Calibration and validation of confocal spectral imaging systems

被引:41
|
作者
Lerner, JM
Zucker, RM
机构
[1] LightForm Inc, Hillsborough, NJ 08844 USA
[2] US EPA, Reprod Toxicol Div, Natl Hlth & Environm Effects Res Lab, Res Triangle Pk, NC 27711 USA
关键词
confocal microscope; wavelength calibration; spectral calibration; spectral imaging; Zeiss; Leica; SKY; PARISS; quality assurance; validation; photomultiplier tube;
D O I
10.1002/cyto.a.20087
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Background: Confocal spectral imaging (CSI) microscopic systems currently on the market delineate multiple fluorescent proteins, labels, or dyes within biological specimens by performing spectral characterizations. However, some CSI systems have been found to present inconsistent spectral profiles of reference spectra within a particular system and between related and unrelated instruments. This variability confirms that there is a need for a standardized, objective calibration and validation protocol. Methods: Our protocol uses an inexpensive multi-ion discharge lamp (MIDL) that contains Hg+, Ar+, and inorganic fluorophores that emit distinct, stable, spectral features in place of a sample. We derived reference spectra from the MIDL data to accurately predict the spectral resolution, ratio of wavelength to wavelength, contrast, and aliasing parameters of any CSI system. We were also able to predict and confirm the influence of pinhole diameter on spectral profiles. Results: Using this simulation, we determined that there was good agreement between observed and theoretical expectations, thus enabling us to identify malfunctioning subsystems. We examined eight CSI systems and one non-confocal spectral system, all of which displayed spectral inconsistencies. No instrument met its optimal performance expectations. In two systems, we established the need for factory realignment that had not been otherwise recognized. Conclusions: We found that using a primary light source that emits an absolute standard "reference spectrum" enabled us to diagnose instrumental errors and measure accuracy and reproducibility under normalized conditions. With this information, a CSI operator can determine whether a CSI system is working optimally and make objective comparisons with the performance of other CSI systems. We determined that, if CSI systems were standardized to produce the same spectral profile of a MIDL lamp, researchers could be confident that the same experimental findings would be obtained on any CSI system. (C) 2004 Wiley-Liss, Inc.
引用
收藏
页码:8 / 34
页数:27
相关论文
共 50 条
  • [1] Spectral characteristics of chromatic confocal imaging systems
    Hillenbrand, Matthias
    Mitschunas, Beate
    Brill, Florian
    Grewe, Adrian
    Sinzinger, Stefan
    APPLIED OPTICS, 2014, 53 (32) : 7634 - 7642
  • [2] Calibration and characterization of spectral imaging systems
    Polder, G
    van der Heijden, GWAM
    MULTISPECTRAL AND HYPERSPECTRAL IMAGE ACQUISITION AND PROCESSING, 2001, 4548 : 10 - 17
  • [3] Wavelength and alignment tests for confocal spectral imaging systems
    Zucker, RM
    Lerner, JM
    MICROSCOPY RESEARCH AND TECHNIQUE, 2005, 68 (05) : 307 - 319
  • [4] New approach for the radiometric calibration of spectral imaging systems
    Kohler, DDR
    Bissett, WP
    Steward, RG
    Davis, CO
    OPTICS EXPRESS, 2004, 12 (11): : 2463 - 2477
  • [5] Reliability of confocal microscopy spectral imaging systems: Use of multispectral beads
    Zucker, Robert M.
    Rigby, Paul
    Clements, Ian
    Salmon, Wendy
    Chua, Michael
    CYTOMETRY PART A, 2007, 71A (03) : 174 - 189
  • [6] Spectral characterization and calibration of AOTF spectrometers and hyper-spectral imaging systems
    Katrasnik, Jaka
    Buermen, Miran
    Pernus, Franjo
    Likar, Bostjan
    CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 2010, 101 (01) : 23 - 29
  • [7] Spectral characterization and calibration of AOTF spectrometers and hyper-spectral imaging systems
    Katrasnik, Jaka
    Pernus, Franjo
    Likar, Bostjan
    DESIGN AND QUALITY FOR BIOMEDICAL TECHNOLOGIES III, 2010, 7556
  • [8] Imaging and Confocal Systems for in vivo Measurements of Human-Iris Spectral Reflectance
    Chen, Yi-Chun
    Jiang, Chong-Jhih
    Huang, Jian-Siang
    Chen, Szu-Yu
    Sun, Ching-Cherng
    JOURNAL OF DISPLAY TECHNOLOGY, 2014, 10 (07): : 595 - 600
  • [9] A practical approach to spectral calibration of short wavelength infrared hyper-spectral imaging systems
    Buermen, Miran
    Pernus, Franjo
    Likara, Bostjan
    DESIGN AND QUALITY FOR BIOMEDICAL TECHNOLOGIES III, 2010, 7556
  • [10] Laboratory Spectral Calibration and Radiometric Calibration of Hyper-spectral Imaging Spectrometer
    Yu, Xiujuan
    Sun, Yanhua
    Fang, Aiping
    Qi, Weijun
    Liu, Caijuan
    2014 2ND INTERNATIONAL CONFERENCE ON SYSTEMS AND INFORMATICS (ICSAI), 2014, : 871 - 875