Comparison of ultrasonic Doppler flow monitor, oscillometric, and direct arterial blood pressure measurements in ill dogs

被引:64
|
作者
Bosiack, Ann P. [1 ]
Mann, F. A. [1 ]
Dodam, John R. [1 ]
Wagner-Mann, Colette C. [1 ]
Branson, Keith R. [1 ]
机构
[1] Univ Missouri, Coll Vet Med, Columbia, MO 65211 USA
关键词
cardiovascular monitoring; critical care monitoring; intensive care medicine; small animal critical care; ISOFLURANE-ANESTHETIZED DOGS; NATIONAL STANDARD; CONSCIOUS DOGS; CATS; SPHYGMOMANOMETERS; HYPERTENSION; ACCURACY;
D O I
10.1111/j.1476-4431.2010.00520.x
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Objective To compare blood pressure measurements obtained via ultrasonic Doppler flow monitor (DOP) and 2 oscillometric noninvasive blood pressure monitors (CAR and PAS) to invasive blood pressure (IBP) in hospitalized, conscious dogs with a range of blood pressures. Design Prospective clinical study. Setting University teaching hospital. Animals Eleven client-owned dogs aged between 4 months and 11.5 years (median 6 y), and weighing between 5.8 and 37.5 kg (median 30.2 kg). Interventions Blood pressure measurement. Measurements and Main Results Three consecutive measurements of systolic, diastolic, and mean arterial pressure (MAP) were recorded for each of the 3 indirect devices (only systolic for DOP), along with concurrent IBP measurements. The data were categorized into 3 groups: hypotensive (direct MAP < 80 mm Hg), normotensive (80 mm Hg < direct MAP >= 100 mm Hg), and hypertensive (direct MAP > 100 mm Hg). Each indirect method was compared with the corresponding direct arterial pressure using the Bland-Altman method. Within the hypotensive group, each indirect method overestimated the corresponding IBP. Within the normotensive group all indirect systolic measurements and the PAS diastolic measurements underestimated the corresponding IBP. The remaining indirect measurements overestimated the corresponding IBP. Within the hypertensive group, DOP and CAR systolic measurements underestimated the corresponding IBP, and the remaining indirect measurements overestimated the corresponding IBP. In hypertensive dogs oscillometric systolic measurements were more accurate than MAP. In hypotensive dogs MAP measurements were more accurate than systolic measurements. All indirect measurements were most accurate in hypertensive dogs. Conclusions The noninvasive blood pressure monitors in our study did not meet the validation standards set in human medicine. However, CAR diastolic and MAP measurements within the normotensive group, CAR MAP measurements within the hypertensive group, and PAS diastolic measurements in all groups were close to these standards. All indirect measurements showed greater bias during hypotension. Precision was poorer for all indirect systolic measurements than for MAP.
引用
收藏
页码:207 / 215
页数:9
相关论文
共 50 条
  • [1] Comparison of direct and indirect (oscillometric) measurements of arterial blood pressure in conscious dogs
    Bodey, AR
    Michell, AR
    Bovee, KC
    Buranakurl, C
    Garg, T
    [J]. RESEARCH IN VETERINARY SCIENCE, 1996, 61 (01) : 17 - 21
  • [2] Comparison of the indirect oscillometric and direct arterial methods for blood pressure measurements in anesthetized dogs
    Meurs, KM
    Miller, MW
    Slater, MR
    [J]. JOURNAL OF THE AMERICAN ANIMAL HOSPITAL ASSOCIATION, 1996, 32 (06) : 471 - 475
  • [3] Evaluation of oscillometric and Doppler ultrasonic devices for blood pressure measurements in anesthetized and conscious dogs
    Vachon, Catherine
    Belanger, Marie C.
    Burns, Patrick M.
    [J]. RESEARCH IN VETERINARY SCIENCE, 2014, 97 (01) : 111 - 117
  • [4] A COMPARISON OF DIRECT AND INDIRECT (OSCILLOMETRIC) MEASUREMENTS OF ARTERIAL BLOOD-PRESSURE IN ANESTHETIZED DOGS, USING TAIL AND LIMB CUFFS
    BODEY, AR
    YOUNG, LE
    BARTRAM, DH
    DIAMOND, MJ
    MICHELL, AR
    [J]. RESEARCH IN VETERINARY SCIENCE, 1994, 57 (03) : 265 - 269
  • [5] Evaluation of an oscillometric blood pressure monitor in anesthetized dogs: Agreement with direct measurements and ability to detect hypotension
    Sedgwick, Samantha
    Lorenzutti, Augusto M.
    Araos, Joaquin B.
    Gleed, Robin D.
    Martin-Flores, Manuel
    [J]. RESEARCH IN VETERINARY SCIENCE, 2021, 135 : 162 - 166
  • [6] Arterial blood pressure in dogs by the oscilometric method and the Doppler ultrasonic
    Cabral, R. R.
    Ciasca, B. D.
    Oliveira, V. M. C.
    Vaz-Curado, A. P.
    Larsson, M. H. M. A.
    [J]. ARQUIVO BRASILEIRO DE MEDICINA VETERINARIA E ZOOTECNIA, 2010, 62 (01) : 64 - 71
  • [7] DIRECT ARTERIAL BLOOD PRESSURE MEASUREMENT USING A CAROTID CATHETER IN COMPARISON TO OSCILLOMETRIC AND DOPPLER METHOD.
    Parker, K.
    Carr, A.
    Duke-Novakovski, T.
    [J]. JOURNAL OF VETERINARY INTERNAL MEDICINE, 2012, 26 (03) : 710 - 711
  • [8] Evaluation of oscillometric and Doppler ultrasonic methods of indirect blood pressure estimation in conscious dogs
    Haberman, Christopher E.
    Kang, Chang W.
    Morgan, Jefferson D.
    Brown, Scott A.
    [J]. CANADIAN JOURNAL OF VETERINARY RESEARCH-REVUE CANADIENNE DE RECHERCHE VETERINAIRE, 2006, 70 (03): : 211 - 217
  • [9] Comparison of direct and Doppler arterial blood pressure measurements in rabbits during isoflurane anaesthesia
    Harvey, Louise
    Knowles, Toby
    Murison, Pamela J.
    [J]. VETERINARY ANAESTHESIA AND ANALGESIA, 2012, 39 (02) : 174 - 184
  • [10] Comparison between oscillometric and intra-arterial blood pressure measurements in ill preterm and full-term neonates
    Lalan, Shwetal
    Blowey, Douglas
    [J]. JOURNAL OF THE AMERICAN SOCIETY OF HYPERTENSION, 2014, 8 (01) : 36 - 44