Superficial dose evaluation of four dose calculation algorithms

被引:14
|
作者
Cao, Ying [1 ,2 ]
Yang, Xiaoyu [2 ]
Yang, Zhen [2 ]
Qiu, Xiaoping [1 ]
Lv, Zhiping [2 ]
Lei, Mingjun [2 ]
Liu, Gui [2 ]
Zhang, Zijian [2 ]
Hu, Yongmei [2 ]
机构
[1] Univ South China, Sch Nucl Sci & Technol, Hengyang 421000, Peoples R China
[2] Cent S Univ, Xiangya Hosp, Oncol Dept, Changsha 410000, Hunan, Peoples R China
关键词
Superficial dose; Skin dose; Algorithm; Monte Carlo simulation; Film; MONTE-CARLO; PHOTON BEAMS; RADIATION TREATMENT; RADIOCHROMIC FILM; RADIOTHERAPY; DOSIMETRY; IMPLEMENTATION; CARCINOMA; IMPACT;
D O I
10.1016/j.radphyschem.2016.02.032
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Accurate superficial dose calculation is of major importance because of the skin toxicity in radiotherapy, especially within the initial 2 mm depth being considered more clinically relevant. The aim of this study is to evaluate superficial dose calculation accuracy of four commonly used algorithms in commercially available treatment planning systems (TPS) by Monte Carlo (MC) simulation and film measurements. The superficial dose in a simple geometrical phantom with size of 30 cm x 30 cm x 30 cm was calculated by PBC (Pencil Beam Convolution), AAA (Analytical Anisotropic Algorithm), AXB (Acuros XB) in Eclipse system and CCC (Collapsed Cone Convolution) in Raystation system under the conditions of source to surface distance (SSD) of 100 cm and field size (FS) of 10 x 10 cm(2). EGSnrc (BEAMnrc/DOSXYZnrc) program was performed to simulate the central axis dose distribution of Varian Trilogy accelerator, combined with measurements of superficial dose distribution by an extrapolation method of multilayer radiochromic films, to estimate the dose calculation accuracy of four algorithms in the superficial region which was recommended in detail by the ICRU (International Commission on Radiation Units and Measurement) and the ICRP (International Commission on Radiological Protection). In superficial region, good agreement was achieved between MC simulation and film extrapolation method, with the mean differences less than 1%, 2% and 5% for 0, 30 degrees and 60 degrees, respectively. The relative skin dose errors were 0.84%, 1.88% and 3.90%; the mean dose discrepancies (0, 30 and 60) between each of four algorithms and MC simulation were (2.41 +/- 1.55%, +/- 3.11 2.40%, and 1.53 +/- 1.05%), (3.09 +/- 3.00%, 3.10 +/- 3.01%, and 3.77 +/- 3.59%), (3.16 +/- 1.50%, 8.70 +/- 2.84%, and 18.20 +/- 4.10%) and (14.45 +/- 4.66%, 10.74 +/- 4.54%, and 334 +/- 3.26%) for AXB, CCC, AAA and PBC respectively. Monte Carlo simulation verified the feasibility of the superficial dose measurements by multilayer Gafchromic films. And the rank of superficial dose calculation accuracy of four algorithms was AXB > CCC > AAA > PBC. Care should be taken when using the AAA and PBC algorithms in the superficial dose calculation. (C) 2016 Elsevier Ltd. All rights reserved.
引用
收藏
页码:23 / 28
页数:6
相关论文
共 50 条
  • [1] Superficial dose verification of four dose calculation algorithms
    Cao, Y.
    Yang, Z.
    Yang, X.
    Qiu, X.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S166 - S166
  • [2] Comparison of four commercial dose calculation algorithms in different evaluation tests
    Rostami, Aram
    De Castro Neto, Aluisio Jose
    Paloor, Satheesh Prasad
    Khalid, Abdul Sattar
    Hammoud, Rabih
    [J]. JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2023, 31 (05) : 1013 - 1033
  • [3] Dose distribution evaluation of various dose calculation algorithms in inhomogeneous media
    Kim, Y. L.
    Suh, T. S.
    Choe, B. Y.
    Choi, B. O.
    Chung, J. B.
    Lee, J. W.
    Bae, Y. K.
    Park, B. M.
    Jung, J. Y.
    Shin, Y. J.
    Kim, J. Y.
    Moon, S. K.
    [J]. INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2016, 14 (04): : 269 - 278
  • [4] Performance Assessment of Four Dose Calculation Algorithms for Spine SBRT
    Fernandez, A.
    Paredes, L.
    Diaz Moreno, R.
    Venencia, C.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E557 - E558
  • [5] Evaluation of the usefulness of dose calculation algorithms in radiotherapy planning
    Kieltyka, B.
    Rawojc, K.
    Kisielewicz, K.
    Markiewicz, I.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2016, 118 : S88 - S89
  • [6] Evaluation of two generic algorithms for CBCT based dose calculation
    Thing, R.
    Nilsson, R.
    Andersson, S.
    Berg, M.
    Lund, M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2022, 170 : S1252 - S1253
  • [7] New evaluation methodology of dose distributions for advanced calculation algorithms
    Jurado-Bruggeman, D.
    Onses-Segarra, A.
    Munoz-Montplet, C.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S975 - S976
  • [8] Dosimetric verification of four dose calculation algorithms for spine stereotactic body radiotherapy
    Hirashima, Hideaki
    Nakamura, Mitsuhiro
    Nakamura, Kiyonao
    Matsuo, Yukinori
    Mizowaki, Takashi
    [J]. JOURNAL OF RADIATION RESEARCH, 2024, 65 (01) : 109 - 118
  • [9] Evaluation of Dose Level Differences Across a Range of Target Sizes for Two Commercial Dose Calculation Algorithms
    Gardner, S.
    Wang, S.
    Snyder, K.
    Huang, Y.
    Smith, C.
    Chetty, I.
    Miller, B.
    [J]. MEDICAL PHYSICS, 2022, 49 (06) : E931 - E931
  • [10] Evaluation of dose calculation algorithms for dynamic are treatments of head and neck tumors
    Linthout, N
    Verellen, D
    Storme, G
    [J]. MEDICAL PHYSICS, 2002, 29 (06) : 1368 - 1368