The National Cardiovascular Center electrohydraulic total artificial heart and ventricular assist device systems: Current status of development

被引:13
|
作者
Tatsumi, E
Taenaka, Y
Homma, A
Nishinaka, T
Takewa, Y
Tsukiya, T
Ohnishi, H
Oshikawa, M
Shirakawa, Y
Kakuta, Y
Shioya, K
Katagiri, N
Mizuno, T
Kamimura, T
Takano, H
Tsukahara, K
Tsuchimoto, K
Wakui, H
Yamaguchi, H
机构
[1] Natl Cardiovasc Ctr, Res Inst, Dept Artificial Organs, Suita, Osaka 5658565, Japan
[2] Aisin Cosmos Res & Dev Co Ltd, Kariya, Aichi 448, Japan
关键词
D O I
10.1097/01.MAT.0000065374.83106.FE
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Electrohydraulic total artificial heart (EHTAH) and electrohydraulic ventricular assist device (EHVAD) systems have been developed in our institute. The EHTAH system comprises a pumping unit consisting of blood pumps and an actuator, as well as an electronic unit consisting of an internal controller, internal and external batteries, and transcutaneous energy transfer (TET) and optical telemetry (TOT) subunits. The actuator, placed outside the pericardial space, reciprocates and delivers hydraulic silicone oil to the alternate blood pumps through a pair of flexible oil conduits. The pumping unit with an external controller was implanted in 10 calves as small as 55 kg. Two animals survived for more than 12 weeks in a good general condition. The assumed cardiac output ranged between 6 and 10 L/min, the power consumption was 12-18 W, and the energy efficiency was estimated to be 9-11%. initial implantation of subtotal system including electronic units was further conducted in another calf weighing 73 kg. It survived for 3 days with a completely tether free system. The EHVAD system is developed by using the left blood pump and the actuator of the EHTAH, which were packaged in a compact metal casing with a compliance chamber. In vitro testing demonstrated maximum output more than 9 L/min and more than 13% maximum efficiency. The initial animal testing lasted for 25 days. These results indicate that our EHTAH and EHVAD have the potential to be totally implantable systems.
引用
收藏
页码:243 / 249
页数:7
相关论文
共 50 条
  • [1] Current status of development and in vivo evaluation of the National Cardiovascular Center electrohydraulic total artificial heart system
    Tatsumi E.
    Taenaka Y.
    Uesho K.
    Homma A.
    Nishinaka T.
    Kakuta Y.
    Tsukiya T.
    Takano H.
    Masuzawa T.
    Nakamura M.
    Koshiji K.
    Fukui Y.
    Tsukahara K.
    Tsuchimoto K.
    Wakui H.
    [J]. Journal of Artificial Organs, 2000, 3 (2) : 62 - 69
  • [2] In vivo evaluation of the national cardiovascular center electrohydraulic total artificial heart
    Tatsumi, E
    Masuzawa, T
    Nakamura, M
    Taenaka, Y
    Nishimura, T
    Endo, S
    Zhang, B
    Kakuta, Y
    Nakata, M
    Nakamura, T
    Nishinaka, T
    Takano, H
    Tsukahara, K
    Tsuchimoto, K
    [J]. ARTIFICIAL ORGANS, 1999, 23 (03) : 242 - 248
  • [3] Development of a pneumatic ventricular assist device and a pnumatic total artificial heart system
    Homma, Akihiko
    Sumikura, Hirohito
    Ohnuma, Kentaro
    Ohgoe, Yasuharu
    Takewa, Yoshiaki
    Tatsumi, Eisuke
    Taenaka, Yoshiyuki
    Fukui, Yasuhiro
    Mukaibayashi, Hiroshi
    Kojima, Kouichi
    [J]. Transactions of Japanese Society for Medical and Biological Engineering, 2014, 52
  • [4] In vivo long-term evaluation of the national cardiovascular center electrohydraulic total artificial heart system
    Tatsumi, E.
    Taenaka, Y.
    Uesho, K.
    Homma, A.
    Nishinaka, T.
    Kakuta, Y.
    Tsukiya, T.
    Katagiri, N.
    Imada, K.
    Takahashi, M.
    Ohnishi, Y.
    Sato, K.
    Takano, H.
    Masuzawa, T.
    Nakamura, M.
    [J]. ASAIO Journal, 2000, 46 (02)
  • [5] Current status of the National Cardiovascular Center totally implantable artificial heart system
    Homma, A
    Taenaka, Y
    Tatsumi, E
    Takewa, Y
    Kamimura, T
    Naitoh, H
    Oshikawa, A
    Mizuno, T
    Shioya, K
    Lee, HS
    Tsukiya, T
    Kakuta, Y
    Katagiri, N
    Takano, H
    Kitamura, S
    Koshiji, K
    Shiba, K
    Fukui, Y
    Wakui, H
    Tsuchimoto, K
    Fukuda, H
    Tsukahara, K
    [J]. SICE 2004 ANNUAL CONFERENCE, VOLS 1-3, 2004, : 436 - 441
  • [6] Gastrointestinal Bleeding in Ventricular Assist Device and Total Artificial Heart Implantations
    Lachar, Jatinder
    Christopher, Sanah
    Flattery, Maureen
    Shah, Keyur
    Tang, Daniel G.
    Smallfield, Melissa C.
    Cooke, Richard H.
    Kasirajan, Vigneshwar
    Cheng, Jianfeng
    Smallfield, George
    [J]. GASTROENTEROLOGY, 2014, 146 (05) : S322 - S322
  • [7] DEVELOPMENT OF ELECTROHYDRAULIC POWER SYSTEMS FOR USE IN ARTIFICIAL HEART AND CIRCULATORY ASSIST DEVICES
    BURNS, WH
    GRIFFITH, NJ
    [J]. MECHANICAL ENGINEERING, 1969, 91 (04): : 55 - &
  • [8] Left Ventricular Assist Device Failure With Total Artificial Heart Bridge to Heart Transplant
    Tushak, Zackary
    Wood, Taylor
    Quader, Mohammed
    Shah, Keyur B.
    [J]. CIRCULATION, 2020, 142
  • [9] USE OF THE TOTAL ARTIFICIAL-HEART AND VENTRICULAR ASSIST DEVICE AS A BRIDGE TO TRANSPLANTATION
    PIFARRE, R
    SULLIVAN, H
    MONTOYA, A
    BAKHOS, M
    GRIECO, J
    FOY, BK
    BLAKEMAN, B
    ALTERGOTT, R
    LONCHYNA, V
    CALANDRA, D
    HINKAMP, T
    GRADY, K
    COSTANZONORDIN, MR
    [J]. JOURNAL OF HEART TRANSPLANTATION, 1990, 9 (06): : 638 - 643
  • [10] Compact Ventricular Assist Device and Total Artificial Heart Using Undulation Pump
    Imachi, Kou
    Saito, Itsuro
    Takiura, Koki
    Chinzei, Tsuneo
    Isoyama, Takashi
    Yambe, Tomoyuki
    Shiraishi, Yasuhiro
    Miura, Hidekazu
    Matsuki, Hidetoshi
    Mitamura, Yoshinori
    Inoue, Yusuke
    Okamoto, Eiji
    Umezu, Mitsuo
    Nemoto, Isao
    Abe, Yusuke
    [J]. WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2006, VOL 14, PTS 1-6, 2007, 14 : 3146 - +