3-D geometry calibration and markerless electromagnetic tracking with a mobile C-arm

被引:2
|
作者
Cheryauka, Arvi [1 ]
Barrett, Johnny [1 ]
Wang, Zhonghua [1 ]
Litvin, Andrew [2 ]
Hamadeh, Ali [2 ]
Beaudet, Daniel [2 ]
机构
[1] GE Healthcare Surg, 384 Wright Brothers Dr, Salt Lake City, UT 84116 USA
[2] GE Healthcare Surg, Lawrence, MA 01843 USA
关键词
mobile C-arm; calibration; motorized motion; electromagnetic tracking;
D O I
10.1117/12.711798
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The design of mobile X-ray C-arm equipment with image tomography and surgical guidance capabilities involves the retrieval of repeatable gantry positioning in three-dimensional space. Geometry misrepresentations can cause degradation of the reconstruction results with the appearance of blurred edges, image artifacts, and even false structures. It may also amplify surgical instrument tracking errors leading to improper implant placement. In our prior publications we have proposed a C-arm 3D positioner calibration method comprising separate intrinsic and extrinsic geometry calibration steps. Following this approach, in the present paper, we extend the intrinsic geometry calibration of C-gantry beyond angular positions in the orbital plane into angular positions on a unit sphere of isocentric rotation. Our method makes deployment of markerless interventional tool guidance with use of high-resolution fluoro images and electromagnetic tracking feasible at any angular position of the tube-detector assembly. Variations of the intrinsic parameters associated with C-arm motion are measured off-line as functions of orbital and lateral angles. The proposed calibration procedure provides better accuracy, and prevents unnecessary workflow steps for surgical navigation applications. With a slight modification, the Misalignment phantom, a tool for intrinsic geometry calibration, is also utilized to obtain an accurate 'image-to-sensor' mapping. We show simulation results, image quality and navigation accuracy estimates, and feasibility data acquired with the prototype system. The experimental results show the potential of high-resolution CT imaging (voxel size below 0.5 mm) and confident navigation in an interventional surgery setting with a mobile C-arm.
引用
收藏
页数:8
相关论文
共 50 条
  • [21] Camera Augmented Mobile C-Arm (CAMC): Calibration, Accuracy Study, and Clinical Applications
    Navab, Nassir
    Heining, Sandro-Michael
    Traub, Joerg
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2010, 29 (07) : 1412 - 1423
  • [22] 3D-imaging using mobile low-end C-arm systems
    Lütjens, J
    Koppe, R
    Klotz, E
    Grass, M
    Rasche, V
    [J]. MEDICAL IMAGING 2002: IMAGE PROCESSING, VOL 1-3, 2002, 4684 : 901 - 912
  • [23] 3D-imaging of the facial skeleton with an isocentric mobile C-arm system (Siremobil ISO-C3D)
    Heiland, M
    Schulze, D
    Adam, G
    Schmelzle, R
    [J]. DENTOMAXILLOFACIAL RADIOLOGY, 2003, 32 (01) : 21 - 25
  • [24] RGB-D/C-arm Calibration and Application in Medical Augmented Reality
    Wang, Xiang
    Habert, Severine
    Ma, Meng
    Huang, Chun-Hao
    Fallavollita, Pascal
    Navab, Nassir
    [J]. 2015 IEEE INTERNATIONAL SYMPOSIUM ON MIXED AND AUGMENTED REALITY, 2015, : 100 - 103
  • [25] Hysterosalpingography with use of mobile C-arm fluoroscopy
    Phillips, Jordana
    Cochavi, Saadia
    Silberzweig, James E.
    [J]. FERTILITY AND STERILITY, 2010, 93 (06) : 2065 - 2068
  • [26] CT Imaging with a mobile C-arm prototype
    Cheryauka, Arvi
    Tubbs, Dave
    Langille, Vinton
    Kalya, Prabhanjana
    Smith, Brady
    Cherone, Rocco
    [J]. MEDICAL IMAGING 2008: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2008, 6913
  • [27] Optimal configuration for dynamic calibration of projection geometry of X-ray C-arm systems
    Mitschke, MM
    Navab, N
    [J]. IEEE WORKSHOP ON MATHEMATICAL METHODS IN BIOMEDICAL IMAGE ANALYSIS, PROCEEDINGS, 2000, : 204 - 209
  • [28] 3D-imaging of the facial skeleton with an isocentric mobile C-arm system (Siremobil Iso-C3D)
    Schulze, D
    Heiland, M
    Adam, G
    Schmelzle, R
    Rother, U
    [J]. RADIOLOGY, 2002, 225 : 565 - 565
  • [29] Fusion Imaging for EVAR with Mobile C-arm
    Kaladji, Adrien
    Villena, Alexandre
    Pascot, Remy
    Lalys, Florent
    Daoudal, Anne
    Clochard, Elodie
    Lucas, Antoine
    Cardon, Alain
    [J]. ANNALS OF VASCULAR SURGERY, 2019, 55 : 166 - 174
  • [30] Precise 3D/2D calibration between a RGB-D sensor and a C-arm fluoroscope
    Xiang Wang
    Severine Habert
    Meng Ma
    Chun-Hao Huang
    Pascal Fallavollita
    Nassir Navab
    [J]. International Journal of Computer Assisted Radiology and Surgery, 2016, 11 : 1385 - 1395