Interference of Thalassemias on Hemoglobin A1c Measurement by Ion-Exchange High-Performance Liquid Chromatography Method Tosoh HLC-723 G8

被引:2
|
作者
Chao, Yan [1 ]
Wan, Zemin [1 ]
Wu, Xiaobin [1 ]
Qiu, Feng [1 ]
Wang, Yunxiu [1 ]
Ke, Peifeng [1 ]
Wu, Xinzhong [1 ]
Xu, Jianhua [1 ]
Zhuang, Junhua [1 ]
Huang, Xianzhang [1 ]
机构
[1] Guangzhou Univ Chinese Med, Dept Lab Sci, Affiliated Hosp 2, Dade Rd 111, Guangzhou 510120, Guangdong, Peoples R China
基金
中国国家自然科学基金;
关键词
Hemoglobin A(1c) (HbA(1c)); thalassemia; ion-exchange high-performance liquid chromatography (IE-HPLC); BETA-THALASSEMIA; HBA(1C);
D O I
10.7754/Clin.Lab.2018.180119
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: The presence of hemoglobinopathies could interfere with some assays for Hemoglobin A(1c) (HbA(1c)) measurement; therefore, the effect of thalassemia on ion-exchange high-performance liquid chromatography (IEHPLC) method Tosoh HLC-723 G8 (Tosoh G8) was evaluated. Methods: A total of 43 normal controls and 101 thalassemia patients were quantified by Premier Hb9210 and Tosoh G8 (variant-mode) systems. At the same time, 7 normal controls and 8 thalassemia patients were confirmed by the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC) reference method for verification. Results: For normal controls, the HbA(1c) values of Tosoh G8 system (y) showed great correlation and agreement with those from Premier Hb9210 (x) (y = 0.9688x + 0.2151, r = 0.9951; mean difference 0.02 +/- 0.30%), and no significant relative bias above 7% was observed; the HbA(1c) values obtained by Tosoh G8 were consistent with the IFCC targets (relative bias < +/- 6%) in all of the samples. However, for thalassemia, the correlation between Tosoh G8 (y) and Premier Hb9210 (x) became relatively low (y = 0.8079x + 1.2897, r = 0.7780); the HbA ie values of 91.1% of the samples (92/101) obtained by Tosoh G8 were higher than those by Premier Hb9210 (mean difference 0.33 +/- 0.48%) and a significant positive bias above 7% was noticed in 43.3% (45/101) thalassemia patients; when compared with the IFCC targets, the 87.5% (7/8) relative bias was > +/- 6%. Conclusions: Thalassemia could directly affect the measurement of HbA(1c) using the IE-HPLC method Tosoh G8 and the clinical laboratorial staff should pay close attention.
引用
收藏
页码:1171 / 1175
页数:5
相关论文
共 50 条
  • [1] Evaluation of newly-developed glycated hemoglobin clinical analytic reagents and chromatography column on Tosoh HLC-723 G8 Analyzer
    Yuan, Bo
    Yang, Wei
    Zhang, Na
    Shi, Hongyan
    Dong, Shuangpeng
    PRACTICAL LABORATORY MEDICINE, 2023, 37
  • [2] Uncommon causes of hemoglobin E flags identified during measurement of hemoglobin A1c by ion-exchange high-performance liquid chromatography
    Herb, Thomas
    Taylor, Alexander S.
    Li, Shih-Hon
    Manthei, David M.
    Gherasim, Carmen
    LABORATORY MEDICINE, 2024, 55 (04) : 528 - 533
  • [3] Evaluation of the analytic performances of the new HPLC system HLC-723 G7 for the measurement of hemoglobin A1C
    Terreni, A
    Paleari, R
    Caldini, A
    Ognibene, A
    Mosca, A
    Messeri, G
    CLINICAL BIOCHEMISTRY, 2003, 36 (08) : 607 - 610
  • [4] HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY METHOD FOR HEMOGLOBIN A1C
    DAVIS, JE
    MCDONALD, JM
    JARETT, L
    DIABETES, 1978, 27 (02) : 102 - 107
  • [5] Hemoglobin S monitoring on TOSOH G8 in hemoglobin A1c mode in case of urgent red blood cell exchange
    Van Aelst, Sophie
    Claerhout, Helena
    Nackers, Elke
    Desmet, Koen
    Kieffer, Davy
    JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2018, 32 (07)
  • [6] Performance of the TOSOH A1c 2.2 plus ion exchange HPLC analyzer for measurement of hemoglobin A1c in diabetic patients.
    Faix, JD
    CLINICAL CHEMISTRY, 1998, 44 : A165 - A165
  • [7] Effect of hemoglobin F and A2 on hemoglobin A1c determined by cation exchange high-performance liquid chromatography
    Can, Murat
    Guven, Berrak
    Akca, Ayse Semra Demir
    JOURNAL OF LABORATORY MEDICINE, 2019, 43 (05) : 265 - 268
  • [8] INTERFERENCE OF ABNORMAL-HEMOGLOBINS ON THE MEASUREMENT OF HEMOGLOBIN-A1C BY ION-EXCHANGE CHROMATOGRAPHY
    ENGBAEK, F
    SORENSEN, GH
    MACINTYRE, B
    CLAUSEN, I
    LUND, HT
    JASTRUP, B
    CLINICA CHIMICA ACTA, 1990, 191 (03) : 239 - 243
  • [9] Measurement of Salicylic Acid by a High-Performance Liquid Chromatography Procedure Based on Ion-Exchange
    M. A. M. Aboul-Soud
    K. Cook
    G. J. Loake
    Chromatographia, 2004, 59 : 129 - 133
  • [10] Measurement of salicylic acid by a high-performance liquid chromatography procedure based on ion-exchange
    Aboul-Soud, MAM
    Cook, K
    Loake, GJ
    CHROMATOGRAPHIA, 2004, 59 (1-2) : 129 - 133