Calibration and validation of confocal spectral imaging systems

被引:41
|
作者
Lerner, JM
Zucker, RM
机构
[1] LightForm Inc, Hillsborough, NJ 08844 USA
[2] US EPA, Reprod Toxicol Div, Natl Hlth & Environm Effects Res Lab, Res Triangle Pk, NC 27711 USA
关键词
confocal microscope; wavelength calibration; spectral calibration; spectral imaging; Zeiss; Leica; SKY; PARISS; quality assurance; validation; photomultiplier tube;
D O I
10.1002/cyto.a.20087
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Background: Confocal spectral imaging (CSI) microscopic systems currently on the market delineate multiple fluorescent proteins, labels, or dyes within biological specimens by performing spectral characterizations. However, some CSI systems have been found to present inconsistent spectral profiles of reference spectra within a particular system and between related and unrelated instruments. This variability confirms that there is a need for a standardized, objective calibration and validation protocol. Methods: Our protocol uses an inexpensive multi-ion discharge lamp (MIDL) that contains Hg+, Ar+, and inorganic fluorophores that emit distinct, stable, spectral features in place of a sample. We derived reference spectra from the MIDL data to accurately predict the spectral resolution, ratio of wavelength to wavelength, contrast, and aliasing parameters of any CSI system. We were also able to predict and confirm the influence of pinhole diameter on spectral profiles. Results: Using this simulation, we determined that there was good agreement between observed and theoretical expectations, thus enabling us to identify malfunctioning subsystems. We examined eight CSI systems and one non-confocal spectral system, all of which displayed spectral inconsistencies. No instrument met its optimal performance expectations. In two systems, we established the need for factory realignment that had not been otherwise recognized. Conclusions: We found that using a primary light source that emits an absolute standard "reference spectrum" enabled us to diagnose instrumental errors and measure accuracy and reproducibility under normalized conditions. With this information, a CSI operator can determine whether a CSI system is working optimally and make objective comparisons with the performance of other CSI systems. We determined that, if CSI systems were standardized to produce the same spectral profile of a MIDL lamp, researchers could be confident that the same experimental findings would be obtained on any CSI system. (C) 2004 Wiley-Liss, Inc.
引用
收藏
页码:8 / 34
页数:27
相关论文
共 50 条
  • [21] SIZE OF THE DETECTOR IN CONFOCAL IMAGING-SYSTEMS
    WILSON, T
    CARLINI, AR
    OPTICS LETTERS, 1987, 12 (04) : 227 - 229
  • [22] THE ROLE OF THE PINHOLE IN CONFOCAL IMAGING-SYSTEMS
    WILSON, T
    HANDBOOK OF BIOLOGICAL CONFOCAL MICROSCOPY, REVISED EDITION, 1989, : 113 - 126
  • [23] 3-DIMENSIONAL IMAGING IN CONFOCAL SYSTEMS
    WILSON, T
    JOURNAL OF MICROSCOPY-OXFORD, 1989, 153 : 161 - 169
  • [24] Deep Spectral Imaging (DSI). An optimized AI based spectral imaging acquisition, transmission and calibration
    Melis, Marcello
    Saggese, Enrico
    Muzi, Marco
    Fois, Matteo
    Scalabrin, Damiano
    SENSORS, SYSTEMS, AND NEXT-GENERATION SATELLITES XXVII, 2023, 12729
  • [25] Onboard spectral calibration and validation of the satellite calibration spectrometer on HY-1C
    Xu, Heyu
    Huang, Wenxin
    Si, Xiaolong
    Li, Xin
    Xu, Weiwei
    Zhang, Liming
    Song, Qingjun
    Gao, Huiting
    OPTICS EXPRESS, 2022, 30 (15) : 27645 - 27661
  • [26] Calibration of MAJIS (Moons And Jupiter Imaging Spectrometer). III. Spectral calibration
    Haffoud, Paolo
    Poulet, Francois
    Vincendon, Mathieu
    Filacchione, Gianrico
    Barbis, Alessandra
    Guiot, Pierre
    Lecomte, Benoit
    Langevin, Yves
    Piccioni, Giuseppe
    Dumesnil, Cydalise
    Rodriguez, Sebastien
    Carter, John
    Stefania, Stefani
    Tommasi, Leonardo
    Tosi, Federico
    Pilorget, Cedric
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2024, 95 (03):
  • [27] An Algorithm for In-Flight Spectral Calibration of Imaging Spectrometers
    Kuhlmann, Gerrit
    Hueni, Andreas
    Damm, Alexander
    Brunner, Dominik
    REMOTE SENSING, 2016, 8 (12):
  • [28] Spectral Calibration of The Coded Aperture Spectra Imaging System
    Wang Jianwei
    Min Huang
    Pei Lin Lin
    Lv Qunbo
    Liu Yangyang
    INTERNATIONAL SEMINAR ON PHOTONICS, OPTICS, AND ITS APPLICATIONS (ISPHOA 2014), 2015, 9444
  • [29] An efficient calibration method for multi-spectral imaging
    Ma, Cui
    Lin, Hui
    Zhang, Guodong
    Du, Ruxu
    OPTICS COMMUNICATIONS, 2018, 420 : 14 - 25
  • [30] Prismatic imaging polarimeter calibration for the infrared spectral region
    Kudenov, Michael W.
    Pezzaniti, Larry
    Dereniak, Eustace L.
    Gerhart, Grant R.
    OPTICS EXPRESS, 2008, 16 (18) : 13720 - 13737