systematic review and meta-analysis
    Willem J Verberk
    Anastasios Kollias
    George S Stergiou