Determining the accuracy of zero-flux and ingestible thermometers in the peri-operative setting

被引:14
|
作者
Jack, James M. [1 ]
Ellicott, Helen [1 ]
Jones, Christopher I. [2 ]
Bremner, Stephen A. [2 ]
Densham, Ian [1 ]
Harper, C. Mark [1 ]
机构
[1] Royal Sussex Cty Hosp, Brighton & Sussex Univ Hosp Trust, Eastern Rd, Brighton BN2 5BE, E Sussex, England
[2] Univ Brighton, Brighton & Sussex Med Sch, Dept Primary Care & Publ Hlth, Brighton BN1 9PH, E Sussex, England
关键词
Temperature measurement; Perioperative; Thermometry; Anaesthesia; DEEP BODY-TEMPERATURE; CORE TEMPERATURE; VALIDITY; SENSOR;
D O I
10.1007/s10877-019-00252-9
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Accurately monitoring peri-operative core temperature is a cornerstone of good practice. Relatively invasive devices such as oesophageal temperature probes and pulmonary artery catheters facilitate this, but are inappropriate for many patients. There remains a need for accurate monitors of core temperature that can be used in awake patients. This study compared the accuracy of two core temperature thermometers that can be used for this purpose: the 3M Bair Hugger (TM) Temperature Monitoring System Zero Flux Thermometer and the CorTempR (TM) Wireless Ingestible Temperature Sensor. Readings were compared with the oesophageal probe, the current intraoperative standard. Thirty patients undergoing elective surgical procedures under general anaesthesia were recruited. The ingestible sensor was ingested prior to induction of anaethesia, and post induction, the zero-flux electrode attached above the right eyebrow and oesophageal probe inserted. During surgery, the temperature on each device was recorded every minute. Measurements were compared using Bland-Altman analysis. The ingestible sensor experienced interference from use of diathermy and fluoroscopy in the operating theatre, rendering 39% of its readings unusable. These were removed from analysis. With remaining readings the bias compared with oesophageal probe was + 0.42 degrees C, with 95% limits of agreement - 2.4 degrees C to 3.2 degrees C. 75.4% of readings were within +/- 0.5 degrees C of the OTP reading. The bias for the zero flux electrode compared to oesophageal probe was + 0.02 degrees C with 95% limits of agreement - 0.5 degrees C to 0.5 degrees C. 97.7% of readings were within +/- 0.5 degrees C of the oesophageal probe. The study findings suggest the zero-flux thermometer is sufficiently accurate for clinical use, whereas the ingestible sensor is not.
引用
收藏
页码:1113 / 1118
页数:6
相关论文
共 50 条
  • [21] Improving communication in the peri-operative setting: the IHI 'Improving joy in work' framework in action
    Viira, D.
    Hill, M.
    ANAESTHESIA, 2019, 74 : 79 - 79
  • [22] Tips and tricks in achieving zero peri-operative opioid used in onco-urologic surgery
    Katims, Andrew B.
    Eilender, Benjamin M.
    Pfail, John L.
    Sim, Alan J.
    Sfakianos, John P.
    WORLD JOURNAL OF UROLOGY, 2022, 40 (06) : 1343 - 1350
  • [23] Tips and tricks in achieving zero peri-operative opioid used in onco-urologic surgery
    Andrew B. Katims
    Benjamin M. Eilender
    John L. Pfail
    Alan J. Sim
    John P. Sfakianos
    World Journal of Urology, 2022, 40 : 1343 - 1350
  • [24] Utilization of base deficit and reliability of base deficit as a surrogate for serum lactate in the peri-operative setting
    Chawla, Lakhmir S.
    Nader, Amirali
    Nelson, Todd
    Govindji, Trusha
    Wilson, Ryan
    Szlyk, Sonia
    Nguyen, Aline
    Junker, Christopher
    Seneff, Michael G.
    BMC ANESTHESIOLOGY, 2010, 10
  • [25] Circulating tumour cell release in the peri-operative setting during curative colorectal cancer surgery
    Spring, K.
    Lim, S.
    Descallar, J.
    Ng, W.
    Chua, W.
    de Souza, P.
    Bokey, L.
    MacKenzie, S.
    ANNALS OF ONCOLOGY, 2015, 26
  • [26] Utilization of base deficit and reliability of base deficit as a surrogate for serum lactate in the peri-operative setting
    Lakhmir S Chawla
    Amirali Nader
    Todd Nelson
    Trusha Govindji
    Ryan Wilson
    Sonia Szlyk
    Aline Nguyen
    Christopher Junker
    Michael G Seneff
    BMC Anesthesiology, 10
  • [27] Prothrombotic mechanisms in a setting of peri-operative bridging with low molecular weight heparins in patients undergoing surgery
    Eijgenraam, P. E.
    ten Cate, H.
    Henskens, Y.
    van den Ham, R.
    ten Cate-Hoek, A.
    JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2015, 13 : 260 - 260
  • [28] Peri-operative Fluid Management in a Regional Acute Hospital Setting - A Single Centre Full Cycle Audit
    Barros, R. Di Marco
    Khan, D. Z.
    Bhatti, M. I.
    Rabie, M. A.
    BRITISH JOURNAL OF SURGERY, 2020, 107 : 177 - 178
  • [29] Differences in patients' and clinicians' research priorities from the anaesthesia and peri-operative care priority setting partnership
    Boney, O.
    Metcalf, L.
    Grocott, M.
    Nathanson, M.
    ANAESTHESIA, 2017, 72 : 68 - 68
  • [30] Efficacy, safety and biomarkers of novel therapeutics and regimens in the peri-operative setting of bone and soft tissue sarcoma
    Bao, Qiyuan
    FRONTIERS IN SURGERY, 2023, 10