Ejection fraction in myocardial perfusion imaging assessed with a dynamic phantom: comparison between IQ-SPECT and LEHR

被引:12
|
作者
Hippelainen, Eero [1 ,2 ]
Makela, Teemu [2 ,3 ]
Kaasalainen, Touko [3 ]
Kaleva, Erna [4 ]
机构
[1] Univ Helsinki, Helsinki Univ Hosp, HUS Med Imaging Ctr, Clin Physiol & Nucl Med, POB 340, FI-00029 Helsinki, Finland
[2] Univ Helsinki, Dept Phys, Helsinki, Finland
[3] Univ Helsinki, HUS Med Imaging Ctr, Radiol, Helsinki Univ Hosp, Helsinki, Finland
[4] Tampere Univ Hosp, Radiat Therapy Unit Lahti, Lahti, Finland
来源
EJNMMI PHYSICS | 2017年 / 4卷
关键词
Myocardial perfusion imaging; SPECT; Gated CT; IQ-SPECT; LEHR; VENTRICULAR VOLUME; PROGNOSTIC VALUE; TIME; PROTOCOL; SYSTEM;
D O I
10.1186/s40658-017-0187-2
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Developments in single photon emission tomography instrumentation and reconstruction methods present a potential for decreasing acquisition times. One of such recent options for myocardial perfusion imaging (MPI) is IQ-SPECT. This study was motivated by the inconsistency in the reported ejection fraction (EF) and left ventricular (LV) volume results between IQ-SPECT and more conventional low-energy high-resolution (LEHR) collimation protocols. IQ-SPECT and LEHR quantitative results were compared while the equivalent number of iterations (EI) was varied. The end-diastolic (EDV) and end-systolic volumes (ESV) and the derived EF values were investigated. A dynamic heart phantom was used to produce repeatable ESVs, EDVs and EFs. Phantom performance was verified by comparing the set EF values to those measured from a gated multi-slice X-ray computed tomography (CT) scan (EFTrue). The phantom with an EF setting of 45, 55, 65 and 70% was imaged with both IQ-SPECT and LEHR protocols. The data were reconstructed with different EI, and two commonly used clinical myocardium delineation software were used to evaluate the LV volumes. Results: The CT verification showed that the phantom EF settings were repeatable and accurate with the EFTrue being within 1% point from the manufacture's nominal value. Depending on EI both MPI protocols can be made to produce correct EF estimates, but IQ-SPECT protocol produced on average 41 and 42% smaller EDV and ESV when compared to the phantom's volumes, while LEHR protocol underestimated volumes by 24 and 21%, respectively. The volume results were largely similar between the delineation methods used. Conclusions: The reconstruction parameters can greatly affect the volume estimates obtained from perfusion studies. IQ-SPECT produces systematically smaller LV volumes than the conventional LEHR MPI protocol. The volume estimates are also software dependent.
引用
收藏
页数:14
相关论文
共 50 条
  • [41] Stress induced depression of the left ventricular ejection fraction assessed with gated myocardial perfusion SPECT
    Lorenz, R.
    Muffert, E.
    Reiners, C.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2007, 34 : S268 - S268
  • [42] Assessment of Left Ventricular Ejection Fraction with multifocal collimators: Comparison between IQ-SPECT, Planar Equilibrium Radionuclide Angiography, and Cardiac Magnetic Resonance
    Pelletier-Galarneau, Matthieu
    Finnerty, Vincent
    Tan, Stephanie
    Authier, Sebastien
    Gregoire, Jean
    Harel, Francois
    JOURNAL OF NUCLEAR MEDICINE, 2018, 59
  • [43] Characteristics of normal myocardial perfusion distribution derived from the IQ-SPECT system: A comparison between gender-combined and specific databases
    Okuda, Koichi
    Nakajima, Kenichi
    Matsuo, Shinro
    Kondo, Chisato
    Sarai, Masayoshi
    Shimotsu, Yoriko Horiguchi
    Onoguchi, Masahisa
    Konishi, Takahiro
    Kinuya, Seigo
    JOURNAL OF NUCLEAR MEDICINE, 2016, 57
  • [44] Assessment of left ventricular ejection fraction with cardiofocal collimators: Comparison between IQ-SPECT, planar equilibrium radionuclide angiography, and cardiac magnetic resonance
    Pelletier-Galarneau, Matthieu
    Finnerty, Vincent
    Tan, Stephanie
    Authier, Sebastien
    Gregoire, Jean
    Harel, Francois
    JOURNAL OF NUCLEAR CARDIOLOGY, 2019, 26 (06) : 1857 - 1864
  • [45] Evaluation of differences between post stress and rest left ventricular ejection fraction assessed with gated myocardial perfusion SPECT
    Verberne, HJ
    Dijkgraaf, MGW
    Van Eck-Smit, BLF
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 2001, 28 (08): : 1169 - 1169
  • [46] Assessment of left ventricular ejection fraction with cardiofocal collimators: Comparison between IQ-SPECT, planar equilibrium radionuclide angiography, and cardiac magnetic resonance
    Matthieu Pelletier-Galarneau
    Vincent Finnerty
    Stephanie Tan
    Sebastien Authier
    Jean Gregoire
    Francois Harel
    Journal of Nuclear Cardiology, 2019, 26 : 1857 - 1864
  • [47] Comparison of ejection fraction calculation between CT and SPECT at high heart rate: A dynamic cardiac phantom study
    Chen, Chien-Cheng
    Shen, Thau-Yun
    Peterson, Christine B.
    Hung, Guang-Uei
    Pan, Tinsu
    JOURNAL OF NUCLEAR CARDIOLOGY, 2021, 28 (01) : 311 - 316
  • [48] Comparison of ejection fraction calculation between CT and SPECT at high heart rate: A dynamic cardiac phantom study
    Chien-Cheng Chen
    Thau-Yun Shen
    Christine B. Peterson
    Guang-Uei Hung
    Tinsu Pan
    Journal of Nuclear Cardiology, 2021, 28 : 311 - 316
  • [49] Differences between stress and rest ejection fraction in gated perfusion SPECT imaging.
    Bokhari, S
    Blood, DK
    Bergmann, SR
    JOURNAL OF NUCLEAR MEDICINE, 1999, 40 (05) : 45P - 45P
  • [50] Gated SPECT myocardial imaging: severity of perfusion abnormalities and difference in ejection fraction in patients with ischemia and myocardial infarction
    Bosnjakovic, V.
    Sobic-Saranovic, D.
    Pavlovic, S.
    Petrasinovic, Z.
    Kozarevic, N.
    Todorovic-Tirnanic, M.
    Ostojic, M.
    Artiko, V.
    Obradovic, V.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2006, 33 : S235 - S236