A large area plastic scintillation detector with 4-corner-readout

被引:5
|
作者
Tang, Shu-Wen [1 ]
Yu, Yu-Hong [1 ]
Zhou, Yong [1 ,2 ]
Sun, Zhi-Yu [1 ]
Zhang, Xue-Heng [1 ]
Wang, Shi-Tao [1 ]
Yue, Ke [1 ]
Liu, Long-Xiang [1 ]
Fang, Fang [1 ]
Yan, Duo [1 ,2 ]
Sun, Yu [1 ,2 ]
Wang, Zhao-Min [1 ,2 ]
机构
[1] Chinese Acad Sci, Inst Modern Phys, Lanzhou 730000, Peoples R China
[2] Univ Chinese Acad Sci, Beijing 100049, Peoples R China
基金
中国国家自然科学基金;
关键词
scintillation detector; 4-corner-readout; time resolution; position resolution; HIGH-RESOLUTION; TIME;
D O I
10.1088/1674-1137/40/5/056001
中图分类号
O57 [原子核物理学、高能物理学];
学科分类号
070202 ;
摘要
A 760 mm x 760 mm x 30 mm plastic scintillation detector viewed by photomultiplier tubes (PMTs) from four corners has been developed, and the detector has been tested with cosmic rays and gamma rays. A position independent effective time T-eff has been found, indicating this detector can be used as a TOF detector. The hit position can also be reconstructed by the time from the four corners. A TOF resolution of 236 ps and a position resolution of 48 mm have been achieved, and the detection efficiency has also been investigated.
引用
收藏
页数:5
相关论文
共 50 条
  • [11] High efficiency plastic scintillator detector with wavelength-shifting fiber readout for the GLAST Large Area Telescope
    Moiseev, A. A.
    Deering, P. L.
    Hartman, R. C.
    Johnson, T. E.
    Nebel, T. R.
    Ormes, J. F.
    Thompson, D. J.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2007, 583 (2-3): : 372 - 381
  • [12] Readout for a large area neutron sensitive microchannel plate detector
    Wang, Yiming
    Yang, Yigang
    Wang, Xuewu
    Li, Yuanjing
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2015, 784 : 226 - 231
  • [13] ELECTRONICS AND READOUT OF A LARGE-AREA SILICON DETECTOR FOR LHC
    BORER, K
    MUNDAY, DJ
    PARKER, MA
    ANGHINOLFI, F
    ASPELL, P
    CAMPBELL, M
    CHILINGAROV, A
    JARRON, P
    HEIJNE, EHM
    SANTIARD, JC
    SCAMPOLI, P
    VERWEIJ, H
    GOSSLING, C
    LISOWSKI, B
    REICHOLD, A
    SPIWOKS, R
    TSESMELIS, E
    BENSLAMA, K
    BONINO, R
    CLARK, AG
    COUYOUMTZELIS, C
    KAMBARA, H
    WU, X
    FRETWURST, E
    LINDSTROEM, G
    SCHULTZ, T
    BARDOS, RA
    GORFINE, GW
    MOORHEAD, GF
    TAYLOR, GN
    TOVEY, SN
    BIBBY, JH
    HAWKINGS, RJ
    KUNDU, N
    WEIDBERG, A
    CAMPBELL, D
    MURRAY, P
    SELLER, P
    TEIGER, J
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 344 (01): : 185 - 193
  • [14] A large dynamic range readout design for the plastic scintillator detector of DAMPE
    Zhou, Yong
    Sun, Zhiyu
    Yu, Yuhong
    Zhang, Yongjie
    Fang, Fang
    Chen, Junling
    Hu, Bitao
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2016, 827 : 79 - 84
  • [15] Scintillation detector with the fiber-optical readout
    Ampilogov N.V.
    Amelchakov M.B.
    Britvich G.I.
    Brudanin V.B.
    Nemchenok I.B.
    Petrukhin A.A.
    Salamatin A.V.
    Soldatov A.P.
    Chernichenko S.K.
    Shein I.V.
    Yashin I.I.
    Bulletin of the Russian Academy of Sciences: Physics, 2009, 73 (05) : 637 - 639
  • [16] UNIFORMITY OF RESPONSE FROM LARGE AREA PLASTIC SCINTILLATION DETECTORS
    GILLESPIE, CR
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1968, 39 (11): : 1724 - +
  • [17] Secondary scintillation readout from GEM and THGEM with a large area avalanche photodiode
    Monteiro, C. M. B.
    Fernandes, L. M. P.
    Veloso, J. F. C. A.
    dos Santos, J. M. F.
    JOURNAL OF INSTRUMENTATION, 2012, 7
  • [18] Energy resolution of scintillation detectors readout with large area avalanche photodiodes and photomultipliers
    Soltan Inst for Nuclear Studies, Swierk-Otwock, Poland
    IEEE Trans Nucl Sci, 3 pt 1 (472-477):
  • [19] Energy resolution of scintillation detectors readout with large area avalanche photodiodes and photomultipliers
    Moszynski, M
    Kapusta, M
    Wolski, D
    Szawlowski, M
    Klamra, W
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1998, 45 (03) : 472 - 477
  • [20] Large area scintillation detector for dosimetric stand with improved light collection
    Machaj, Bronislaw
    Mirowicz, Jan
    Kowalska, Ewa