Noninvasive Autoregulation Monitoring with and without Intracranial Pressure in the Naive Piglet Brain

被引:43
|
作者
Brady, Ken M. [1 ]
Mytar, Jennifer O. [1 ]
Kibler, Kathleen K. [1 ]
Hogue, Charles W., Jr. [1 ]
Lee, Jennifer K. [1 ]
Czosnyka, Marek [2 ]
Smielewski, Peter [2 ]
Easley, R. Blaine [1 ]
机构
[1] Johns Hopkins Univ Hosp, Dept Anesthesiol & Crit Care Med, Baltimore, MD 21287 USA
[2] Addenbrookes Hosp, Dept Acad Neurosurg, Cambridge, England
来源
ANESTHESIA AND ANALGESIA | 2010年 / 111卷 / 01期
关键词
CEREBRAL-BLOOD-FLOW; PERFUSION-PRESSURE; HEAD-INJURY; LOWER LIMIT; REACTIVITY; TIME;
D O I
10.1213/ANE.0b013e3181e054ba
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
BACKGROUND: Cerebrovascular autoregulation monitoring is often desirable for critically ill patients in whom intracranial pressure (ICP) is not measured directly. Without ICP, arterial blood pressure (ABP) is a substitute for cerebral perfusion pressure (CPP) to gauge the constraint of cerebral blood flow across pressure changes. We compared the use of ABP versus CPP to measure autoregulation in a piglet model of arterial hypotension. METHODS: Our database of neonatal piglet (5-7 days old) experiments was queried for animals with naive ICP that were made lethally hypotensive to determine the lower limit of autoregulation (LLA). Twenty-five piglets were identified, each with continuous recordings of ICP, regional cerebral oximetry (rSo(2)), and cortical red cell flux (laser Doppler). Autoregulation was assessed with the cerebral oximetry index (COx) in 2 ways: linear correlation between ABP and rSo(2) (COx(ABP)) and between CPP and rSo(2) (COx(CPP)). The lower limits of autoregulation were determined from plots of red cell flux versus ABP. Averaged values of COx(ABP) and COx(CPP) from 5 mm Hg ABP bins were used to show receiver operating characteristics for the 2 methods. RESULTS: COx(ABP) and COx(CPP) yielded identical receiver operating characteristic curve areas of 0.91(95% confidence interval [Cl], 0.88-0.95) for determining the LLA. However, the thresholds for the 2 methods differed: a threshold COx(ABP) of 0.5 was 89% sensitive (95% Cl, 81%-94%) and 81% specific (95% Cl, 73%-88%) for detecting ABP below the LLA. A threshold COx(CPP) of 0.42 gave the same 89% sensitivity (95% Cl, 81%-94%) with 77% specificity (95% Cl, 69%-84%). CONCLUSIONS: The use of ABP instead of CPP for autoregulation monitoring in the naive brain with COx results in a higher threshold value to discriminate ABP above from ABP below the LLA. However, accuracy was similar with the 2 methods. These findings support and refine the use of near-infrared spectroscopy to monitor autoregulation in patients without ICP monitors. (Anesth Analg 2010;111:191-5)
引用
收藏
页码:191 / 195
页数:5
相关论文
共 50 条
  • [41] Noninvasive detection of alarming intracranial pressure changes by auditory monitoring in early management of brain injury: a prospective invasive versus noninvasive study
    Giraudet, Fabrice
    Longeras, Francois
    Mulliez, Aurelien
    Thalamy, Aurelie
    Pereira, Bruno
    Avan, Paul
    Sakka, Laurent
    CRITICAL CARE, 2017, 21
  • [42] Validation of Noninvasive Absolute Intracranial Pressure Measurements in Traumatic Brain Injury and Intracranial Hemorrhage
    Kienzler, Jenny C.
    Zakelis, Rolandas
    Babler, Sabrina
    Remonda, Elke
    Ragauskas, Arminas
    Fandino, Javier
    OPERATIVE NEUROSURGERY, 2019, 16 (02) : 186 - 195
  • [43] Noninvasive detection of alarming intracranial pressure changes by auditory monitoring in early management of brain injury: a prospective invasive versus noninvasive study
    Fabrice Giraudet
    François Longeras
    Aurélien Mulliez
    Aurélie Thalamy
    Bruno Pereira
    Paul Avan
    Laurent Sakka
    Critical Care, 21
  • [44] Noninvasive Monitoring of Dynamic Cerebrovascular Autoregulation and ‘Optimal Blood Pressure’ in Normal Adult Subjects
    Paul Pham
    Jessica Bindra
    Anders Aneman
    Alwin Chuan
    John M. Worthington
    Matthias Jaeger
    Neurocritical Care, 2019, 30 : 201 - 206
  • [45] Noninvasive Monitoring of Dynamic Cerebrovascular Autoregulation and "Optimal Blood Pressure' in Normal Adult Subjects
    Pham, Paul
    Bindra, Jessica
    Aneman, Anders
    Chuan, Alwin
    Worthington, John M.
    Jaeger, Matthias
    NEUROCRITICAL CARE, 2019, 30 (01) : 201 - 206
  • [46] A Novel Noninvasive Technique for Intracranial Pressure Waveform Monitoring in Critical Care
    Brasil, Sergio
    Solla, Davi Jorge Fontoura
    Nogueira, Ricardo de Carvalho
    Teixeira, Manoel Jacobsen
    Malbouisson, Luiz Marcelo Sa
    Paiva, Wellingson da Silva
    JOURNAL OF PERSONALIZED MEDICINE, 2021, 11 (12):
  • [47] Identifying Biomarkers for Intracranial Pressure Changes in the Development of a Noninvasive Monitoring Device
    Hawks, Jeff A.
    Bashford, Greg
    Kedar, Sachin
    Thorell, William
    Ghate, Deepta
    JOURNAL OF MEDICAL DEVICES-TRANSACTIONS OF THE ASME, 2016, 10 (02):
  • [48] IMPEDANCE MEASUREMENT AS A NONINVASIVE TECHNIQUE FOR THE MONITORING OF INTRACRANIAL-PRESSURE VARIATIONS
    MAGNANO, M
    ALBERA, R
    LACILLA, M
    GABINI, A
    NADDEO, M
    BRUNO, D
    AUDIOLOGY, 1994, 33 (04): : 237 - 243
  • [49] Glasgow Coma Score in Patients with Severe Traumatic Brain Injury with or without Intracranial Pressure Monitoring
    Scap, Miroslav
    Matas, Marijana
    Jednacak, Hrvoje
    Vukic, Miroslav
    BRITISH JOURNAL OF ANAESTHESIA, 2012, 108 : 73 - 74
  • [50] Noninvasive intracranial pressure monitoring for HIV-associated cryptococcal meningitis
    Bollela, V. R.
    Frigieri, G.
    Vilar, F. C.
    Spavieri, D. L., Jr.
    Tallarico, F. J.
    Tallarico, G. M.
    Andrade, R. A. P.
    de Haes, T. M.
    Takayanagui, O. M.
    Catai, A. M.
    Mascarenhas, S.
    BRAZILIAN JOURNAL OF MEDICAL AND BIOLOGICAL RESEARCH, 2017, 50 (09)