An innovative tool for groundwater velocity measurement compared with other tools in laboratory and field tests

被引:8
|
作者
Essouayed, E. [1 ]
Annable, M. D. [2 ]
Momtbrun, M. [1 ]
Atteia, O. [3 ]
机构
[1] Univ Bordeaux Montaigne, EA Georessources & Environm 4592, Bordeaux INP ENSEGID, INNOVASOL, 1 Allee Daguin, F-33600 Pessac, France
[2] Univ Florida, Dept Environm Engn Sci, Gainesville, FL 32611 USA
[3] Univ Bordeaux Montaigne, Carnot ISIFoR, EA Georessources & Environm 4592, Bordeaux INP ENSEGID, 1 Allee Daguin, F-33600 Pessac, France
来源
JOURNAL OF HYDROLOGY X | 2019年 / 2卷
关键词
Direct velocity tool; Darcy flux measurement; Laboratory and field comparison; Innovative;
D O I
10.1016/j.hydroa.2018.100008
中图分类号
P [天文学、地球科学];
学科分类号
07 ;
摘要
An innovative solution for groundwater velocity measurement in wells, the Direct Velocity Tool (DVT), is developed and tested. The DVT allows measurement of Darcy flux at centimeter scale vertical resolution requiring only a few minutes for each measurement (typically 5-10 min). Results are generated in real time through the use of an equation. The theoretical functional Darcy flux range is between 1 and 40 cm.day (1) for an aquifer with a hydraulic conductivity, K, less than 10 (3) m.s (1). Laboratory tests showed a linear response between Darcy flux imposed and Darcy flux measured with the DVT, with a high correlation coefficient (R-2 = 0.99). The DVT was tested for velocities ranging from 5 to 30 cm.day (1) where a higher standard deviation was observed for higher velocity (+/- 6.46 cm.day 1 for 30 cm.day (1)). At a hydraulically controlled field site, tracer tests, borehole dilutions, Passive Flux Meter (PFM) and DVT were tested under constant flow conditions. The PFM, tracer test and DVT technologies measured similar Darcy flux ranging from 10 to 17 cm.day (1). The DVT and PFM had uncertainties of 2.70 cm.day (1) and 2.50 cm.day (1), respectively. At a contaminated field site, velocities measured with the PFM range between 5 and 9 cm.day (1), while the DVT provided velocities of 4 to 8 cm.day (1). Velocity measurement showed similar results at the contaminated field site between PFM and DVT. The DVT offers an innovative solution for Darcy flux measurement and can be deployed easily. It provides the capability to characterize the vertical distribution of the horizontal velocity for a well in a period of 1-2 h. (C) 2018 The Author(s). Published by Elsevier B.V.
引用
收藏
页数:9
相关论文
共 44 条
  • [21] A bioassay method validation framework for laboratory and semi-field tests used to evaluate vector control tools
    Matope, Agnes
    Lees, Rosemary S.
    Spiers, Angus
    Foster, Geraldine M.
    MALARIA JOURNAL, 2023, 22 (01)
  • [22] A bioassay method validation framework for laboratory and semi-field tests used to evaluate vector control tools
    Agnes Matope
    Rosemary S. Lees
    Angus Spiers
    Geraldine M. Foster
    Malaria Journal, 22
  • [23] LABORATORY AND FIELD TESTS OF ZINEB AND OTHER FUNGICIDES AGAINST DOWNY MILDEW OF LETTUCE (BREMIA LACTUCAE REGEL)
    CHANNON, AG
    WEBB, MJW
    ANNALS OF APPLIED BIOLOGY, 1967, 59 (03) : 355 - &
  • [24] SUGAR-BEET (BETA-VULGARIS L) SEED VIGOR COMPARED IN LABORATORY AND FIELD-TESTS
    LOVATO, A
    CAGALLI, S
    SEED SCIENCE AND TECHNOLOGY, 1993, 21 (01) : 61 - 67
  • [25] Effect of natural cold source on groundwater source heat pump according to laboratory and field geotechnical thermal physical tests
    Zhang, Qing
    Zhang, Yanjun
    Gao, Ping
    Yu, Ziwang
    ENERGY AND BUILDINGS, 2014, 84 : 557 - 566
  • [26] Validation of nutritional screening Malnutrition Screening Tool compared to other screening tools and the nutritional assessment in different social and health areas
    Castro-Vega, Iciar
    Veses Martin, Silvia
    Cantero Llorca, Juana
    Salom Vendrell, Christian
    Banuls, Celia
    Hernandez-Mijares, Antonio
    NUTRICION HOSPITALARIA, 2018, 35 (02) : 351 - 358
  • [27] Total Cost of Ownership (tco) measurement is a new evidence-based tool to know the real cost of laboratory tests
    Stankevich, L.
    Gontard, P.
    Gorodetsky, B.
    CLINICA CHIMICA ACTA, 2019, 493 : S674 - S675
  • [28] Validation of an innovative Polysomnography System for patient-side Self-Application in the Home compared to established supervised Sleep Laboratory Measurement
    Schoebel, C.
    Sundrup, Grosse M.
    Zentgraf, C.
    Dietz-Terjung, S.
    Taube, C.
    PNEUMOLOGIE, 2023, 77 : S95 - S96
  • [29] System for In Situ Measurement of Neutral Temperature in Continuous-Welded Rail: Results from Laboratory and Field Tests
    Nucera, Claudio
    Phillips, Robert
    di Scalea, Francesco Lanza
    Fateh, Mahmood
    Carr, Gary
    TRANSPORTATION RESEARCH RECORD, 2013, (2374) : 154 - 161
  • [30] Interferon Gamma Release Assay, A Powerful Tool for the Detection of Human and Bovine Tuberculosis in the Greater Cairo Area Compared to Other Diagnostic Tools
    Metwally, Mirihan A.
    Yassin, Aymen S.
    Riad, Emad M.
    Hamouda, Hayam M.
    Amin, Magdy A.
    SAINS MALAYSIANA, 2020, 49 (04): : 785 - 792