Ejection fraction in myocardial perfusion imaging assessed with a dynamic phantom: comparison between IQ-SPECT and LEHR

被引:12
|
作者
Hippelainen, Eero [1 ,2 ]
Makela, Teemu [2 ,3 ]
Kaasalainen, Touko [3 ]
Kaleva, Erna [4 ]
机构
[1] Univ Helsinki, Helsinki Univ Hosp, HUS Med Imaging Ctr, Clin Physiol & Nucl Med, POB 340, FI-00029 Helsinki, Finland
[2] Univ Helsinki, Dept Phys, Helsinki, Finland
[3] Univ Helsinki, HUS Med Imaging Ctr, Radiol, Helsinki Univ Hosp, Helsinki, Finland
[4] Tampere Univ Hosp, Radiat Therapy Unit Lahti, Lahti, Finland
来源
EJNMMI PHYSICS | 2017年 / 4卷
关键词
Myocardial perfusion imaging; SPECT; Gated CT; IQ-SPECT; LEHR; VENTRICULAR VOLUME; PROGNOSTIC VALUE; TIME; PROTOCOL; SYSTEM;
D O I
10.1186/s40658-017-0187-2
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Developments in single photon emission tomography instrumentation and reconstruction methods present a potential for decreasing acquisition times. One of such recent options for myocardial perfusion imaging (MPI) is IQ-SPECT. This study was motivated by the inconsistency in the reported ejection fraction (EF) and left ventricular (LV) volume results between IQ-SPECT and more conventional low-energy high-resolution (LEHR) collimation protocols. IQ-SPECT and LEHR quantitative results were compared while the equivalent number of iterations (EI) was varied. The end-diastolic (EDV) and end-systolic volumes (ESV) and the derived EF values were investigated. A dynamic heart phantom was used to produce repeatable ESVs, EDVs and EFs. Phantom performance was verified by comparing the set EF values to those measured from a gated multi-slice X-ray computed tomography (CT) scan (EFTrue). The phantom with an EF setting of 45, 55, 65 and 70% was imaged with both IQ-SPECT and LEHR protocols. The data were reconstructed with different EI, and two commonly used clinical myocardium delineation software were used to evaluate the LV volumes. Results: The CT verification showed that the phantom EF settings were repeatable and accurate with the EFTrue being within 1% point from the manufacture's nominal value. Depending on EI both MPI protocols can be made to produce correct EF estimates, but IQ-SPECT protocol produced on average 41 and 42% smaller EDV and ESV when compared to the phantom's volumes, while LEHR protocol underestimated volumes by 24 and 21%, respectively. The volume results were largely similar between the delineation methods used. Conclusions: The reconstruction parameters can greatly affect the volume estimates obtained from perfusion studies. IQ-SPECT produces systematically smaller LV volumes than the conventional LEHR MPI protocol. The volume estimates are also software dependent.
引用
收藏
页数:14
相关论文
共 50 条
  • [31] Feasibility of an eighth-time gated myocardial perfusion SPECT functional imaging using IQ-SPECT system
    Caobelli, F.
    Thackeray, J. T.
    Soffientini, A.
    Bengel, F. M.
    Pizzocaro, C.
    Guerra, U. P.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2015, 42 : S229 - S230
  • [32] Feasibility of one-eighth time gated myocardial perfusion SPECT functional imaging using IQ-SPECT
    Caobelli, Federico
    Thackeray, James T.
    Soffientini, Alberto
    Bengel, Frank M.
    Pizzocaro, Claudio
    Guerra, Ugo Paolo
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2015, 42 (12) : 1920 - 1928
  • [33] Application of IQ.SPECT in Myocardial Perfusion Imaging: Comparison with conventional SPECT
    Jin, Xiaoxiao
    Wei, Hua
    Wu, Zhifang
    Liu, Jianzhong
    Hu, Guang
    Wang, Jin
    Li, Sijin
    JOURNAL OF NUCLEAR MEDICINE, 2017, 58
  • [34] IQ-SPECT protocol allows to greatly reduce administered dose or acquisition time in Myocardial Perfusion Imaging: evidences from a phantom study.
    Caobelli, F.
    Kaiser, S. Ren
    Chieregato, M.
    Zoccarato, O.
    Galelli, M.
    Soffientini, A.
    Pizzocaro, C.
    Savelli, G.
    Guerra, U. P.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2014, 41 : S612 - S612
  • [35] Feasibility of an eighth-time gated myocardial perfusion SPECT functional imaging using IQ-SPECT system.
    Caobelli, Federico
    Soffientini, Alberto
    Thackeray, James
    Bengel, Frank
    Pizzocaro, Claudio
    Guerra, Ugo
    JOURNAL OF NUCLEAR MEDICINE, 2015, 56 (03)
  • [36] Short time stress only myocardial perfusion IQ-SPECT imaging in clinical routine: initial results
    Valenta, Ines
    Rajaram, Ramya
    Platsch, Guenther
    Schindler, Thomas
    JOURNAL OF NUCLEAR MEDICINE, 2017, 58
  • [37] Optimisation of IQ. SPECT in Myocardial Perfusion Imaging - Comparison with Conventional SPECT and Echocardiography Imaging
    Joel, E.
    Bouchareb, Y.
    Haroon, A.
    Luqman, M.
    Newell, M.
    Jan, H.
    2014 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2014,
  • [38] Evaluation of patients with coronary artery disease IQ-SPECT protocol in myocardial perfusion imaging: Preliminary results
    Caobelli, F.
    Pizzocaro, C.
    Paghera, B.
    Guerra, U. P.
    NUKLEARMEDIZIN-NUCLEAR MEDICINE, 2013, 52 (05): : 178 - 185
  • [39] Is a misposition of the heart likely to affect results of myocardial perfusion imaging when using IQ-SPECT protocol?
    Soffientini, A.
    Mostarda, A.
    Massetti, V.
    Andreoli, M.
    Zoccarato, O.
    Pizzocaro, C.
    Guerra, U. P.
    Sayellil, G.
    Caobellli, F.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2014, 41 : S208 - S208
  • [40] Comparison of Left Ventricular Ejection Fraction Calculation Methods between CT and SPECT with a Dynamic Cardiac Phantom
    Pan, Tinsu