Determining the accuracy of zero-flux and ingestible thermometers in the peri-operative setting

被引:14
|
作者
Jack, James M. [1 ]
Ellicott, Helen [1 ]
Jones, Christopher I. [2 ]
Bremner, Stephen A. [2 ]
Densham, Ian [1 ]
Harper, C. Mark [1 ]
机构
[1] Royal Sussex Cty Hosp, Brighton & Sussex Univ Hosp Trust, Eastern Rd, Brighton BN2 5BE, E Sussex, England
[2] Univ Brighton, Brighton & Sussex Med Sch, Dept Primary Care & Publ Hlth, Brighton BN1 9PH, E Sussex, England
关键词
Temperature measurement; Perioperative; Thermometry; Anaesthesia; DEEP BODY-TEMPERATURE; CORE TEMPERATURE; VALIDITY; SENSOR;
D O I
10.1007/s10877-019-00252-9
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Accurately monitoring peri-operative core temperature is a cornerstone of good practice. Relatively invasive devices such as oesophageal temperature probes and pulmonary artery catheters facilitate this, but are inappropriate for many patients. There remains a need for accurate monitors of core temperature that can be used in awake patients. This study compared the accuracy of two core temperature thermometers that can be used for this purpose: the 3M Bair Hugger (TM) Temperature Monitoring System Zero Flux Thermometer and the CorTempR (TM) Wireless Ingestible Temperature Sensor. Readings were compared with the oesophageal probe, the current intraoperative standard. Thirty patients undergoing elective surgical procedures under general anaesthesia were recruited. The ingestible sensor was ingested prior to induction of anaethesia, and post induction, the zero-flux electrode attached above the right eyebrow and oesophageal probe inserted. During surgery, the temperature on each device was recorded every minute. Measurements were compared using Bland-Altman analysis. The ingestible sensor experienced interference from use of diathermy and fluoroscopy in the operating theatre, rendering 39% of its readings unusable. These were removed from analysis. With remaining readings the bias compared with oesophageal probe was + 0.42 degrees C, with 95% limits of agreement - 2.4 degrees C to 3.2 degrees C. 75.4% of readings were within +/- 0.5 degrees C of the OTP reading. The bias for the zero flux electrode compared to oesophageal probe was + 0.02 degrees C with 95% limits of agreement - 0.5 degrees C to 0.5 degrees C. 97.7% of readings were within +/- 0.5 degrees C of the oesophageal probe. The study findings suggest the zero-flux thermometer is sufficiently accurate for clinical use, whereas the ingestible sensor is not.
引用
收藏
页码:1113 / 1118
页数:6
相关论文
共 50 条
  • [31] Peri-operative diagnostic accuracy of magnetic resonance imaging and ultrasonography in breast cancer: A meta-analysis
    Mouabbi, Jason A.
    Sako, Zeyad
    Abdalla, Ahmed
    Hadid, Tarik
    Rimawi, Mothaffar F.
    CANCER RESEARCH, 2021, 81 (04)
  • [32] Utility of VExUS score in the peri-operative care unit, intensive care unit, and emergency setting - A systematic review
    Gupta, Bhavna
    Ahluwalia, Pallavi
    Gupta, Anish
    Ranjan, Nikhil
    Kakkar, Kamna
    Aneja, Parag
    INDIAN JOURNAL OF ANAESTHESIA, 2023, 67 (SUPPL 4) : S218 - S226
  • [33] Differences between patients' and clinicians' research priorities from the Anaesthesia and Peri-operative Care Priority Setting Partnership
    Boney, O.
    Nathanson, M. H.
    Grocott, M. P. W.
    Metcalf, L.
    ANAESTHESIA, 2017, 72 (09) : 1134 - 1138
  • [34] Effect of thermometry method on recognition of peri-operative hypothermia and patient throughput: Traditional infra-red tympanic vsnon-invasive zero heat flux thermometry
    Searle, J.
    Fawcett, E.
    Gove, R.
    Mandour, Y.
    MacKinnon, J.
    ANAESTHESIA, 2014, 69 : 41 - 41
  • [35] Normal values of plasma methoxyamines in the setting of renal insufficiency and peri-operative stress.: Consequences for the etiological diagnosis of hypertension
    Mulliez, E
    Rouaix-Emery, N
    Pisanu, A
    Wémeau, J
    Proye, C
    ANNALES D ENDOCRINOLOGIE, 2001, 62 (05) : 437 - 441
  • [36] Immune checkpoint inhibitors in the peri-operative setting in renal cell carcinoma: A meta-analysis of the randomized clinical trials.
    Esteban Villarrubia, Jorge
    Carril-Ajuria, Lucia
    Carretero-Gonzalez, Alberto
    Romero Ferreiro, Carmen
    Alvarez, Pablo
    Ruiz-Vico, Maria
    Castellano, Daniel
    de Velasco, Guillermo
    JOURNAL OF CLINICAL ONCOLOGY, 2023, 41
  • [37] Speaking up for safety in surgery: understanding demographic and professional variables influencing team communication and 'speaking up' in a peri-operative setting
    Heritage, Rebecca
    Broadley, Rachel
    ANAESTHESIA, 2024, 79 : 38 - 38
  • [38] Development of a Preoperative Predictive Model for Intra- or Peri-operative Major Complications with High Accuracy Validated with 558 ASD Patients
    Ames, Christopher Pearson
    Scheer, Justin
    Smith, Justin
    Schwab, Frank
    Lafage, Virginie
    Shaffrey, Christopher
    Bess, Robert
    Ailon, Tamir
    Burton, Douglas
    Klineberg, Eric
    JOURNAL OF NEUROSURGERY, 2016, 124 (04) : A1156 - A1157
  • [39] NOD2 DEFICIENCY IMPAIRS PERI-OPERATIVE OUTCOME AND TRANSEPITHELIAL ION FLUX BALANCE IN A MURINE MODEL OF SHORT BOWEL SYNDROME
    Berlin, Peggy
    Witte, Maria B.
    Reiner, Johannes
    Steinig, Julia
    Lindemann, Sabeth
    Bannert, Karen
    Bartsch, Madeleine
    Jaster, Robert
    Lamprecht, Georg
    GASTROENTEROLOGY, 2017, 152 (05) : S188 - S188
  • [40] Peri-operative aromatase inhibitor treatment in determining or predicting longterm outcome in early breast cancer - The POETIC* Trial (CRUK/07/015)
    Robertson, John F. R.
    Dowsett, Mitchell
    Bliss, Judith M.
    Morden, James P.
    Wilcox, Maggie
    Evans, Abigail
    Holcombe, Chris
    Horgan, Kieran
    Kirwan, Cliona
    Mallon, Elizabeth
    Sibbering, Mark
    Skene, Anthony
    Vidya, Raghavan
    Cheang, Maggie
    Banerji, Jane
    Kilburn, Lucy
    Dodson, Andrew
    Smith, Ian
    CANCER RESEARCH, 2018, 78 (04)