Clinical application of a model-based cardiac stroke volume estimation method

被引:1
|
作者
Smith, Rachel [1 ]
Balmer, Joel [1 ]
Pretty, Christopher G. [1 ]
Shaw, Geoffrey M. [2 ]
Chase, J. Geoffrey [1 ]
机构
[1] Univ Canterbury, Dept Mech Engn, Canterbury, New Zealand
[2] Christchurch Hosp Intens Care Unit, Christchurch, New Zealand
来源
IFAC PAPERSONLINE | 2020年 / 53卷 / 02期
基金
欧盟地平线“2020”;
关键词
Pulse contour analysis; Pressure contour analysis; Windkessel model; Stroke volume; Cardiac output; Hemodynamic monitoring; Intensive care; PRESSURE; OUTPUT; PULSE; ANESTHESIA; ACCURACY; SHOCK;
D O I
10.1016/j.ifacol.2020.12.435
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
A system is needed for monitoring stroke volume (SV) and cardiac output (CO) in unstable patients which is non-additionally invasive, reproducible and reliable in a variety of physiological states. This study evaluates SV estimation accuracy of a non-additionally invasive pulse contour analysis method implemented using a 3-element Windkessel model. The model lumps the properties of the arterial system into 3 parameters: characteristic impedance of the proximal aorta (Z), and resistance (R) and compliance (C) of the systemic arteries. Parameter products ZC and RC are dynamically identified from measured femoral arterial pressure waveforms, and Z is a static parameter obtained by calibration. The accuracy of the model is evaluated for a cohort of 9 liver transplant patients, using thermodilution as a reference method. Data were obtained from Vital Data Bank (VitalDB). The study thus provides independent assessment of a pulse contour analysis, proven in animal studies, in an uncontrolled clinical environment. The model tracked trends in SV well over the course of the surgery. However, the 95% range for percentage error was -88% to +53%, outside acceptable limits of +/- 45%. Main areas contributing to error for the model include the changing extent of reflected waves in the arterial system, dynamic response characteristics of fluid-filled pressure catheters, and the assumption of fixed Z parameter. Further investigation is needed to consider the contribution of these factors to SV estimation error by the model. Copyright (C) 2020 The Authors.
引用
收藏
页码:16137 / 16142
页数:6
相关论文
共 50 条
  • [1] Incorporating pulse wave velocity into model-based pulse contour analysis method for estimation of cardiac stroke volume
    Smith, Rachel
    Balmer, Joel
    Pretty, Christopher G.
    Mehta-Wilson, Tashana
    Desaive, Thomas
    Shaw, Geoffrey M.
    Chase, J. Geoffrey
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2020, 195
  • [2] Clinically applicable model-based method, for physiologically accurate flow waveform and stroke volume estimation
    Balmer, Joel
    Pretty, Christopher G.
    Davidson, Shaun
    Mehta-Wilson, Tashana
    Desaive, Thomas
    Smith, Rachel
    Shaw, Geoffrey M.
    Chase, J. Geoffrey
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2020, 185
  • [3] EVALUATION OF A MODEL-BASED ESTIMATION METHOD FOR DYNAMIC CARDIAC STUDIES
    CHIAO, P
    MUZIK, O
    ROGERS, WL
    JOURNAL OF NUCLEAR MEDICINE, 1994, 35 (05) : P185 - P185
  • [4] Model-based Estimation of Ventricular Cerebrospinal Fluid Volume
    Flurenbrock, Fabian
    Muntwiler, Simon
    Korn, Leonie
    Daners, Marianne Schmid
    Zeilinger, Melanie N.
    2024 IEEE CONFERENCE ON CONTROL TECHNOLOGY AND APPLICATIONS, CCTA 2024, 2024, : 302 - 309
  • [6] Maternal mortality estimation at the subnational level: a model-based method with an application to Bangladesh
    Ahmed, Saifuddin
    Hill, Kenneth
    BULLETIN OF THE WORLD HEALTH ORGANIZATION, 2011, 89 (01) : 12 - 21
  • [7] Model-Based Estimation of the Cardiac Output in Exercise Scenario
    Dan, Ana Maria
    2016 12TH IEEE INTERNATIONAL SYMPOSIUM ON ELECTRONICS AND TELECOMMUNICATIONS (ISETC'16), 2016, : 285 - 288
  • [8] A model-based method for myocardium flow estimation
    Santarelli, M. F.
    Landini, L.
    Lombardi, M.
    Positano, V.
    L'Abbate, A.
    Benassi, A.
    MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2000, 11 (01) : 87 - 88
  • [9] A model-based method for myocardium flow estimation
    M. F. Santarelli
    L. Landini
    M. Lombardi
    V. Positano
    A. L’Abbate
    A. Benassi
    Magma: Magnetic Resonance Materials in Physics, Biology, and Medicine, 2000, 11 (1-2): : 87 - 88
  • [10] Model-based Clinical Assist System for Cardiac Ablation
    Wu, Yutong
    Tang, Renzhi
    Kang, Eunsuk
    Jiang, Zhihao
    ICCPS'21: PROCEEDINGS OF THE 2021 ACM/IEEE 12TH INTERNATIONAL CONFERENCE ON CYBER-PHYSICAL SYSTEMS (WITH CPS-IOT WEEK 2021), 2021, : 110 - 120